Honeywell SmartLine ST 800 User Manual page 109

Pressure transmitters
Table of Contents

Advertisement

Meter Body Comm fault
Communications between the
meter body and the electronics
module has failed.
ST 800 Pressure Transmitter User's Manual
99
This could be the result of a failure
on either of these modules or the
cable that connects them.
Use a HART, DE, or FF
communicator to read the detailed
status information from the
transmitter. Refer to the
appropriate communicator manual
to get more information about the
possible causes of the failure.
Check the ribbon cable that
connects the meter body to the
electronics module. Make sure that
the cable is securely plugged into
the electronics module. Make sure
that all pins are plugged into the
connector (i.e., make sure that the
connector is not offset in a way that
leaves some pins unconnected).
Cycle power to the transmitter.
If the problem continues to occur
replace the electronics module.
If this does not fix the problem,
replace the meter body.
Revision 15

Advertisement