Page 2
CAUTION This contains information for avoiding risks when handling the equipment that could result in minor or moderate injury to the operator or damage to property. DA00062341-002 1 / 11 IV-202F4 / IV-204F3...
──── To ensure safe use, be sure to observe the following safety precautions. NF Corporation shall not be held liable for damages that arise from a failure to observe these safety precautions or warnings or cautions in the instruction manual.
Page 4
This product does not include batteries and is designed in accordance with the RoHS Directive (EU). Dispose of the product in accordance with the laws and regulations of the area where you use it. DA00062341-002 3 / 11 IV-202F4 / IV-204F3...
Page 5
If the orientation of the connector is correct, excessive force is not required. Please be careful of the orientation of the connector. Securely connect the cable to the power connector of the main unit. DA00062341-002 4 / 11 IV-202F4 / IV-204F3...
Page 6
Securely connect the cable to the power connector of the main unit. This product does not operate on a single power supply (+30 V / GND). Be sure to use a dual power supply (+15 V / GND / −15 V). DA00062341-002 5 / 11 IV-202F4 / IV-204F3...
Page 7
If a large parallel capacitance is connected to the input of this product, overshooting, ringing, or oscillation may occur. Make the input wiring as short as possible. We recommend 50-Ω coaxial cables with lengths of approximately 10 cm and 40 cm for the IV-202F4 and IV-204F3, respectively.
Page 8
遮断周波数=1.5MHz 100k 100M 100k Frequency[Hz] Frequency[Hz] Figure 2-1 Frequency characteristic (IV-202F4) Figure 2-2 Frequency characteristic (IV-204F3) 2.2 Equivalent Input Noise Current Density The characteristics shown below are calculated by dividing the output noise voltage density by the current gain. 100p 70fA/√Hz typ.
Page 9
3.1 Overview The IV-202F4 and IV-204F3 are compact current-to-voltage conversion amplifiers with low noise characteristics. Wide bandwidths of up to 10 MHz and 1 MHz are realized for the IV-202F4 and IV-204F3, respectively. These products stably operate for up to an input parallel capacitance of approximately 20 pF (IV-202F4) or 50 pF (IV-204F3) without adjustment.
Page 10
To compensate for input parallel capacitance, make adjustments by turning the adjuster clockwise. However, the upper cut-off frequency becomes lower as you turn the adjuster clockwise. * 3-1: approximately 20 pF or higher (IV-202F4). Approximately 50 pF or higher (IV-204F3). DA00062341-002...
Page 11
* For the SMB connector used as the input and output connector, use a connector with a diameter of φ6.2 mm or smaller. * The external view shown above is that of the IV-202F4. The external dimensions of the IV-204F3 are the same; only the model name is different.
Page 13
NF or an agent designated by NF. The purchaser shall prepay all shipping charges, duties, and taxes for the product to either NF or an agent from another country, and the shipping charge for returning the product to the purchaser shall be paid by NF.