Olympus EVIS LUCERA GIF N260 Operation Manual

Gastrointestinal videoscope, colonovideoscope

Advertisement

Refer to the endoscope's companion manual, the "REPROCESSING MANUAL" with
your endoscope model listed on the cover, for reprocessing information.
INSTRUCTIONS
EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE
OLYMPUS GIF TYPE N260
OLYMPUS GIF TYPE XP260
OLYMPUS GIF TYPE PQ260
OLYMPUS GIF TYPE XQ260
OLYMPUS GIF TYPE Q260
OLYMPUS GIF TYPE H260
OLYMPUS GIF TYPE H260Z
EVIS LUCERA COLONOVIDEOSCOPE
OLYMPUS CF TYPE Q260AL/I
OLYMPUS CF TYPE Q260DL/I
OLYMPUS CF TYPE H260AL/I
OLYMPUS CF TYPE H260AZL/I
OLYMPUS CF TYPE H260DL/I
OLYMPUS PCF TYPE PQ260L/I
OLYMPUS PCF TYPE Q260AL/I
OLYMPUS PCF TYPE Q260AZI
OLYMPUS PCF TYPE Q260JL/I

Advertisement

Table of Contents
loading

Summary of Contents for Olympus EVIS LUCERA GIF N260

  • Page 1 INSTRUCTIONS EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE N260 OLYMPUS GIF TYPE XP260 OLYMPUS GIF TYPE PQ260 OLYMPUS GIF TYPE XQ260 OLYMPUS GIF TYPE Q260 OLYMPUS GIF TYPE H260 OLYMPUS GIF TYPE H260Z EVIS LUCERA COLONOVIDEOSCOPE OLYMPUS CF TYPE Q260AL/I...
  • Page 3: Table Of Contents

    Contents Contents Symbols..................Important Information — Please Read Before Use....Intended use .................... Applicability of endoscopy and endoscopic treatment ......Instruction manual ..................User qualifications..................Instrument compatibility ................Reprocessing before the first use/reprocessing and storage after use ..Spare equipment..................Maintenance management ...............
  • Page 4 Contents Chapter 5 Troubleshooting ............110 Troubleshooting guide ..............Withdrawal of the endoscope with an irregularity......Returning the endoscope for repair..........Appendix..................119 System chart .................... EMC information..................EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL...
  • Page 5: Symbols

    Symbols Symbols The meaning(s) of the symbol(s) shown on the component packaging, the back cover of the instruction manual, and/or the instrument are as follows: Refer to instructions. Caution Endoscope TYPE BF applied part Manufacturer Authorized representative in the European Community Lot number Serial number EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL...
  • Page 6: Important Information - Please Read Before Use

    Important Information — Please Read Before Use Intended use These instruments have been designed to be used with an Olympus video system center, magnification controller (for CF-H260AZL/I only), endoscope position detecting unit (for CF-Q260DL/I, CF-H260DL/I only), light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment.
  • Page 7: Applicability Of Endoscopy And Endoscopic Treatment

    Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, please contact Olympus.  Terms used in this manual NBI (Narrow Band Imaging) observation mode: This is an observation mode using narrowband light.
  • Page 8: User Qualifications

    Important Information — Please Read Before Use User qualifications If there are official standards for user qualifications to perform endoscopy and endoscopic treatment that are defined by the hospital’s medical administrators or other official institutions, such as academic societies on endoscopy, follow those standards.
  • Page 9: Spare Equipment

    Equipment that has been disassembled, repaired, altered, changed, or modified by persons other than Olympus’ own authorized service personnel is excluded from Olympus’ limited warranty and is not warranted by Olympus in any manner. Signal words The following signal words are used throughout this manual: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
  • Page 10: Warnings And Cautions

    Important Information — Please Read Before Use Warnings and cautions Follow the warnings and cautions given below when handling this instrument. This information is to be supplemented by the warnings and cautions given in each chapter. • After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL”...
  • Page 11 Important Information — Please Read Before Use  Transnasal insertion is accompanied by the risk of inflammation of the nasal cavity. If this happens, the nasal passage will be constricted, making it more difficult to withdraw the endoscope. In this case, do not use force to withdraw the endoscope because patient injury, bleeding, and/or perforation may result.
  • Page 12 Important Information — Please Read Before Use • When using the PCF-PQ260L/I, be sure to check the passive bending section moving smoothly by touching it with your hands before inserting the endoscope into the patient. If any irregularity is observed on the distal end motion of the endoscope, immediately stop using the endoscope and withdraw it from the patient.
  • Page 13 Important Information — Please Read Before Use • The endoscope position detecting unit is designed only to assist the insertion of an endoscope. Never insert the endoscope into the patient’s body by observing only the endoscope position display of the endoscope position detecting unit.
  • Page 14 Important Information — Please Read Before Use • Do not touch the electrical contacts inside the electrical connector. CCD damage may result. • Do not apply shock to the distal end of the insertion section, particularly the objective lens surface at the distal end. Visual irregularities may result.
  • Page 15 Important Information — Please Read Before Use • Do not pull the UPD cable during an examination. The endoscope connector will be pulled out from the output socket of the light source and the endoscopic image will not be visible. The UPD cable will be pulled out from the UPD scope connector and the scope model will not be visible (for CF-Q260DL/I, CF-H260DL/I only).
  • Page 16 Important Information — Please Read Before Use • When using this equipment in domestic environments and connected to the public mains network, electromagnetic interference with other devices may occur. If electromagnetic interference occurs, mitigation measures may be necessary, such as reorienting or relocating this equipment, or shielding the location.
  • Page 17: Examples Of Inappropriate Handling

