7. To calibrate the device, scroll to Set Span** and change the span
gas calibration value using this procedure;
Using the right arrow, move the cursor to xxx  PPM (span 
a.
value field).  Press Enter to select the field (it is editable 
when flashing). 
Use the up or down arrows to increase or decrease the val‐
b.
ue, press Enter to validate the new value. 
Move the cursor back to Set Span and press Enter to start 
c.
the calibration.  
 
 
 
The device Span is being calibrated.  The screen will display the de‐
vice's status as "In calib..." until the calibration is complete. 
 
**When selecting Set Span, make sure that the device has been sup‐
plied with the appropriate calibration gas before and during the cali‐
bration process.   
 
 