Symbols used in the Digital Display MAN: manual measuring range selection is active MIN/MAX value storage HOLD: display memory, “freeze” measured value Digital display with decimal point and polarity display Diode measurement selected appears when acoustic signal is active Selected type of current Unit of measure Unit of measure C / F for temperature measurement max.
Table of Contents Page Safety Features and Precautions You have selected an instrument which provides you with a Safety Features and Precautions .........3 high level of safety. This instrument fulfills the requirements of the applicable Initial Start-Up ..............5 European and national EC guidelines. We confirm this with Selecting Measuring Functions and Measuring Ranges ..5 the CE marking.
Page 4
• Be absolutely certain that the measuring ranges are not Meanings of symbols on the instrument: overloaded beyond their allowable capacities. Limit values Warning concerning a source of danger can be found in the “Measuring Ranges” table in chapter (attention: observe documentation) 12 “Characteristic Values”.
Initial Start-Up The instrument is automatically switched to: (6039 digits + 1 digit) – The next highest range at typically Battery (560 digits – 1 digit) – The next lowest range at typically Your instrument is supplied with two 1,5 V AA size batteries in accordance with IEC LR 6, and is ready for operation.
3.4 Relative Measurement REL Measured Value Memory “HOLD” A reference value for relative measurements can be stored to By pressing the HOLD key, the currently displayed memory with the keys MAN / AUTO and HOLD. measurement value can be “frozen” in the display, and “Hold” is simultaneously indicated.
Voltage Measurement ➭ Depending upon the voltage to be measured, set the rotary selector switch to V or V 20.00 ➭ TRMS Connect the measurement cables as shown. The “” connector jack should be grounded. Note! The 600 mV measuring range can only be selected 6 V ...
Voltage Measurements at Above 600 V Voltages of greater than 600 V can be measured with a high- voltage probe, e.g. the HV3 or the HV30 from GMC-I 0.000 Messtechnik GmbH. It is absolutely essential to ground the AC ~ ...6 A bonding terminal.
Page 9
connector jacks.The following symbol appears at the digital display in this case: • If a fuse blows, eliminate the cause of overload before 00.00 placing the instrument back into service! AC ~ ... 60 mA • Refer to chapter 13 “Maintenance”, regarding fuse TRMS replacement.
Measuring Alternating Current with (Clip-On) Current Resistance Measurement Transformers ➭ Make sure that the device under test is voltage-free. Interference voltages distort measurement results! 8.1.1 Transformer Output mA / A Set the rotary selector switch to . ➭ ➭ Connect the device under test as shown. Attention! If current transformers are operated without being ...
Continuity and Diode Testing ➭ Make sure that the device under test is voltage-free. Interference voltages distort measurement results! FUNC 030.0 ➭ Set the rotary selector switch to ➭ Connect the device under test as shown. brief 0.000 0 V ! acoustic R <...
Temperature Measurement The multimeter provides for the measurement of temperatures within a range of – 50°C to + 400°C with the help of a type K temperature sensor. ➭ Set the rotary selector switch to „°C“. ➭ Connect the sensor to the two accessible jacks. The device indicates the measured temperature in C at the digital display.
Maintenance 13.2 Fuses Attention! If one of the fuses blows, this condition is displayed at the LCD as soon as a measured quantity with a voltage of greater Disconnect the instrument from the measuring circuit before opening to replace batteries or fuses! than 4 V is applied to the corresponding connector jacks.
If you use batteries or rechargeable batteries in your instrument Accessories or accessories which no longer function properly, they must be duly disposed of in compliance with the applicable 15.1 General national regulations. The extensive accessories available for our measuring Batteries or rechargeable batteries may contain harmful instruments are checked for compliance with currently valid substances or heavy metal such as lead (PB), cadmium (CD)
Repair and Replacement Parts Service, DAkkS Calibration Certificate Reprints (upon request) Calibration Center* and Rental Instrument Service If you order a DAkkS calibration certificate reprint for your instrument, please provide us with the reference numbers If required please contact: indicated in the upper and lower most fields of the calibration seal.