Alternatively, you can remove the charging wand from
your implant site, which will cause your Vesta Charger to
time out and shut off automatically.
Note: If you wish to resume charging your OPTIMIZER
Smart Mini IPG after you terminate a charging session,
please wait for approximately 10 minutes before initiating
a new charging session to allow the temperature of your
OPTIMIZER Smart Mini IPG to return to its baseline
temperature.
6.8.2
Termination of Charging Session Due to IPG
Temperature
To ensure your safety while charging your OPTIMIZER
Smart Mini IPG, the temperature of your IPG is monitored
during the charging process. If the reported temperature
of your OPTIMIZER Smart Mini IPG at the beginning of
the charging session is outside the acceptable
temperature range or if the temperature of your implanted
OPTIMIZER Smart Mini IPG remaining consistently high
for more than 10 minutes while it is being charged, then
your Vesta Charger will emit 3 long beeping tones and
display the Charging IPG Temperature Error screen,
indicated by a thermometer icon in the center of the
screen (see Figure 19). Your Charger will then shut off. If
this should occur, please wait for approximately 10
minutes before initiating a new charging session.
Figure 19: Charging IPG Temperature Error Screen
23