Internal Adjustments - Tektronix DM2510 Service Procedure

Hide thumbs Also See for DM2510:
Table of Contents

Advertisement

TEST EQUIPMENT REQUIREMENTS
Description
Performance Specifications
Calibration
DC Voltage : Range ,
to
Generator
1000V , accuracy : ± 0.005 % .
AC Voltage : Range
to 1000V,
accuracy: ± 0.1 % , 20 Hz
20kHZ
AC Power
Amplifier
Digital
Multimeter
Shorting
plug
Temperature
Table 1
DM2510/ DM2510G
Page 4 of 17
AC Current : Range
to 2A ,
accuracy: ± 0.1 % , 20Hz- 10kHz
DC Current: Range
to 2A,
accuracy: ± 0.01 %
Resistance:
to 20 MQ ,
accuracy: ± 0.01 %
DC Volts accuracy: ≥ 0.5%
Temperature bath :
° C +/- 0.1 ° C ,
100 ° C +/- 0.1 ° C
Used for
DC Volts
AC Volts
Set DC offset
voltage
Example
Fluke Model
5700A
Fluke Model
5100B
Adjustment Procedure
May 6 , 1994
Fluke 5215A
Tektronix
DM251
Tektronix P/N
134-0012-00
Temperature
RTE 4 Neslab
probe
Internal Adjustments
The entire calibration procedure may be performed without having to make
any internal adjustments if AC compensation has been verified .
Calibration
can be performed by either of two methods :
(1 ) from the front panel or (2 ) over the IEEE-488 bus (GPIB ).
To make internal adjustments , you must remove the top of the cabinet to access
the component side of the main board .
All other calibrations need to be performed with the top of the cabinet closed .
Revision A
¡