    Important Information — Please Read Before Use Examples of inappropriate handling Details on clinical endoscopic technique are the responsibility of trained specialists. Patient safety in endoscopic examinations and endoscopic treatment can be ensured through appropriate handling by the physician and the medical facility.
  • Page 18: Chapter 1 Checking The Package Contents

    Match all items in the package with the components shown below. Inspect each item for damage. If the instrument is damaged, a component is missing, or you have any questions, do not use the instrument; immediately contact Olympus. This instrument was not disinfected or sterilized before shipment.
  • Page 19 Chapter 1 Checking the Package Contents GIF-H260Z Endoscope CF-Q260AL/I, CF-H260AL/I, PCF-Q260AL/I Endoscope CF-H260AZL/I Endoscope EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL...
  • Page 20 Chapter 1 Checking the Package Contents CF-Q260DL/I, CF-H260DL/I Endoscope PCF-Q260AZI Endoscope PCF-Q260JL/I Endoscope EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL...
  • Page 21 Chapter 1 Checking the Package Contents Channel cleaning brush Water-resistant cap (BW-20T) (MH-553) Injection tube (MH-946) AW channel cleaning Channel-opening cleaning adapter (MH-948) brush (MH-507) Suction valve Air/water valve Channel plug (MH-944) (MH-443, 2 pcs) (MH-438, 2 pcs) Biopsy valve Suction cleaning adapter Chain for water-resistant cap (MB-358, 10 pcs)
  • Page 22: Chapter 2 Instrument Nomenclature And Specifications

