Equipment Required; Check Procedure - Tektronix 2A61 Instruction Manual

Differential amplifier
Table of Contents

Advertisement

Type
2A61
SECTION
5
PERFORMANCE
CHECK
Introduction
This section of the manual
provides a means of rapidly
checking the performance of the Type 2A61.
It is intended to
check the calibration of the instrument without the need for
performing the complete Calibration Procedure.
The Perform-
ance Check does not provide for the adjustment of any inter-
nal controls.
Failure to meet the requirements given in this
procedure
indicates the need for internal checks or adjust-
ments, and the user should refer to the Calibration Procedure
in this manual.
EQUIPMENT
REQUIRED
The following (or equivalent) equipment is required for a
complete
performance
check of the Type 2A61.
Equipment
specifications given here are the minimum
requirements for
making
the
performance
check.
All
test
instruments
are
assumed
to be calibrated and operating within their rated
specifications.
If substitute equipment is used, it must meet or
exceed the specifications of the recommended equipment.
1. Test Oscilloscope, Tektronix Type
545B with Tektronix
Type H Plug-In Unit. Minimum alternate requirements: Delay-
ing Sweep feature, with a delayed sweep range from .5 ms to
5s and the delaying sweep range from .2ms to .2s; mini-
mum
deflection factor, 100 millivolts/div;
2. Tektronix Type 561A Oscilloscope.
3. Tektronix Type 2B67 Plug-In Time Base Unit, for use in
the Type
561A.
4. Square Wave Generator, Tektronix Type 106. Minimum
alternate requirements; risetime not more than 200 ns. Amp!i-
tude at least 500 millivolts into 50.0; Repetition Rate range
from 10 Hz to 25 kHz.
5. Standard
Amplitude
Calibrator.
Tektronix
Calibration
Fixture, Tektronix
Part Number
067-0502-00.
Minimum
alter-
nate requirements; amplitude accuracy, 0.25%; signal ampli-
tude range, 2 millivolts to 100 millivolts; output signal, 1 kHz
square wave and +DC.
6. Audio Sine Wave Generator, Heath Co. 1G72. Minimum
alternate
requirements;
sine
wave
output;
frequency
range,
60Hz
to 10kHz;
amplitude
range,
100 millivolts, p/p to 5
volts, p/p.
7. 10X
Probe,
Tektronix
P6006.
8. Cable, special 4 pin connector on one end and
BNC
connector
on
the
other.
Tektronix
Calibration
Fixture,
Tek-
tronix part number 067-0531 -00.
9. 1000:1
Divider, Tektronix Calibration
Fixture, Tektronix
part number
067-0529-00.
10. Adapter,
dual
binding
post to BNC
connector.
Tek-
tronix part number 013-0035-00.
©
11. Adapter, BNC female to BNC female.
Tektronix part
number 013-0028-00.
12. Termination,
509,
2 Watt,
General
Radio
Type
874
connector
to BNC
connector.
Tektronix
part number
017-
0083-00.
13. Three patch cords, 18 inch, with banana type terminals
on both
ends.
Tektronix
part number
012-0031-00.
14, Cable,
special
low
noise input.
Tektronix part num-
ber 012-0072-00.
PERFORMANCE
CHECK
PROCEDURE
The following
procedure
uses the equipment listed under
Equipment
Required.
If substitute equipment
is used,
con-
nections and control settings may need to be altered to cor-
respond to the characteristics of the equipment used.
Preliminary Procedure
1. Install a Type 2B67 in the right hand plug-in compart-
ment in the Type 561A Oscilloscope.
2. Install the Type 2A61
plug-in to be checked in the left
hand
plug-in compartment
in the Type 561A
Oscilloscope.
3. Turn the Type 561A
Oscilloscope Power switch to On
and allow a warm-up period of 20 minutes at 25° C, 5°
C (77°
F, +9°
Fh.
4, Set the
Instrument
controls as follows:
Type
561A
Focus
Well defined trace
Intensity
Good
viewing brilliance
Type
2A61
POSITION
Midrange
FREQ
RESPONSE
FROM
06 ~
TO
3 MHz
INPUT SELECTOR
GND
MILLIVOLTS/DIV
.02
VARIABLE
CALIB
Type
2B67
Position
Midrange
Time/div
1 ms
Variable
Calibrated
Pull 5X
Mag
Pushed in
Mode
Norm
5-1