SINGLE: Single-sweep mode
RESET:
NOTE: Dual-trace single-sweep observation is
not possible if the vertical operation
MODE is set to ALT. Be sure to set the
MODE to CHOP.
Reset mode of single-sweep operation. When
reset, the switch returns to the SINGLE posi-
tion, with the READY LED @) lighting until com-
pletion of sweep.
READY Indicator
When reset in single-sweep operation, this lamp lights
and remains lit until the sweep operation is completed.
HOR!IZ MODE
(HORIZONTAL MODE) Switch
Used to select the horizontal display mode.
A:
ALT:
Only A sweep is operative, with the B sweep
dormant.
A sweep alternates with the B sweep. For this
mode of operation, the B sweep appears as an
intensified section on the A sweep.
Only delayed B sweep is operative.
Channel 1 becomes the Y-axis and channel 2
becomes the X-axis for X-Y operation. The set-
ting of the vertical MODE
and TRIG MODE
switches have no effect.
CURSORS Switch (CS-5130 only)
Cursor measurement mode select switch.
NOTE:
Cursor measurement is impossible if the HO-
RiZ MODE selector @ is set to ALT or B.
Fig. 4
OFF:
AV 1:
Cursor measurement
is not performed.
The
cursor, and cursour measurement mode and
cursor measurement value are not displayed
on the CRT.
Two horizontal cursor lines are displayed on
the CRT, and voltage difference
and voltage
ratio between them are displayed in the upper
right on
the
CRT
posterior
to
the
cursor
measurement mode display.
Setting the CH1 VARIABLE controller (3) to the
CAL position causes voltage diffarence meas-
urement, and a value calculated in accordance
with setting of the CH1 VOLTS/DEV control @
is displayed posterior to AV1.
Setting the CH1 VARIABLE contio Her G) to the
UNCAL position causes voltage ntio measure-
ment, and a value calculated assuming that
5-division as 100% is displayed posterior to
the RATIO.
When the A cursor is below the 4 REF cursor,
a negative value is displayed.
NOTE: Setting of the MODE selct switch @)
to the CH2 position causes 4 V2 mode
cursor measurement.