2.1 Overview

This section discusses the considerations involved with deploying a Honeywell ST 800 SmartLine pressure
transmitter in a process system. The following areas are covered:
Safety
Input and output data
Reliability
Environmental limits
Installation considerations
Operation and maintenance
Repair and replacement

2.1.1 Accuracy

The ST 800 SmartLine pressure transmitter measures the gauge, differential, or absolute pressure of a
process and reports the measurement to a receiving device.

2.1.2 Diagnostic Messages

Transmitter standard diagnostics are reported in the two basic categories listed in
detected as critical diagnostics drive the analog output to the programmed burnout level. Problems
detected as non-critical diagnostics may affect performance without driving the analog output to the
programmed burnout level. Informational messages (not listed in
or setting conditions. The messages listed in
associated with HART and DE protocols. HART and DE diagnostic messages are listed and described in
the ST 800 SmartLine Pressure Transmitter HART/DE Option User Manual, #34-ST- 25-38.
Critical Diagnostics (Failure Conditions)
Electronic Module DAC Failure
Meter Body NVM Corrupt
Config Data Corrupt
Electronic Module Diag Failure
Meter Body Critical Failure
Sensor Comm Timeout
5

2 Application Design

Table 3

Table 3 - ST 800 Standard Diagnostics Messages

Non-Critical Diagnostics (Warning Conditions)
No DAC Compensation
No Factory Calibration
PV Out of Range
Fixed Current Mode
Sensor Over Temperature
Meter Body Excess Correct
Local Display
Low Supply Voltage
No DAC Calibration
Tamper Alarm
Meter Body Unreliable Comm
ST 800 Pressure Transmitter User's Manual
Table
3) report various transmitter status
are specific to the transmitter, exclusive of those
Loop Current Noise
AO Out of Range
URV Set Err. Span Config Button
LRV Set Err. Zero Config Button
PILD Blockage Detected
Table 3.