- Check and make sure that the drain valve is closed (ro-
tate clockwise to the stop position, see Fig� 10)�
Note: Please note information in section "Commissioning
and operating"�
- Connect the power plug and turn on the device with
power switch�
- The low level icon (
WiCo� Meanwhile, the bottom LED segment of fluid
level indicator lights up in red color to indicate the low
fluid level�
- Open the filling opening lid and remove the stopper
from the filling opening� Then, fill fluid to the bath�
Fig. 10
Note: Pay attention to the fluid level information�
Fluid level information on the WiCo display:
— Low level
— High level
Fluid level information on the fluid level indicator:
Red (High Level)
Red (Low Level)
Fig. 11
) appears on the display of the
Drain valve
Drain port
- To drain the fluid from the bath, connect a hose to the
drain port and turn the drain valve in counter clockwise
direction with a straight screwdriver�
Warning: Do not empty the bath while it is still cold,
there is a risk of freezing