Test 8 checks the AC attenuator flatness on the 3V AC range.
A 5V test signal is generated at A2U123
pin 13 and routed through A2U402 to the input of the A3 assembly.
A measurement
is made on the A3
assembly output signal and the Main Controller (A1U501) checks for convergence by checking to DAC
(A2U217) count limits of 512 +462.
This limit assures being within the calibration range of the DAC.
Figure 8-10 illustrates some of the circuitry used for this test.
On error, if the AUXERR? command
is executed, the number 256 will be displayed on the control panel
if their are no other errors (See Table 8-14). This indicates an AC attenuator flatness error that could
possibly be corrected by performing an internal mechanical adjustment on A3C332 and/or by perform-
ing the ACAL
command.
To adjust
A3C332,
execute
DIAGNOSTIC
8 and adjust
until the control
panel display indicates "PASSED" (refer to the AC Converter
Frequency
Response
Adjustment
in sec-
tion 5). If adjustment does not correct the error, executing DIAGNOSTIC 8 puts this test into a loop-
ing condition,
This is useful while troubleshooting as signals can be monitored.
Test 9 checks the ohms precharge circuitry on assembly A2/1.
The unit is set to the 300 volt DC range
with
I power
line cycle integration
time and the latest auto calibration
constant
is sent to the A/D
Controller
(A2U501)
from
the Main
Controller
(AI1U501).
The
input
hybrid
(A2U101)
precharge
amplifier
is selected
by input
hybrid switch
PRE
as the input
to input amplifier
QIIL
and
a zero
measurement
is made.
Next, a measurement
is made through input hybrid switch
MC.
The difference
between
the two readings is computed
and the Main Controller
performs error checking to limits of
512 +512 DAC counts.
This limit assures being within the calibration range of the DAC.
Figure 8-11
il-
lustrates some of the circuitry used for this test.
On error, if the AUXERR?
command
is executed, the number
512 is displayed on the control
panel if
their are no other errors (See Table 8-14),
This indicates an ohms precharge failure that could possibly
be corrected by performing the ACAL command,
8-78.
Self Test Changes for New Main Controller (03457-66511)
Due to a new
main
controller
board
change, minor changes
have been
made
to self test execution.
Auxiliary error codes 2048, 4096 and 8192 are no longer used.
Designations for components used in the
description of self tests 0, | and 2 have changed.
The sections that follow
include an updated table of
error codes and an updated description of tests 0, 1, and 2.
8-79. Updated Power-on Self Tests
At instrument power-on, an abbreviated form of the complete Self Test procedure is executed to check
the fundamental
digital hardware
kernel and the communication
link between
the Main Controller
(ALLU601) and A/D Controller (A2U501).
This consists of executing Self Tests 0, | and 2 in that order.
These 3 tests are discussed in detail in the following Self Test paragraphs.
Upon satisfactory completion
of the power-on
test routine, the instrument
momentarily
displays the
HP-IB address it is currently set to (the address is set to decimal 22 at the factory), emits a single "beep"
and sets itself to the predefined power-on
state.
The front panel indications of this state are; the DC
Voltage function, Auto-Range and automatic trigger are selected.
For a complete list of power-on con-
ditions, refer to Section III, Table 3-1 in this manual.