Manual for device software updates
ARM 3.14 → ARM 3.16
___________________________________________________________________
Manual for device software updates
ARM 3.14 → ARM 3.16
The update may only be carried out on the following devices:
Name of the device
PRIMEDIC HeartSave AS
PRIMEDIC HeartSave PAD
PRIMEDIC HeartSave AED
PRIMEDIC HeartSave AED-M
Attention: Please read the instructions in advance as well as the information in Section '8 - Information'
carefully and in-full.
If you have any questions, you are welcome to use the following contact options to ask us.
Term definitions: In the following instructions, for all devices concerned in the HeartSave series
(HeartSave PAD / HeartSave AED / HeartSave AS / HeartSave AED-M) the term 'AED' is used consistently
and representatively.
The memory card which was included in the original scope of delivery of your device is described as the
'SaveCard'.
'Battery 3', 'Battery 6' and 'AkuPak LITE' are jointly referred to with the term 'Power module'.
Attention: Please check the status display of your AED (Fig. A) as fast as possible.
If it shows the depiction in Fig. (D), go to Section '1 - Remove the power module'.
However, if the status display shows one of the two depictions Fig. (B) or Fig. (C), please remove the
power module as described in '1 - Remove the power module', wait a few seconds and then re-insert it
into the AED as described in '4.2 - Re-inserting the power module'. As a result, your AED carries out
an automatic self-test (see Section '6.3 - Automatic self-test' and following).
If the status display shows the depiction in Fig. (D), you can start the update as described from
'1 - Preparations'.
If one of the two depictions Fig. (B) or Fig. (C) remains visible in the status display of your AED,
please contact the Metrax GmbH technical service.
In this case, your AED must be sent to Metrax.
Metrax technical service
Telephone:
E-mail:
Service@primedic.com
Anleitung Update ARM 3.16 - Download v2.2 EN.docx
Serial numbers affected
73943100001 - 73943100367
73944100000 - 73944100594
73945100000 - 73945100928
73946100000 - 73946100569
+49 741 257 275
Released on: 06.04.2022
Page 1 of 7

Advertisement

loading
Need help?

Need help?

Do you have a question about the HeartSave AS and is the answer not in the manual?

Questions and answers

Summary of Contents for Primedic HeartSave AS

  • Page 1 If you have any questions, you are welcome to use the following contact options to ask us. Term definitions: In the following instructions, for all devices concerned in the HeartSave series (HeartSave PAD / HeartSave AED / HeartSave AS / HeartSave AED-M) the term 'AED' is used consistently and representatively.
  • Page 2: Removing The Power Module

    Manual for device software updates ARM 3.14 → ARM 3.16 ___________________________________________________________________ Removing the power module • Lay the AED on its reverse side. • Press the release mechanism button (1) to the right until the power module is released and pro- trudes from the slot.
  • Page 3 Update package for the HeartSave AED, HeartSave AS, HeartSave AED-M If you carry out the update on a HeartSave AED, HeartSave AS, or HeartSave AED-M, the following folders and files should then be on the SaveCard: aedpadb.ulf bootloader.bin Remove the SaveCard safely Use the ‘Eject removable media' function before you remove the SaveCard from the card reader device.
  • Page 4 Manual for device software updates ARM 3.14 → ARM 3.16 ___________________________________________________________________ Implementing the update Insertion of the SaveCard with the software update it incorporates and the power module, the AED starts and automatically initiates the update. • During the update, a regular clicking sound can be heard from the AED loudspeaker and the depiction in the status display changes.
  • Page 5 Manual for device software updates ARM 3.14 → ARM 3.16 ___________________________________________________________________ Restoring readiness for use of the AED Removing the update package from the SaveCard Now, carry out the following steps in sequence: • Remove the power module, as described for '1 - Removal of the power module'. •...
  • Page 6 Manual for device software updates ARM 3.14 → ARM 3.16 ___________________________________________________________________ Switching the AED off When switched off, your AED returns to standby mode. • Close the cover of your AED as shown in illustration (12). • The verbal instructions become silent shortly afterwards. •...
  • Page 7: Additional Information

    After successfully completed updating, check the verbal instructions on your AED. If required, reset the starting language by pressing the language selection button (for the HeartSave AED and HeartSave AS) or via the menu (for the HeartSave AED-M). Additional information...

This manual is also suitable for:

Heartsave padHeartsave aedHeartsave aed-m739431000017394310036773944100000 ... Show all