HP 70205A Installation And Verification Manual page 64

Graphics display
Hide thumbs Also See for 70205A:
Table of Contents

Advertisement

The utility has a 2-second timeout. Allow 2 seconds for the utility to respond to th
softkey. The utility responds by displaying one of the following messages. (Example m
are shown in Figure 5-6.)
Module not accepting data
This message indicates that there is an element at the
specified address, but it cannot communicate on the
HP-MSIB.
Communication complete
This message indicates the HP-MSIB of the element works
correctly. Note that the active light of the module should be
ON.
Module not present
This message indicates that there is no element at the
specified address.
This
utifity
may
inhibit
normal
system
A
operation.
Use
¢ to
exit.
ACTIVE
ON:
Row
address?
4
8
Column
address?
5
Module
not
present
ACTIVE
ON:
C
Row
address?
1
Column
address?
1C
Communication
complete
D
ACTIVE
OFF:
Row
address?
1
Column
address?
iC
Communication
complete
E
F
ENTER
Figure 5-6. Compieted HP-MSIB Communication
interpreting the Utility Results
If only one module cannot communicate on HP-MSIB, that module is faulty. Refer to the
appropriate service manual for troubleshooting and repair information.
If none of the elements on the HP-MSIB can communicate, either a module is disrupting the
entire HP-MSIB, the mainframe HP-MSIB circuits are faulty, or the display instrument has a
faulty HP-MSIB interface and the utility is giving erroneous results. Use the following steps to
further isolate the problem.
1. If an abnormal CRT screen is a symptom, refer to both the HP 70001A Mainframe Service
Manual and to the HP 70205A and HP 70206A Graphics Displays Service Manual for
further troubleshooting and repair information.
2. If flashing error lights are a symptom, the faulty element or mainframe can be isolated by
using the steps given below.
5-10
Troubleshooting