Vesta Charger Wireless Nominal Specifications
Characteristic
MICS MedRadio
Frequency Band
Bandwidth
Modulation
Radiated Power
Range
Transcutaneous Energy Transfer
Frequency Band
Modulation
Radiated Power
Range
Recharge Channel Communication
Frequency Band
Bandwidth
Modulation
Radiated Power
Range
402 – 405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio
Communication Service
(MedRadio)
240 kHz
FSK
-20.6 dBm EIRP
0 to at least 1.5 m
13.56 MHz
Industrial, Scientific, and Medical
radio band (ISM)
Amplitude (slow to optimize
coupling)
< 0.6 W reactive near-field
5 mm to 40 mm
13.56 MHz ± 9.2 ppm
Industrial, Scientific, and Medical
radio band (ISM)
<0.014 MHz
PPM
-6.93 dBm EIRP
5 mm to 40 mm
55