    Chapter 2 Instrument Nomenclature and Specifications Chapter 2 Instrument Nomenclature and Specifications Nomenclature  GIF-N260 Universal cord 1. Suction connector 5. Electrical connector Air pipe 26. Chain connector 3. Air supply connector 3. Water supply connector Product name (model) and serial number Light guide Contact pins 4.
  • Page 23 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock 6. UP/DOWN angulation control knob Control section Suction cylinder Grip section Air/water cylinder 14. Color code Biopsy valve (MB-358) Boot 10. Instrument channel Instrument channel port 11.
  • Page 24 Chapter 2 Instrument Nomenclature and Specifications  GIF-XP260 GIF-PQ260 GIF-XQ260 GIF-Q260 GIF-H260 GIF-XP260 1. Suction connector 5. Electrical connector 26. Chain connector Air pipe Universal cord 3. Air supply connector 3. Water supply connector Light guide Product name (model) and serial number Contact pins GIF-PQ260 GIF-XQ260...
  • Page 25 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 16. RIGHT/LEFT angulation control knob Control Air/water cylinder 15. RIGHT/LEFT angulation lock section Grip section 14.
  • Page 26 Chapter 2 Instrument Nomenclature and Specifications  GIF-H260Z Universal cord 19. Auxiliary water inlet 1. Suction connector Auxiliary water inlet cap (MAJ-215) 5. Electrical connector Air pipe 2. S-cord connector mount 3. Air supply connector 3. Water supply connector Product name (model) and serial number Light guide Contact pins 4.
  • Page 27 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 18. Zoom lever 16. RIGHT/LEFT angulation control knob Control section Air/water cylinder 15. RIGHT/LEFT angulation lock Grip section 14.
  • Page 28 Chapter 2 Instrument Nomenclature and Specifications  CF-Q260AL CF-Q260AI CF-H260AL CF-H260AI PCF-Q260AL PCF-Q260AI Universal cord 1. Suction connector 5. Electrical connector 2. S-cord connector mount Air pipe 3. Air supply connector 3. Water supply connector Product name (model) and serial number Light guide Contact pins 4.
  • Page 29 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 16. RIGHT/LEFT angulation control knob Control Air/water cylinder 15. RIGHT/LEFT angulation lock section Grip section 14.
  • Page 30 Chapter 2 Instrument Nomenclature and Specifications  CF-H260AZL CF-H260AZI Mark 3 22. Seal ring Front view of the zoom connector Universal cord 1. Suction connector 2. S-cord connector mount 5. Electrical connector 20. Zoom connector Air pipe 21. Water-resistant cap (MAJ-583) 3.
  • Page 31 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 18. Zoom lever 16. RIGHT/LEFT angulation control knob Control section 15. RIGHT/LEFT angulation lock Air/water cylinder Grip section 14.
  • Page 32 Chapter 2 Instrument Nomenclature and Specifications  CF-Q260DL CF-Q260DI CF-H260DL CF-H260DI Mark 3 25. Seal ring Front view of the UPD scope connector Universal cord 1. Suction connector 2. S-cord connector mount 23. UPD scope connector 5. Electrical connector Air pipe 24.
  • Page 33 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 16. RIGHT/LEFT angulation control knob Control Air/water cylinder 15. RIGHT/LEFT angulation lock section Grip section 14.
  • Page 34 Chapter 2 Instrument Nomenclature and Specifications  PCF-PQ260L PCF-PQ260I Universal cord 1. Suction connector 5. Electrical connector 2. S-cord connector mount Air pipe 3. Air supply connector 3. Water supply connector Product name (model) and serial number Light guide Contact pins 4.
  • Page 35 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 16. RIGHT/LEFT angulation control knob Control Air/water cylinder 15. RIGHT/LEFT angulation lock section Grip section 14.
  • Page 36 Chapter 2 Instrument Nomenclature and Specifications  PCF-Q260AZI Universal cord 19. Auxiliary water inlet 1. Suction connector Auxiliary water inlet cap (MAJ-215) 5. Electrical connector Air pipe 2. S-cord connector mount 3. Air supply connector 3. Water supply connector Product name (model) and serial number Light guide Contact pins 4.
  • Page 37 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 18. Zoom lever 16. RIGHT/LEFT angulation control knob Control section 15. RIGHT/LEFT angulation lock Air/water cylinder Grip section 14.
  • Page 38 Chapter 2 Instrument Nomenclature and Specifications  PCF-Q260JL PCF-Q260JI Universal cord 19. Auxiliary water inlet 1. Suction connector Auxiliary water inlet cap (MAJ-215) 5. Electrical connector Air pipe 2. S-cord connector mount 3. Air supply connector 3. Water supply connector Product name (model) and serial number Light guide Contact pins...
  • Page 39 Chapter 2 Instrument Nomenclature and Specifications 9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob 16. RIGHT/LEFT angulation control knob Control Air/water cylinder 15. RIGHT/LEFT angulation lock section Grip section 14.
  • Page 40: Endoscope Functions

    Connects the endoscope to the suction tube of the suction pump. 2. S-cord connector mount (except GIF-N260, GIF-XP260) Connects the endoscope with the Olympus electrosurgical unit via the S-cord. The S-cord conducts leakage current from the endoscope to the electrosurgical unit. To connect the S-cord, refer to the instruction manual for the electrosurgical unit.
  • Page 41 Chapter 2 Instrument Nomenclature and Specifications 10. Instrument channel The instrument channel functions as:  Channel for the insertion of EndoTherapy accessories  Suction channel  Fluid feed channel (from a syringe via the biopsy valve) 11. Insertion section limit mark This mark shows the maximum point to that the endoscope may be inserted into the patient’s body.
  • Page 42 Chapter 2 Instrument Nomenclature and Specifications 17. Flexibility adjustment ring (for endoscopes with flexibility adjustment only) Turn this ring to adjust the flexibility of the insertion section. When the “” mark on the ring is aligned with the “ ” mark at the bottom of the grip section, the insertion section is the most flexible.
  • Page 43 Chapter 2 Instrument Nomenclature and Specifications 26. Chain connector (for GIF-N260, GIF-XP260 only) This connector connects the fitting part of the chain for water-resistant cap as required. Do not connect the S-cord of the electrosurgical unit to the chain connector. 27.
  • Page 44: Specifications

