Primary And Secondary Parameters; Tertiary Parameter; Drive Level Dependency Test - HP E4915A User Manual

Crystal impedance meter. crystal impedance/lcr meter
Table of Contents

Advertisement

ACI
AFr
Drive Level
Figure 5-5. Drive Level Dependency Test
Primary and Secondary Parameters
Steps to specify the primary and secondary parameters are described
later in "Handler Interface (Built-in Comparator Function; HP E4916A
Only)".
Tertiary Parameter
You can specify a limit for both Fr and Cl at each drive level. The
HP E4915A/E4916A checks the measured value at each level against
the limit as this level changes.
1. Sequentially press
((blue)]
and
f'm/k (Comprtrj).
2- Use
or (<^t) to display A F_H:n.nn Hz
•L Press A F_H:n.nn Hz to enter the upper limit for frequency.
Press f<^T) to display AZ_H:n.nn Hz .
Press AZ_H:n.nn Hz to enter the upper limit for impedance.
6- Use fljjO or f^in to display Lm A F:0FF and Lm A Z:0FF.
7- Press Lm A F:0FF to toggle it Lm A F:0N .
8- Press Lm A Z:0FF to toggle it Lm A Z:0N .
9- Use (2hD or (^0 to display COMP:OFF .
i0- Press COMP:OFF to toggle it for turning ON the comparator
function.
Applications
5-17