Warranty
The Blood Pressure Monitor is
guaranteed
Blood Pressure Monitor does not function properly due to defective components or poor
workmanship, we will repair or replace it freely. The warranty does not cover damages to
your Blood Pressue Monitor due to improper handling. Please contact local retailer for
details.
Electromagnetic Compatibility Information
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environm-
ent.
IEC 60601
Compliance
IMMUNITY test
test level
level
± 8 kV
± 8 kV
Electrostatic
contact
contact
discharge (ESD)
±2 kV,±4 kV,
±2 kV,±4 kV,
IEC 61000-4-2
±8 kV,
±8 kV,
±15 kV air
±15 kV air
± 2 kV ,
± 2 kV ,
Electrostatic
transient/burst
100kHz, for
100kHz, for
IEC 61000-4-4
AC power port
AC power port
±0.5kV, ±1kV
±0.5kV, ±1kV
Surge
(differential
(differential
IEC 61000-4-5
mode)
mode)
0 % UT;
0 % UT;
0,5 cycle
0,5 cycle
At 0°, 45°, 90°,
At 0°, 45°, 90°,
135°, 180°, 225°
135°, 180°, 225°
, 270° and 315°
, 270° and 315°
Voltage dips,
0 % UT; 1 cycle
0 % UT; 1 cycle
short interrupti-
and
and
ons and voltage
70 % UT;
70 % UT;
variations on p-
25/30 cycles
25/30 cycles
ower supply in-
Single phase:
Single phase:
put lines
at 0°
at 0°
IEC 61000-4-11
0 % UT;
0 % UT;
250/300 cycle
250/300 cycle
Power frequency
(50/60 Hz)
30 A/m; 50Hz
30 A/m; 50Hz
magnetic field
or 60Hz
or 60Hz
IEC 61000-4-8
for 2-year from the date of purchase. If the
Electromagnetic environment
-guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels charactertic of a
typical location in a typical comme-
rcial or hospital environment.
31
Electromagnetic Compatibility Information
The device satisfies the EMC requirements of the international standard IEC
60601-1-2. The requirements are satisfied under the conditions described in the
table below. The device is an electrical medical product and is subject to special
precautionary measures with regard to EMC which must be published in the
instructions for use. Portable and mobile HF communications equipment can affect
the device. Use of the unit in conjunction with non-approved accessories can affect
the device negatively and alter the electromagnetic compatibility. The device
should not be used directly adjacent to or between other electrical equipment.
Table 1
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emissions test
Radiated emission CISPR 11
Conducted emission CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
33
Electromagnetic Compatibility Information
Table 2(continued)
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environm-
ent.
IEC 60601
IMMUNITY test
test level
Radiated RF EM
3V/m or 10 V/m
80MHz-2.7
fields
Ghz 80%AM at
IEC 61000-4-3
1kHz
3 V in 0.15
MHz- 80 MHz
Conducted
6 V in ISM
disturbances
and/or amateur
Induced by RF
radio bands
fields
between 0.15
IEC 61000-4-6
MHz and 80
MHz 80 %
AM at 1kHz
Electromagnetic environment
Compliance
-guidance
The device uses RF energy only for
its internal function. Therefore, its
emissions are very low and are not
Group 1,
likely to cause any interference in
class B.
nearby electronic equipment.
The device is suitable for use in all
establishments, including domestic
establishments and those directly
Group 1,
connected to the public low-voltage
class B.
power supply network that supplies
buildings used for domestic
purposes.
Class A
Complies
Compliance
Electromagnetic environment
level
-guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended seperation distance 80 MHz
3V/m or 10 V/m
to 800 MHz 800 MHz to 2.7 Ghz where P is
80MHz-2.7
the maximum output power rating of the
Ghz 80%AM at
transmitter in watts (W) according to the
1kHz
transmitter manufacturer and d is the
recommended separation distance in
metres (m). Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a should be
less than the compliance level in each
frequency range. Interference may occur in
the vicinity of equipment marked with the
following symbol:
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance
calculated from the equation applicable to
3 V in 0.15
the frequency of the transmitter.
MHz- 80 MHz
Recommended seperation distance 80 MHz
6 V in ISM
to 800 MHz 800 MHz to 2.7 Ghz where P is
and/or amateur
the maximum output power rating of the
transmitter in watts (W) according to the
radio bands
transm- itter manufacturer and d is the
between 0.15
recommended separation distance in
MHz and 80
metres (m). Field strengths from fixed RF
MHz 80 %
transmitters, as determined by an
AM at 1kHz
electromagnetic site survey, a should be less
than the compliance level in each frequency
range. Interference may occur in the
vicinity of equipment marked with the
following symbol:
32
34