Tektronix DM2510 Service Procedure page 17

Hide thumbs Also See for DM2510:
Table of Contents

Advertisement

a .
b.
C.
d .
e .
f.
g .
h .
i .
j.
k.
1.
8
Set the DMM to Temperature measurement function by pressing
the PROG and TEMP key.
DM2510/ DM2510G
Page 17 of 17
Connect the P6602 Probe to the DMM LOW and V / N /TEMP
jacks.
Put the tip
of the
temperature
probe
about
1
inch
into the
temperature bath of
° C (± 0.1 ° C).
Wait for the measurement to stabilize .
Press the DOWN and UP of Autoranging keys simultaneously . The
DMM will then calibrate the offset.
After writing the new settings in EEPROM the readout will show :
" doNE "
A few
second later , the readout returns to 0.0° C. The offset
calibration of the temperature function is now completed .
Put the tip of the temperature
probe about
1
inch into the
temperature bath of 100 ° C (±0.1 ° C).
Wait for the measurement to stabilize .
Press the DOWN and UP of Autoranging keys simultaneously . The
DMM will then calibrate the gain .
After completion of the calibration the readout will show : " doNE " ,
and the positive gain is now successfully written in EEPROM .
As a result, the DMM display should now read +100.0 ° C ± 1
count.
When all previously mentioned calibration steps are successfully performed ,
re- calibration of the DM2510G/ DM2510 is completed .
Adjustment Procedure
May 6 , 1994
Revision A