9 Service and maintenance
For a reliable function and high measurement quality it is necessary to
inspect and service the analyser regularly.
Besides the regular routine control by the operator (see chapter 9.1.-) MRU
recommends a regular bi-annual maintenance at a minimum, which must
carry by a qualified technician.
Apply personality safety equipment, against high H
concentrations before open the device.
Under special conditions the device can accumulate with toxic
gases.
► Measure the atmosphere, around the device, before
opening it.
► After the opening of the device, ventilate sufficiently.
9.1. Preparing and information about the maintenance
It is important to power off the mains supply before the maintenance can be
started. Even if the main fuse is powered off, dangerous voltage is present.
It can be required to cut off the electric supply and save this from an
accidental switch-on.
For maintenance work on the gas analyser, please take care that dangerous
and toxic gases may leak. The gas supply must be cut-off.
It is important to comply with all country and local codes.
9.2. Regular maintenance work by the operator
All inspections- and maintenance works are dependent from individual
operating conditions, and site. The specified intervals below are only
benchmarks.
Review
Moisture in the analyzer.
Dirt and depositions in analyzer.
Dirt and moisture in the filter-
unit
Testing the gas pipes of leakage
with a "Gas Leak Detector".
Inspecting the conditions of the
gas filters
Inspecting the conditions of the
Filter mats
USER MANUAL SWG100
DANGER
Recommended
intervals
Weekly
Remove the moisture. Call
vendor specialist.
Weekly
Remove the dirt, prevent
further penetration of dirt.
Inform vendor.
Weekly
Exchange the filter- unit.
Weekly
Tighten or exchange the gas
pipe if it is necessary
Every month
Exchange if is necessary
Weekly
Exchange if necessary
S
2
Actions
Page 105 of 158
Need help?
Do you have a question about the SWG100 BIOGAS and is the answer not in the manual?