    Chapter 2 Instrument Nomenclature and Specifications Specifications Environment Operating Ambient temperature 10 – 40C (50 – 104F) environment Relative humidity 30 – 85% Atmospheric pressure 700 – 1060 hPa (0.7 – 1.1 kgf/cm (10.2 – 15.4 psia) Standard storage Ambient temperature 5 –...
  • Page 45 Chapter 2 Instrument Nomenclature and Specifications Specifications  Endoscope functions Model 1 GIF-N260 Optical system Field of view  Direction of view Forward viewing Depth of field 3 – 100 mm Insertion Distal end outer diameter ø 4.9 mm section Distal end enlarged 1.
  • Page 46 Chapter 2 Instrument Nomenclature and Specifications Model 1 GIF-XP260 Optical system Field of view  Direction of view Forward viewing Depth of field 3 – 100 mm Insertion Distal end outer diameter ø 5 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 47 Chapter 2 Instrument Nomenclature and Specifications Model GIF-PQ260 Optical system Field of view  Direction of view Forward viewing Depth of field 3 – 100 mm Insertion Distal end outer diameter ø 7.9 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 48 Chapter 2 Instrument Nomenclature and Specifications Model GIF-XQ260 Optical system Field of view  Direction of view Forward viewing Depth of field 3 – 100 mm Insertion Distal end outer diameter ø 9 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 49 Chapter 2 Instrument Nomenclature and Specifications Model GIF-Q260 Optical system Field of view  Direction of view Forward viewing Depth of field 3 – 100 mm Insertion Distal end outer diameter ø 9.2 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 50 Chapter 2 Instrument Nomenclature and Specifications Model GIF-H260 Optical system Field of view  Direction of view Forward viewing Depth of field 3 – 100 mm Insertion Distal end outer diameter ø 9.8 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 51 Chapter 2 Instrument Nomenclature and Specifications Model GIF-H260Z Optical system Field of view 140 (WIDE position) 75 (TELE position) Direction of view Forward viewing Depth of field 7 – 100 mm (WIDE position) 1.5 – 3 mm (TELE position) (The focus range is switchable.) Insertion Distal end outer diameter ø...
  • Page 52 Chapter 2 Instrument Nomenclature and Specifications Model CF-Q260AL CF-Q260AI Optical system Field of view  Direction of view Forward viewing Depth of field 4 – 100 mm Insertion Distal end outer diameter ø 12.2 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 53 Chapter 2 Instrument Nomenclature and Specifications Model CF-Q260DL CF-Q260DI Optical system Field of view  Direction of view Forward viewing Depth of field 5 – 100 mm Insertion Distal end outer diameter ø 12.2 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 54 Chapter 2 Instrument Nomenclature and Specifications Model CF-H260AL CF-H260AI Optical system Field of view  Direction of view Forward viewing Depth of field 5 – 100 mm Insertion Distal end outer diameter ø 13.2 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 55 Chapter 2 Instrument Nomenclature and Specifications Model CF-H260AZL CF-H260AZI Optical system Field of view  (WIDE position)  (TELE position) Direction of view Forward viewing Depth of field 7 – 100 mm (WIDE position) 2 – 3 mm (TELE position) (The focus range is switchable.) Insertion Distal end outer diameter...
  • Page 56 Chapter 2 Instrument Nomenclature and Specifications Model CF-H260DL CF-H260DI Optical system Field of view  Direction of view Forward viewing Depth of field 5 – 100 mm Insertion Distal end outer diameter ø 13.6 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 57 Chapter 2 Instrument Nomenclature and Specifications Model PCF-PQ260L PCF-PQ260I Optical system Field of view  Direction of view Forward viewing Depth of field 5 – 100 mm Insertion Distal end outer diameter ø 9.2 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 58 Chapter 2 Instrument Nomenclature and Specifications Model PCF-Q260AL PCF-Q260AI Optical system Field of view  Direction of view Forward viewing Depth of field 5 – 100 mm Insertion Distal end outer diameter ø 11.3 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 59 Chapter 2 Instrument Nomenclature and Specifications Model PCF-Q260AZI Optical system Field of view  (WIDE position)  (TELE position) Direction of view Forward viewing Depth of field 7 – 100 mm (WIDE position) 2 – 3.5 mm (TELE position) (The focus range is switchable.) Insertion Distal end outer diameter ø...
  • Page 60 Chapter 2 Instrument Nomenclature and Specifications Model PCF-Q260JL PCF-Q260JI Optical system Field of view  Direction of view Forward viewing Depth of field 5 – 100 mm Insertion Distal end outer diameter ø 10.5 mm section Distal end enlarged 1. Air/water nozzle 2.
  • Page 61 Chapter 2 Instrument Nomenclature and Specifications Medical This device complies with the requirements Devices of Directive 93/42/EEC concerning medical Directive devices. Classification: Class II a Applied standard: This instrument complies with the IEC 60601-1-2: 2001 standards listed in the left column. IEC 60601-1-2: 2007 CISPR 11 of emission: 1...
  • Page 62: Attaching The Chain For Water-Resistant Cap (Maj-1119)

