Termination Of Charging Session Due To Charging Ipg Timeout; Termination Of Charging Session Due To Low Charger Battery Level - Impulse Dynamics OPTIMIZER Lite System Patient Manual

For the treatment of moderate to severe heart failure
Table of Contents

Advertisement

7.8.3
Termination of Charging Session Due to
Charging IPG Timeout
If the duration of the charging session exceeds 5 hours,
your Vesta Charger will emit 3 long beeping tones and
display the Charging IPG Timeout Error screen, indicated
by a flashing hourglass icon in the center of the screen
(see Figure 20). Your Charger will then shut off. If this
should occur, please wait for approximately 10 minutes
before initiating a new charging session.
Figure 20: Charging IPG Timeout Error Screen
7.8.4
Termination of Charging Session Due to Low
Charger Battery Level
If the battery charge level of your Vesta Charger drops
below 10% during a charging session, your Vesta
Charger will emit 3 long beeping tones and display the
Low Charger Battery Alert screen, indicated by an empty
battery icon with a flashing "X" over it (see Figure 21).
Your Vesta Charger will then shut off.
Figure 21: Low Charger Battery Alert Screen
If this occurs, recharge the battery of your Vesta Charger
as described in Section 7.6.