General
Information
Table 1-1. Specifications
@
TRUE RMS AC CURRENT (Cont*d)
90 DAY: Tcal + 5°C (ACI Function) After 1 hour warm-up
1 A range. (1 amp maximum input)
FREQUENCY
* PERCENT
NUMBER OF COUNTS
OF
>= 1 PLC
0.1 PLC
-005 PLC
-0005 PLC
READING
|6 1/2 digits|5
1/2 digits|4
1/2 digits|3
1/2 digits
20Hz-45Hz
95
2800
290
32
46H2- 100H2
0.4
2800
290
32
101Hz-20KHz
0.35
2800
290
32
**
GOOHZ-20KHz
0.35
2800
750
80
12
* Specifications
apply
within
one
week
of
AC
auto-cal
(ACAL
2)
for
stable
temperature
conditions.
If AC
auto-cal
is not
used,
add
0.6
to
the
percent
of
reading
figures.
For 1 year specifications, add .08% to the 90 day Percent of Reading figures.
**
Using
AC
fast
filter
(ACBAND
400)
for
frequencies
above
400Hz.
TEMPERATURE
COEFFICIENT
- AC COUPLED: ALL Ranges.
FREQUENCY
PERCENT
NUMBER OF COUNTS
READING
|6 1/2 digits|5
1/2 digits [4 1/2 digits [3 1/2 digits
&
20H2-100KH: | 0.017
|
150
15
|
1.5
|
0.15
MEASUREMENT ACCURACY - DC COUPLED:
+(% of reading + number of counts)
After
two
hour
warm-up.
Sine-wave inputs > 10% of range.
AC slow filter ON (ACBAND 20).
Auto-zero ON.
For pre-programmed settling times, error is 0.1% of input current step.
HOUR:
Tcal + 1°C (ACDCI
Function) After 2 hour warm-up
Accuracy relative to calibration std
30
mA
and
300
mA
ranges.
FREQUENCY
PERCENT
NUMBER OF COUNTS
OF
>= 1 PLC
0.1 PLC
-005 PLC
-0005 PLC
READING
|6 1/2 digits]5
1/2 digits|4
1/2 digits]3
1/2 digits
20H2-45Hz
1.5
16000
1600
165
20
46Hz-100Hz
0.35
16000
1600
165
20
101Hz-20KHz
0.25
16000
1600
165
20
**
GOOHZ-20KHz
0.6
16000
3750
375
42
21KHz-100KHz
0.9
17500
1750
180
22