    Chapter 2 Instrument Nomenclature and Specifications Attaching the chain for water-resistant cap (MAJ-1119) • Do not lift the endoscope by the chain for water-resistant cap. Doing so may result in the fitting part of the chain detaching from the S-cord connector mount or the chain connector of the GIF-N260, GIF-XP260, causing the endoscope to fall.
  • Page 63 Chapter 2 Instrument Nomenclature and Specifications Notch Chain part Fitting part Connecting plate Hole Figure 2.1 To ensure that you do not forget to attach the water-resistant cap, it is recommended that you connect the chain for water-resistant cap to the endoscope’s S-cord connector mount or the chain connector of the GIF-N260, GIF-XP260.
  • Page 64 Chapter 2 Instrument Nomenclature and Specifications Place the hole on the fitting part over the endoscope’s S-cord connector mount or the chain connector of the GIF-N260, GIF-XP260 (see Figure 2.3). Fitting part Hole S-cord connector mount or chain connector Figure 2.3 Confirm that the fitting part is securely attached to the foot of the S-cord connector mount or the chain connector of the GIF-N260, GIF-XP260 and can be smoothly rotated.
  • Page 65: Chapter 3 Preparation And Inspection

    Should any irregularity be observed after inspection, follow the instructions as described in Chapter 5, “Troubleshooting”. If this instrument malfunctions, do not use it. Return it to Olympus for repair as described in Section 5.3, “Returning the endoscope for repair” on page 118.
  • Page 66 Chapter 3 Preparation and Inspection Monitor Video system center Light source Magnification controller Water container (for CF-H260AZL/I only) Suction pump Endoscope Endoscope position detecting unit (for CF-Q260DL/I, CF-H260DL/I only) Mouthpiece (for GIF models only) EndoTherapy accessories Water pump (OFP) or a syringe Auxiliary water tube (for endoscopes with auxiliary (for endoscopes with auxiliary...
  • Page 67: Inspection Of The Endoscope

    Chapter 3 Preparation and Inspection Inspection of the endoscope Clean and disinfect or sterilize the endoscope as described in the “REPROCESSING MANUAL” with your endoscope model listed on the cover. Then remove the water-resistant cap from the endoscope connector. Inspection of the endoscope Inspect the control section and the endoscope connector for excessive scratching, deformation, loose parts, or other irregularities.
  • Page 68 Chapter 3 Preparation and Inspection Figure 3.3 When inspecting the PCF-PQ260L/I, confirm that the passive bending section bends smoothly by holding both ends of the passive bending section (about 100 mm from the distal end of the endoscope and the distal end of the insertion tube) using both hands and bending it lightly as shown in Figure 3.4 (the passive bending section can be bent almost roughly about 60).
  • Page 69 Chapter 3 Preparation and Inspection Inspect the objective lens and light guide lens at the distal end of the endoscope’s insertion section for scratches, cracks, stains, or other irregularities. Inspect the air/water nozzle at the distal end of the endoscope’s insertion section for abnormal swelling, bulges, dents, or other irregularities.
  • Page 70 Chapter 3 Preparation and Inspection Set the insertion section to the most flexible and most rigid conditions, respectively. In each case, hold the insertion section with two hands between 30 and 50 cm from the distal end, and bend it gently as shown in Figure 3.6.
  • Page 71 Chapter 3 Preparation and Inspection  Inspection for smooth operation Confirm that both the UP/DOWN and RIGHT/LEFT angulation locks move all the way in the “F ” direction (the GIF-N260 has only the UP/DOWN angulation lock). Turn the UP/DOWN and RIGHT/LEFT angulation control knobs slowly in each direction until they stop, and return them to their respective neutral positions (the GIF-N260 has only the UP/DOWN angulation control knob).
  • Page 72: Preparation And Inspection Of Accessories

    Chapter 3 Preparation and Inspection  Inspection of the UP/DOWN angulation mechanism Move the UP/DOWN angulation lock all the way in the opposite direction of the “F ” mark. Then turn the UP/DOWN angulation control knob in the “ U” or the “D ”...
  • Page 73 Chapter 3 Preparation and Inspection Confirm that the holes of the valves are not blocked (see Figures 3.9 and 3.10). Confirm that the valves are not deformed or cracked (see Figures 3.9 and 3.10). Check for excessive scratching or tears in the air/water valve’s seals (see Figure 3.9).
  • Page 74 Chapter 3 Preparation and Inspection Inspection of the biopsy valve The biopsy valve is a consumable that should be inspected as follows before each use. Replace it with a new one if any irregularity is observed during the inspection. An irregular, abnormal, or damaged valve can reduce the efficacy of the endoscope’s suction system, and may leak or spray patient debris or fluids, posing an infection control risk.
  • Page 75 Chapter 3 Preparation and Inspection Inspection of the auxiliary water inlet cap (for endoscopes with auxiliary water feeding only) Confirm that the auxiliary water inlet cap attached to the endoscope connector has no dents, cracks, or other irregularities (see Figure 3.13). Should any irregularity be observed, replace it with a new one as described in “Attaching the auxiliary water inlet cap (for endoscopes with auxiliary water feeding only)”...
  • Page 76 Chapter 3 Preparation and Inspection Inspection of the mouthpiece (for GIF models only) Do not use a mouthpiece that is damaged, deformed, or reveals other irregularities. Doing so may cause patient injury and/or equipment damage. Placing the mouthpiece in the patient’s mouth before the procedure prevents the patient from biting and/or damaging the endoscope’s insertion section.
  • Page 77: Attaching Accessories To The Endoscope

