Specifications - Kenwood CS-5130 Instruction Manual

40mhz dual trace readout oscilloscope
Hide thumbs Also See for CS-5130:
Table of Contents

Advertisement

SPECIFICATIONS
CS-5130
CS-5135
CRT
150 mm rectangular with internal graticule
Acceleration Voltage
12 kV
Display Area
8x 10 div (1 div=10 mm)
VERTICAL
AXIS (CH1 and CH2)
Sensitivity
|
1 mV/div to 5 V/div:
1 mV to 2 mV/div +5%, 5 mV/div to 5 V/div +3%
Attenuator
12 steps, 1 mV/div to 5 V/div in 1-2-5 sequence
Vernier control for fully adjustable sensitivity between steps
Input Impedance
1 MQ +2%,
approx. 30 pF
Frequency
| DC
DC to 40 MHz, within —3 dB (5 mV/div to 5 V/div)
Response
DC to 15 MHz, within —3 dB (1 mV/div to 2 mV/div)
| AC
5 Hz to 40 MHz, within —3 dB (5 mV/div to 5 V/div)
5 Hz to 15 MHz, within —3 dB (1 mV/div to 2 mV/div)
Rise Time
8.7 ns (5 mV/div to 5 V/div)
17.5 ns (1 mV/div & 2 mV/div)
Signal Delay Time
Adequate to identify leading edge
Crosstalk
— 40 dB or less (at 1 kHz)
Operating Modes | CH1
Single trace
CH2
Single trace
ALT
Two-waveform
display, alternately
CHOP
Two-waveform display, chopped
ADD
CH1 +(+CH2) added display
Chop Frequency
Approx. 300 kHz
Channel Polarity
Normal or inverted, channel 2 only inverted
/\, Maximum
Input Voltage
500 Vp-p or 250 V (DC +AC
peak)
HORIZONTAL AXIS Input thru CH2, x 10 MAG not included
Operating Modes
X-Y operation is selectable with HORIZ MODE
switch
CH1 : Y axis
CH2 : X axis
Same as vertical axis (CH2)
Sensitivity
Input Impedance
Same as vertical axis (CH2)
Frequency
| DC
DC to 1 MHz, within —3 dB
Response
| AC
5 Hz to 1 MHz, within —3 dB
X-Y Phase Difference
|
3° or less at 100 kHz
A Maximum
Input Voltage
Same as vertical axis (CH2)
SWEEP
Type
A
A sweep
ALT
A sweep (intensified for duration of B sweep) and B sweep (delayed sweep) alternating
B
Delayed sweep
X-Y
X-Y oscilloscope operation
Sweep Time
A
0.2 yps/div to 0.5 s/div +3%,
in 20 ranges, in 1-2-5 sequence
Vernier control for fully adjustable sweep time between steps
B
I
0.2 ys/div to 50 ms/div +3%, in 17 ranges, in 1-2-5 sequence
Sweep Magnification
x 10 (ten times)
+5% (+8% in 0.2 ys-to-0.5 us range)
Linearity
|
+3%
(+5% for x10 magnification)
Holdoff
Continuously variable from NORM to more than ten time (MAX)
Trace Separation
Shifts
B sweep trace continuously in vertical direction by 4 divisions orrnore with
respect to
A sweep
Delayed Sweep
Continuous delay (AFTER DELAY) & triggered delay (B TRIG' D: triggered by A trigger)
Delay Time
Continuous adjustable from 0.2 ys/div to 0.5 s/div

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cs-5135

Table of Contents