Ambulatory Blood Pressure Monitor TM2440-EX,
Instruction Manual, Corrections and Additions
Thank you for purchasing the Ambulatory Blood Pressure Monitor TM2440-EX. The Instruction Manual that
accompanies the product has been corrected and had content added. Please read it in conjunction with the
details shown below.
Instruction manual Page 22 6.1. Automatic Blood Pressure Measurement (A-BPM)
The following changes have been made.
The initial pressurization value is set to 180 mmHg at the
factory.
Instruction manual Page 37 8.2.3. Initial Pressurization Value
The following changes have been made.
The initial pressurization value is set to 180 mmHg at the
factory.
Instruction Manual Page 37 8.3. A-BPM Preset Programs
The following is added after
Wrong
Wrong
A-BPM can use only while the automated measurement can be performed.
"
The Start time and Operation time are initially set calculating from the
time at which the
automatic measurement is made, this needs to be set again.
The initial pressurization value for automatic blood pressure
measurement is AUTO.
In the case of AUTO, the appropriate pressurization value
is set automatically.
The initial pressurization value setting is AUTO
(appropriate pressurization value set automatically) when
the
mark is displayed and 180 mmHg when the
mark is hidden.
Addition
Note
mark
was
displayed. When the
1
Correct
Correct
next
1WMPD4004497
"

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TM2440-EX and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for AND TM2440-EX

  • Page 1 Instruction Manual, Corrections and Additions Thank you for purchasing the Ambulatory Blood Pressure Monitor TM2440-EX. The Instruction Manual that accompanies the product has been corrected and had content added. Please read it in conjunction with the details shown below. Instruction manual Page 22 6.1. Automatic Blood Pressure Measurement (A-BPM) The following changes have been made.
  • Page 2 Example 2 Automatic measurement " " Addition NB. Automatic blood pressure measurement will continue even if turned "OFF" and "ON" again during automatic measurement. The following notes were added to "Example 3 Automatic measurement" Addition NB. Automatic blood pressure measurement will not be performed once the Operation time has been exceeded even if turned "OFF"...

Table of Contents