HP 3457A Service Manual page 109

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

PERFORMANCE
TEST CARD
1 YEAR LIMITS
Hewlett-Packard Model 3457A
Digital Mutimeter
Serial Number_
AC VOLTAGE TEST (Cont'd)
Test Performed by
Date
Reference Temperature_
Step#
Input
to
Set-Up
and
High
Reading
Low
Test
Test
34570
Configuration
Limit
Limit
Pass
Fail
FREQUENCY
RESPONSE
TEST
(Cont'd)
38
Set
the
3457A
to
AC
Slow
Response
(ACBAND
< 400)
39
-3V, 100Hz
3 VAC Range
3.008820 v
2.991180 Vv
40
-3V,45Hz
3 VAC Range
3.019320 v
2.980680 V
41
-3V,20Hz
3 VAC Range
3.019320 Vv
=
2.980680 V
42
300mv,100Hz
300 mVAC Range
300.8820 mv
299.1180 mv
43
300mv,45Hz
300 mVAC Range
301.9320 mv
298.0680 mv
44
300mv,20Hz
300 mVAC Range
301.9320 mv
298.0680 mv
>
FREQUENCY
TEST
Step#
Input to
Set-Up and
High
Reading
Low
Test
Test
3457A
Configuration
Limit
Limit
Pass
Fail
1
1, 20Hz
3 VAC Range
20.01000 Hz
_19.99000 Hz
2
WV, 1MHz
3 VAC Range
1.000100 MHz
0.999900 MHz