Advanced Configuration Using Display (Signal Properties)
7) Press Right Arrow (at least six times) to scroll to False Echo Edit.
8) Press OK to edit the False EchoCurve.
9) Enter the Starting Point Distance (Start) using the Up Arrow and Right
Arrow.
a) Press Up Arrow to increase the value of the number highlighted. Up
Arrow will scroll from 0 to 9 and back again.
b) Press Right Arrow to move one segment to the right. Right Arrow
will scroll left to right and then back to the first segment.
10) When the value is correct, press OK to save the setting. Enter the Starting
Point Intensity (Amp) using the Up Arrow and Right Arrow.
a) Press Up Arrow to increase the value of the number highlighted. Up
Arrow will scroll from 0 to 9 and back again.
b) Press Right Arrow to move one segment to the right. Right Arrow
will scroll left to right and then back to the first segment.
11) When the value is correct, press OK to save the setting. Enter the Ending
Point Distance (End) using the Up Arrow and Right Arrow.
a) Press Up Arrow to increase the value of the number highlighted. Up
Arrow will scroll from 0 to 9 and back again.
b) Press Right Arrow to move one segment to the right. Right Arrow
will scroll left to right and then back to the first segment.
c) When the value is correct, press OK to save the setting.
12) Enter the Ending Point Intensity (Amp) using the Up Arrow and Right
Arrow.
a) Press Up Arrow to increase the value of the number highlighted. Up
Arrow will scroll from 0 to 9 and back again.
b) Press Right Arrow to move one segment to the right. Right Arrow
will scroll left to right and then back to the first segment.
c) When the value is correct, press OK to save the setting.
13) Review the new false EchoCurve verifying the signals you want removed
are identified.
a) Press OK to continue.
14) Under Save? select YES if the New EchoCurve is correct or select NO if
the New EchoCurve is not correct.
a) Use the Right Arrow and Up Arrow to toggle between Yes and NO.
b) When the value is correct, press OK to save the setting.
15) When done, press ESC to return to the Signal Properties Menu, press
ESC to return to the Advanced Menu, press ESC again to return to the
Configuration Menu and press ESC a fourth time to return to the Main
Screen.
|
40