AND TALKING PREMIUM+ Instruction Manual page 29

Blood pressure monitor upper arm
Table of Contents

Advertisement

Available languages

Available languages

Technical Data
EMD Technical Data
Battery-operated or AC Adapter-operated Blood Pressure Monitor
Medical Electrical Equipment needs special precautions regarding EMD and
needs to be installed and put into service according to the EMD information
provided in the following.
Portable and mobile RF communication equipment (e.g. cell phones) can
affect Medical Electrical Equipment.
The use of accessories and cables other than those specified may result in
increased emissions or decreased immunity of the unit.
Table 1 - EMISSION Limits -
Phenomenon
Conducted and radiated RF EMISSION
CISPR 11
Harmonic distortion IEC 61000-3-2
Voltage fluctuations and flicker
IEC 61000-3-3
Table 2 - IMMUNITY TEST LEVELS : Enclosure Port -
Phenomenon
Electrostatic discharge IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
Rated power frequency magnetic fields
IEC 61000-4-8
Table 3 - IMMUNITY TEST LEVELS : Input a.c. power Port -
Phenomenon
Electrical fast transients / bursts
IEC 61000-4-4
Compliance
Group 1, Class B
Class A
Compliance
IMMUNITY TEST LEVELS
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
See table 4
30 A/m
50 Hz or 60 Hz
IMMUNITY TEST LEVELS
±2 kV
100 kHz repetition frequency
E-29

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Ua-1030t

Table of Contents