Orion's Belt Connection [ bELT ]
1. The default setting is Off.
2. On: Orion's Belt function is enabled.
3. Off: Orion's Belt function is disabled.
Belt Setting [ bELT.P ]
You may set the Orion Nano as the master unit or subordinating unit when establishing a connection.
Select [ 1 ] to set the Orion Nano as the master unit. Select [ 2 ] or [ 3 ] to set the Orion Nano as the
secondary unit.
Calibration [ CAL ]
The Orion Nano is a precision instrument, and regular calibration is recommended for frequent use.
Before performing a smart calibration, select Weighing Mode and ensure the weighing unit is in grams,
and follow the steps below:
1
Press the Dose
rapidly 5 times in
the weighing mode
to enter the user
calibration mode.
Note: You may use a 100 g or 500 g weight for calibration. To ensure accuracy, use Acaia calibration
weights or F1 or F2 Class calibration weights.
About [ About ]
Shows the current Orion Nano Firmware version.
To update to the latest version, please refer to "Updating the Orion Nano's Firmware"
2
While LED display
shows [ CAL ] and
then [ 100 ]. Place
a 100 g calibration
weight and wait for the
scale to execute the
calibration procedure.
3
The Display shows
[ -End- ] when the
calibration procedure
is completed.