Cause
Issue: not loud enough
Low volume
Low battery
Custom molds/domes
not inserted properly
Change in hearing
Custom molds/domes
blocked with ear wax
Plugged microphones
46
Possible remedy
Turn up volume; see hearing
care professional if problem
persists
Replace battery
See "Putting your hearing
aids on your ears. " Remove
and reinsert carefully
Consult your hearing care
professional
Clean custom molds/domes.
See "Care and maintenance. "
Consult your hearing care
professional
Consult your hearing care
professional
Cause
Issue: intermittent sound
Low battery
Dirty battery contact
Issue: two long beeps
Low battery
Issue: whistling
Custom molds/domes
not inserted properly
Hand/clothing near ear
Poorly fitting custom
molds/domes
Possible remedy
Replace battery
Consult your hearing care
professional
Replace battery
See "Putting your hearing
aids on your ears. " Remove
and reinsert carefully
Move hand/clothing away
from ear
Consult your hearing care
professional