The Call Doctor Alert screen will be followed by the Snooze
Buzzer Alert screen (see Figure 29), or if at night, the Snooze
Alert screen (see Figure 30).
Figure 29: Snooze Buzzer
Figure 30: Snooze Alert Screen
Alert Screen
If a Call Doctor Alert Code appears on the screen of your Vesta
Charger, take note of the code that is displayed and then press
the Power Button on your Vesta Charger to snooze the activated
alert. Afterwards, use the information below to determine your
next course of action.
If the Call Doctor Alert Code "A9", "A19", "A21", "A23",
"A25", or "A27" is displayed, please call the 24-hour
Support Hotline (866-312-5370) and inform them of the
alert code displayed by your Vesta Charger. Then, unless
the alert code displayed is "A9", proceed to charge your
implanted OPTIMIZER Lite IPG.
If the Call Doctor Alert Code "A31" is displayed, it means
that your Vesta Charger has detected repeated internal
errors during its operation. Please call the 24-hour
Support Hotline (866-312-5370) to obtain a replacement
Vesta Charger.
If the Call Doctor Alert Code "A32" is displayed, it means
that you are attempting to use your Vesta Charger on an
unpaired device. If this code is displayed by your Vesta
Charger, perform the following steps:
1. Verify that the Vesta Charger you are using is the
one that was assigned to you and then restart the
charging process.
2. If this code is still displayed after the Charging
Wand has been placed over your implanted
OPTIMIZER Lite IPG and the charging process
has been restarted, please call the 24-hour
Support Hotline (866-312-5370).