PERFORMANCE TEST CARD
90 DAY LIMITS
Hewlett-Packard Model 3457A
Digital
Mutimeter
Serial Number
Test Performed by _
Date
Reference Temperature_
AC VOLTAGE TEST (Cont'd)
Step#
Input
to
Set-Up and
High
Reading
Low
Test
Test
34578
Configuration
Limit
Limit
Pass
Fail
FREQUENCY
RESPONSE
TEST
(Cont'd)
37
Set
the
3457A
to
AC
Slow
Response
(ACBAND
< 400)
38
3V, 100Hz
3 VAC Range
3.008800 Vv
2
991200
V
7
a
39-3, 45Hz
3 VAC Range
3.026800 v
2.973200
v
40
3v,20Hz
3 VAC Range
3.026800 V
_
2.973200 v
41
300mV,100Hz
300 mVAC Range
300.8800 mv
:
299.1200
mv
-
=
42
300mV,45Hz
300 mVAC Range
302.6800 mv
297.3200
mv
43
300mv,20Hz
300 mVAC Range
302.6800 mv
297.3200
mv
-
FREQUENCY
TEST
Step#
Input to
Set-Up and
High
Reading
Low
Test
Test
3457A
Configuration
Limit
Limit
Pass
Fail
1
W,20Hz
3 VAC Range
20.01000 Hz
19.99000 Hz
2
VV, IMHz
3 VAC
Range
1.000100
MHz
0.999900