Invacare® AVIVA™ FX Series
6.10.6 Replacing Circuit Breaker
•
4 mm Allen key
•
Phillips screw driver
1. Make sure that circuit breaker is in the OFF position.
2. Remove side shroud and side panel shroud. Refer to
6.6.5 Replacing Side Shrouds, page 23
3. Loosen screws B securing battery cables A to circuit
breaker C.
4. Remove battery cables from each side of circuit breaker.
5. Pull tab D on back right side of circuit breaker and
rotate right side of circuit breaker away from bracket
(not shown).
6. Remove circuit breaker from bracket (not shown).
7. Remove foam E from top of circuit breaker C.
8. Install parts in reverse order.
If needed, loosen circuit breaker screws fully
for additional clearance to seat battery cables
during installation.
Tighten the hardware as listed below:
•
Screws securing battery cables to 2 Nm
•
Side panel shroud screws to 3 Nm
50
A B
C
B A
Fig. 6-53
Fig. 6-54
Fig. 6-55
1675507-A