Summary of Contents for JOYTECH DBP-6191

  • Page 1 INSTRUCTIONS FOR USE Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 2: Table Of Contents

    7.2 WHO Blood Pressure Classification Indicator ........14 7.3 Health Reminder ................15 BLOOD PRESSURE Q&A ................15 MAINTENANCE ..................16 10. SPECIFICATIONS ..................17 11. ELECTROMAGNETIC COMPATIBILITY INFORMATION ......19 12. ADDITIONAL NOTES ................23 WARRANTY CARD ................... 26 Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 3: Safety Notice

    1. SAFETY NOTICE Dear Customer, thank you for purchasing the DBP-6191 Blood Pressure Monitor. The unit has been constructed using reliable circuitry and durable materials. Used properly, this unit will provide yeas of satisfactory use. This device is intended for non-invasive measuring adults and adolescents over 12 years of age individual’s systolic, diastolic blood pressure and heart rate using the oscillometric...
  • Page 4 Monitoring the trends in your blood pressure taken at either the arm or the wrist is nevertheless useful and important. Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 5 6. People suffering from vascular constriction, liver disorders or diabetes, people with cardiac pacemakers or a weak pulse, and women who are pregnant should consult their physician before measuring their blood pressure themselves. Different values may be obtained due to their condition. 7.
  • Page 6 Advising operator that Instruction manual/ Booklet must be consulted. 30. Essential performance: Electrosurgery Refer 202.6.2.101 IEC 80601-2-30 interference recovery Limits of the error of Refer. 202.12.1.102 IEC 80601-2-30 the manometer Reproducibility of the BLOOD Refer 201.12.1.107 IEC 80601-2-30 PRESSURE DETERMINATION Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 7: Unit Illustration

    2. UNIT ILLUSTRATION INSTRUKCJA UŻYWANIA Automatyczny cyfrowy INSTRUKCJA INSTRUKCJA ciśnieniomierz naramienny DBP-6191 UŻYCIA UŻYCIA POLSKI Automatic Arm cuff 4 × batteries Carrying case Instruction Medical AC blood 22–42 cm type AAA 3 for use Adapter (DC 5.0 V, pressure (M/L)
  • Page 8: Display

    9. This blood pressure monitor is not recommended for people with severe arrhythmia. 10. Do not use this blood pressure monitor if the device is damaged. Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 9: Quick Start

    5. QUICK START 1. Install batteries. (See Figure A) 2. Insert cuff air plug into the behind side of monitor unit. (See Figure B) Fig. A Fig. B 3. Remove thick clothing from the arm area. 4. Rest for several minutes prior to testing. Sit down in a quiet place, preferably at a desk or table, with your arm resting on a firm surface and your feet flat on the floor.
  • Page 10: Unit Operation

    Press“START/STOP” button again to set the Time/Date mode. Set the year first by adjusting the “MEM” button. Press“START/STOP” button again to confirm cur- rent month. Continue setting the date, hour and minute in the same way. Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 11: Applying The Arm Cuff

    Every time the “START/STOP” button is pressed, it will lock in your selection and continue in succession (month, day, hour, minute, 12/24 hours) 3. Time Format setting Press “START/STOP”button again to set the time format setting mode. Set the time format by adjusting the“MEM” button. EU means European Time US means U.S Time.
  • Page 12: Testing

    25% slower or faster than the average rhythm detected while measuring systolic blood pressure and diastolic blood pressure. Consult your physician if the Irregular Heartbeat Symbol (( )) frequently appears with your test results. 10 | Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 13: Power Off

    6.7 Power Off The “START/STOP“ button can be pressed to turn off the unit in any mode. The unit can turn off the power itself about 3 minutes no operation in any mode. Safety Precaution: If pressure in arm cuff becomes too extreme while testing, press the “START/STOP“...
  • Page 14: Memory Deletion

    ”.will be displayed when the cuff is properly wound. When the cuff is too loose, „ ”will be displayed. At this time, please wear the cuff correctly and start measuring again. 12 | Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 15: Troubleshooting

