Invacare® Fox™
12 Troubleshooting
12.1 Resetting the circuit breaker
WARNING!
– NEVER defeat or bypass the circuit breaker.
– ONLY replace with a circuit breaker of the same
rating.
Resetting the circuit breaker may be needed if the mobility device
does not turn on and the reset button has popped out about 6 mm.
1.
Switch off remote.
2.
To reset, press the circuit breaker button A located on the rear
of the battery box.
58
1577074-B