    Chapter 3 Preparation and Inspection Attaching accessories to the endoscope The air/water valve and the suction valve do not require lubrication. Lubricants can cause swelling of the valves’ seals, which will impair valve function. Attaching the suction valve Align the two metal ridges on the underside of the suction valve with the two holes in the suction cylinder.
  • Page 78 Chapter 3 Preparation and Inspection Attaching the air/water valve Attach the air/water valve to the air/water cylinder of the endoscope (see Figure 3.17). Confirm that the valve fits properly without any bulging of the skirt. Air/water valve Suction valve Skirt Air/water cylinder Suction cylinder Figure 3.17...
  • Page 79: Inspection And Connection Of Ancillary Equipment

    Chapter 3 Preparation and Inspection Attaching the auxiliary water inlet cap (for endoscopes with auxiliary water feeding only) If the auxiliary water inlet cap is not attached, attach the fitting ring to the auxiliary water inlet on the endoscope connector (see Figure 3.19). Fitting ring Auxiliary water inlet cap Auxiliary water inlet...
  • Page 80 Chapter 3 Preparation and Inspection The NBI observation mode is available when the video system center CV-260SL and the light source CLV-260SL, CLV-260NBI are used with the endoscope. The CLV-260NBI may not be available in some areas. Connection of the endoscope and ancillary equipment Firmly connect the suction tube from the suction pump to the suction connector on the endoscope connector.
  • Page 81 Chapter 3 Preparation and Inspection Air supply connector Water supply connector Water container’s connection adapter Endoscope connector Air supply channel Metal tip Water supply channel Figure 3.20 Align the mark on the videoscope cable LUCERA or 200 with mark 1 on the endoscope connector and push it in until it stops (see Figure 3.21).
  • Page 82 Chapter 3 Preparation and Inspection Align the notch on the endoscope plug of the magnification control cable with mark 3 on the zoom connector of the endoscope and push the endoscope plug into the zoom connector until it stops (for CF-H260AZL/I only, see Figure 3.22).
  • Page 83 Chapter 3 Preparation and Inspection Connect the suction tube from the suction pump to the suction connector on the endoscope connector (see Figure 3.24). Suction pump Suction connector Suction tube Figure 3.24 Open the auxiliary water inlet cap (for endoscopes with auxiliary water feeding only, see Figure 3.25).
  • Page 84: Inspection Of The Endoscopic System

    Chapter 3 Preparation and Inspection Inspection of the endoscopic system Inspection of the endoscopic image Do not stare directly into the distal end of the endoscope while the examination light is ON. Otherwise, eye injury may result. Turn the video system center, light source, monitor, magnification controller (for CF-H260AZL/I only) and endoscope position detecting unit (for CF-Q260DL/I, CF-H260DL/I only) ON and inspect the WLI and NBI endoscopic image as described in their respective instruction manuals.
  • Page 85 If this fails to stop air bubbles from being emitted, do not use the endoscope because there may be a malfunction. Contact Olympus. When the distal end of the insertion section is immersed less than 10 cm below the surface of the sterile water, a small amount of air bubbles may be emitted from the air/water nozzle even when the air/water valve is not operated.
  • Page 86 If the reattached or replaced suction valve fails to operate smoothly, the endoscope may be malfunctioning; stop using it and contact Olympus. • If the biopsy valve leaks, replace it with a new one. A leaking biopsy valve can reduce the efficacy of the endoscope’s...
  • Page 87 Chapter 3 Preparation and Inspection Inspection of the instrument channel Keep your eyes away from the distal end when inserting EndoTherapy accessories. Extending the EndoTherapy accessory from the distal end could cause an eye injury. Insert the EndoTherapy accessory through the biopsy valve. Confirm that the EndoTherapy accessory extends smoothly from the distal end.
  • Page 88 Chapter 3 Preparation and Inspection  For GIF-H260Z, PCF-Q260AZI Move the zoom lever toward “ T” until it stops and confirm that the image of an object located about 2 mm from the distal end is clearly visible. Move the zoom lever toward “W ”...
  • Page 89 Chapter 3 Preparation and Inspection If the auxiliary water channel is used for feeding water, never disconnect the auxiliary water tube during an examination; leave it attached until the endoscope is precleaned. If the auxiliary water tube is detached before precleaning, water remaining in the auxiliary water channel may be spilled on the surrounding equipment.
  • Page 90: Chapter 4 Operation

