Removing an SpO2 sensor

Removing sensor model FLEXMAX
Press the sides of the sensor together to open the sensor.
Remove the sensor from the finger.
Disconnect the sensor cable from the interface cable.
Remove the interface cable plug from the DefiMonitor XD.
Clean and disinfect the sensor. Please also observe the operating instructions for the sensor.
✓ Store the sensor until it is next used.
The expected shelf life of the sensor is 1 year.
7.8

Alarms and alarm limits

Extremely set alarm limits
No alarm in case of a patient's critical state.
The patient's critical state is not recognised.
➢ Select purposeful alarm thresholds.
The DefiMonitor XD has an intelligent alarm system. High, medium and low priority alarms exist and
also informative notifications and error notifications. If several alarms are triggered simultaneously, all
the alarms are displayed on the monitor. However, only the heart alarm signal (heart frequency and
VF/VT) is played back. Alarm signalling takes place for all alarms in less than 10 seconds.
Physiological alarms
Physiological alarms are triggered by patient monitoring. Different alarm signals exist for alarms which
are triggered via heart monitoring (heart frequency and VF/VT) and for alarms which are triggered via
the SpO2 monitoring.
Technical alarms
Technical alarms are triggered by the monitoring of the DefiMonitor XD, e.g. battery charge level low.
These alarms are signalised acoustically and optically.
Informative notifications and error notifications
Informative notifications and error notifications are signalised acoustically and optically. If your Defi-
Monitor XD has SpO2, the notifications will be displayed instead of the SpO2 curve. The notifications
will be shown until the cause is resolved.

7.8.1 Alarms with high priority

A high-priority alarm indicates that it is necessary for the user to react immediately.
Parameter
ECG, SpO2
(physiological alarm)
DefiMonitor XD | 24050 | GB | I
WARNING
Status
Upper alarm limit for
heart rate exceeded
Fallen below lower alarm
limit for heart rate
Notification
Blinking heart rate value
Blinking alarm bell
Blinking alarm LED
Alarm signal
Blinking heart rate value
Blinking crossed-out
alarm bell
Blinking heart rate value
Blinking alarm bell
Mode
ECG MON
MAN
SYNC
PACE
AED
ECG MON
MAN
45 / 108