HP 3457A Service Manual page 66

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

B457A
Performance Tests
2-Wire Ohms Performance Tests Cont'd
4-32. 2-Wire Ohms Function - Gain Test
4-33. Equipment Required. Resistance Standards of 30 Ohms (+ .2%), 300 Ohms ( .02%), 3 Kohms (+ .003%)
30 Kohms (+ 001%), 300 Kohms (+ 001%) 3 Mohms
(+ .001%), and 30 Mohms
(+ .009%) are required for
this procedure.
LSet the HP 3457A to the 2-wire ohms function (OHM) and the number of digits displayed (DIGITS
DISP) to six.
2. Connect the Resistance Standard to the HP 3457A front panel HI and LO input terminals. (The
connecting wires should be as short as possible to reduce lead resistance).
3. Use the Test Record to record the 30 Ohm, 300 Ohm, 3 Kohm, 30 Kohm, 300 Kohm, 3 Mohm and
30 Mohm readings.
4. If any of the readings are beyond the limits specified in Table 4-8 and on the Test Record, the in-
strument should be calibrated or repaired. Refer to Section V for calibration procedures.
5. Remove the Resistance standard from the HP 3457A front panel input terminals.
34570
3457A
| 34578
90 day limits
1 year Limits
Input
Range
| Set Up
High
Low
High
Low
30 ohm
30 ohm
OHM
30.20530
ohm
29.79470
ohm
30.20560
ohm
29.7954400hm
300 ohm | 300 ohm | OHM
300.2170 ohm
299.7830 ohm
300.2200 ohm
299.7800 ohm
3 Kohm
3 Kohm|
OHM
3.000312 Kohm
2.999688 Kohm
3.000357 Kohm
2.999643 Kohm
30 Kohm}|
30 Kohm|
OKM
30.00132 Kohm
29.99868 Kohm
30.00177 Kohm
29.99823 Kohm
300 Kohm|
300 Kohm|
OKM
300.0130 Kohm
299.9870 Kohm
300.0160 Kohm
299.9840 Kohm
3 Mohm
3 Mohm|
OHM
3.000179 Mohm
2.999821 Mohm
3.000209 Mohm
2.999791 Mohm
30 Mohm}|
30 Mohm|
OKM
30.00833 Mohm
29.99167 Mohm
30.01283 Mohm
29.98717 Mohm
Table 4-8. 2-Wire Ohms Gain Test Limits
4-Wire Ohms Performance Tests
@ SENSE
INPUT
tevin,
a vie.
HATH
RECALL
STORE)
_ sno
AEAN
oR SOEV
RNSAMP | ERROR
CLEAR
CJ C4) Ge)
TAST
cont
FuLteR | _ entry
Figure 4-6. 4-Wire Ohms Offset Performance Test Connections

Advertisement

Table of Contents
loading

Table of Contents