    Chapter 4 Operation Chapter 4 Operation This manual does not explain or discuss clinical endoscopic procedures. It only describes basic operation and precautions related to the operation of this instrument. Therefore, the operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique.
  • Page 91 Chapter 4 Operation • Never insert or withdraw the endoscope under any of the following conditions. Otherwise, patient injury, bleeding, and/or perforation can result.  While the EndoTherapy accessory extends from the distal end of the endoscope.  While the bending section is locked in position. ...
  • Page 92 Chapter 4 Operation • The bending section will never bend to the RIGHT or LEFT direction but to the UP or DOWN direction. To insert or withdraw, operate the endoscope by considering the direction in which the bending section is angulated. Never apply excessive force to the RIGHT or LEFT direction when inserting or withdrawing the endoscope.
  • Page 93 Chapter 4 Operation • Regardless of the flexibility of the endoscope’s insertion section, it can cause patient injury, bleeding, and/or perforation if it is forcibly inserted, withdrawn, and/or twisted with excessive force. It is generally believed that an endoscope with a more rigid insertion section is easier to manipulate in the intestines if used properly.
  • Page 94 Chapter 4 Operation • The endoscopic image may be disrupted while switching between WLI observation mode and NBI observation mode. Therefore, do not perform any endoscopic operation or treatment while switching between the WLI observation mode and NBI observation mode. Otherwise, injury in the body cavity may result.
  • Page 95: Insertion

    Chapter 4 Operation Insertion Holding and manipulating the endoscope The control section of the endoscope is designed to be held in the left hand. The air/water and suction valves can be operated using the left index finger. The UP/DOWN angulation control knob and the zoom lever can be operated using the left thumb.
  • Page 96 Chapter 4 Operation • Transnasal insertion of the endoscope should be performed carefully. If resistance to insertion is felt, or the patient reports pain, stop the insertion immediately. Otherwise, patient injury can result or the endoscope could become lodged and difficult to withdraw.
  • Page 97 Chapter 4 Operation > 10 cm Figure 4.2 Endoscopes with flexibility adjustment > 10 cm Figure 4.3  For GIF models Move the zoom lever toward “W ” until it stops so that the widest possible angle is shown in the endoscopic image (for GIF-H260Z only). If necessary, apply a medical-grade, water-soluble lubricant to the insertion section.
  • Page 98 Chapter 4 Operation  For CF/PCF models To determine the correct splinting tube to use with the endoscope, select one of the combinations shown in the “System chart” in the Appendix. Move the zoom lever toward “W ” until it stops so that the widest possible angle is shown in the endoscopic image (for CF-H260AZL/I, PCF-Q260AZI only).
  • Page 99 Chapter 4 Operation Flexibility adjustment (for endoscopes with flexibility adjustment only) • Do not change the insertion section’s flexibility abruptly. Otherwise, patient pain, injury, bleeding, and/or perforation can result. • If the endoscopic image moves suddenly or is lost while you are changing the insertion section’s flexibility, stop changing the flexibility and restore the optimum field of view.
  • Page 100 Chapter 4 Operation Before inserting or withdrawing the endoscope, set the insertion section to an appropriate level of flexibility by turning the flexibility adjustment ring as required (see Figure 4.4). Rigid Flexible Figure 4.4 When changing the insertion section’s flexibility during a procedure, turn the flexibility adjustment ring slowly, and closely monitor the position of the flexibility marks, the endoscopic image, and the patient’s condition.
  • Page 101 Chapter 4 Operation • If the biopsy valve is left uncapped during the procedure, debris or fluids could leak or spray from it, posing an infection control risk. When the valve is uncapped, place a piece of sterile gauze over it to prevent leakage. •...
  • Page 102 “Inspection of the suction function” on page 82, before using the endoscope again. If the thick fluid or solid matter cannot be discharged, stop using the suction function and contact Olympus. • When aspirating, maintain the suction pressure at the lowest level necessary to perform the procedure.
  • Page 103 Chapter 4 Operation Performing both air feeding and suction at the same time sometimes makes it easier to remove water droplets from the objective lens surface.  Auxiliary water feeding (for endoscopes with auxiliary water feeding only) Nothing other than sterile water should be used for auxiliary water feeding.
  • Page 104: Using Endotherapy Accessories