    6.15 Troubleshooting Abnormal Cause analysis Processing method phenomenon The armband is tied too tight or too loose, Roll the armband correctly Or the arm strap is tied incorrectly; Move the arm during measurement or Stay quiet, keep your arm steady, and do Electronic sphygmomanometer not move the monitor Speaking, nervous or emotional during...
  • Page 16: Blood Pressure Information

    Example: fluctuation within a day (male, 35 years old) 7.2 WHO Blood Pressure Classification Indicator The DBP-6191 is equipped with a classification indicator based on established guidelines from the World Health Organization. The chart below (color coded on monitor unit) indicates test results.
  • Page 17: Health Reminder

    7.3 Health Reminder Hypertension is a dangerous disease that can affect the quality of life. It can lead to a lot of problems including heart failure, kidney failure, and cerebral hemorrhaging. By maintaining a healthy lifestyle and visiting your physician on a regular basis, hypertension and relative diseases are much easier to control when diagnosed in their early stages.
  • Page 18: Maintenance

    Avoid extreme temper- Remove batteries when atures. Do not expose not in operation for an unit directly under sun- extended period of time. shine. Do not disassemble product. 16 | Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 19: Specifications

    Service and maintenance require parts, repair, technical support will be provided. 10. SPECIFICATIONS Product Description Arm-type Fully Automatic Blood Pressure Monitor Model DBP-6191 Display LCD Digital Display Size: 47,6×63,8 mm (1.87" × 2.51") Measurement Method Oscillometric Method Systolic Pressure 60–260 mmHg...
  • Page 20 EN 1060-4 Non-invasive sphygmomanometers – Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers. ISO 81060-2, non-invasive sphygmomanometers – part 2: clinical validation of automated measurement type. 18 | Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 21: Electromagnetic Compatibility Information

    Correct Disposal of This Product (Waste Electrical & Electronic Equipment) This marking shown on the product indicates that it should not be disposed with other household waste at the end of its life. To prevent potential harm to the environment or to human health, please separate this product from other types of wastes and recycle it responsibly.
  • Page 22 30 A/m; 50Hz Power frequency magnetic fields should be Hz) magnetic field or 60Hz or 60Hz at levels charactertic of a typical location in a IEC 61000-4-8 typical commercial or hospital environment. 20 | Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 23 Table 2 Guidance and declaration of manufacturer-electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Radiated RF EM fields 3 V/m or 10 3 V/m or 10 Portable and mobile RF communications...
  • Page 24 DECT; LTE Band 1, 3, 4, 25; 1970 UMTS 2450 2400–2570 Bluetooth, Pulse WLAN, modulation 802.11 b/g/n, 217Hz RFID 2450, LTE Band 7 5240 5100–5800 WLAN 802.11 Pulse modulation 5500 217Hz 5785 22 | Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 25: Additional Notes

    Table 4 Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated therefore disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
  • Page 26 EXPECTED SERVICE LIFE of the ME EQUIPMENT 11. Do not place the blood pressure monitor and cuff at will. It will cause asphyxiation if the child swallows or twine around his neck. 24 | Arm-type Fully Automatic Blood Pressure Monitor DBP-6191...
  • Page 27 12. The cuff and the case of the blood pressure monitor have been tested for biocom- patibility and do not contain allergenic or harmful materials. Please stop using it if allergy occurs during use. 13. WARNING: Non-professionals do not modify the equipment, otherwise it will make the equipment measurement is not accurate.
  • Page 28: Warranty Card

    DATE OF SALE: ..........................WARRANTY TERMS 1. Diagnosis S.A. grants a warranty: • 2 years for blood pressure monitor DBP-6191 • 2 years for cuffs Hardware defects revealed during the warranty period will be rectified free of charge within 21 days.
  • Page 29 DOOR TO DOOR WARRANTY INFORMATION This warranty applies to products labelled “DOOR TO DOOR WARRANTY” 1. The warranty, under the terms and conditions set out in this card, is provided by Diagnosis S.A., with registered office under the address: Gen. W. Andersa 38A, 15-113 Białystok, Poland (hereinafter referred to as the “Warrantor”). 2.
  • Page 30 The product is in compliance with the requirements of MDD 93/42/EEC, „0197” is the identification number of notify body. Manufacturing address JOYTECH Healthcare Co., Ltd. No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China European Authorized Representative Shanghai International Holding Corp.