    Chapter 4 Operation Operating the zoom function (for GIF-H260Z, CF-H260AZL/I, PCF-Q260AZI only) Refer to the magnification controller’s instruction manual for the operation of the magnification controller (for CF-H260AZL/I only). When magnifying the endoscopic image, move the zoom lever toward “ T”...
  • Page 105 Chapter 4 Operation • When using EndoTherapy accessories, keep the distance between the distal end of the endoscope and the mucous membrane greater than the endoscope’s minimum visible distance so that the EndoTherapy accessory remains visible in the endoscopic image. If the distal end of the endoscope is placed closer than its own minimum visible distance, the position of the accessory cannot be seen in the endoscopic image, which could cause serious patient injury and/or...
  • Page 106 Chapter 4 Operation • When using a distal attachment, the distal end of the endoscope becomes longer and its outer diameter is larger. Handle the endoscope carefully so as not to cause perforation or other patient injury. When performing endoscopic treatment using this equipment, take extra care. •...
  • Page 107 Chapter 4 Operation • When the biopsy valve’s cap is detached from the main body, it may cause patient debris or fluids to leak or spray from the endoscope, posing an infection control risk. When the biopsy valve’s cap has to be detached, place a piece of sterile gauze over it to prevent leakage.
  • Page 108 Chapter 4 Operation • When the tip of the EndoTherapy accessory extends approximately 1 cm from the distal end of the endoscope, the accessory will appear in the endoscopic image. • When the accessory appears in the endoscopic image, it may also reflect the light from the endoscope and/or cast its shadow in the endoscopic image.
  • Page 109 Chapter 4 Operation Use of nonflammable gases (for CF/PCF models only) Performing treatment while the intestines are filled with a flammable gas could result in an explosion, fire, and/or serious patient injury. If the intestines contain a flammable gas, replace it with air or a nonflammable gas such as CO before performing high-frequency treatment or laser cauterization treatment.
  • Page 110 Chapter 4 Operation • The application of high-frequency current may interfere with the endoscopic image. This does not indicate a malfunction. • When the endoscope is used with the electrosurgical unit ESG-100, it is not necessary to use the S-cord. Laser cauterization treatment (except GIF-N260, GIF-XP260) •...
  • Page 111: Withdrawal Of The Endoscope

    Chapter 4 Operation Withdrawal of the endoscope • If blood unexpectedly adheres to the surface of the insertion section of the withdrawn endoscope, carefully check the condition of the patient. • If it becomes impossible to withdraw the transnasally inserted endoscope, pull its distal end out of the mouth, cut the flexible tube using wire cutters, and after ensuring that the cut section will not injure the body cavity or nasal cavity of the...
  • Page 112: Transportation Of The Endoscope

    Chapter 4 Operation Transportation of the endoscope Transporting within the hospital Set the insertion section to the most flexible condition (for endoscopes with flexibility adjustment only). When carrying the endoscope with the auxiliary water tube connected to the auxiliary water inlet, attach the clip of the auxiliary water tube to the universal cord (for endoscopes with auxiliary water feeding only, see Figure 4.6).
  • Page 113 Chapter 4 Operation Transporting outside the hospital Transport the endoscope in the carrying case. Always clean, disinfect, or sterilize the endoscope after removing it from the carrying case. If the endoscope is not cleaned, disinfected, or sterilized, it could pose an infection control risk.
  • Page 114: Chapter 5 Troubleshooting

    Some problems that appear to be malfunctions may be correctable by referring to Section 5.1, “Troubleshooting guide”. If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to Olympus for repair. Olympus does not repair accessory parts. If an accessory part becomes damaged, contact Olympus to purchase a replacement.
  • Page 115 Chapter 5 Troubleshooting Endoscope functions  Angulation Irregularity Possible cause Solution description Resistance is The angulation lock(s) is Rotate angulation lock(s) in the encountered when engaged. “F ” direction. rotating angulation control knob(s).  Air/water feeding Irregularity Possible cause Solution description No air feeding.
  • Page 116 Chapter 5 Troubleshooting  Suction Irregularity Possible cause Solution description The suction is The biopsy valve is not Attach it correctly. absent or attached properly. insufficient. The biopsy valve is Replace it with a new one. damaged. The suction pump is not set Adjust the suction pump’s setting as properly.
  • Page 117 Chapter 5 Troubleshooting  Image quality or brightness Irregularity Possible cause Solution description There is no video Not all power switches are Turn all power switches ON. image. An image is not The objective lens is dirty. Feed water to remove mucus, etc. clear.
  • Page 118 Chapter 5 Troubleshooting  Auxiliary water feeding (for endoscopes with auxiliary water feeding only) Irregularity Possible cause Solution description The auxiliary water The auxiliary water inlet cap Replace it with a new one. inlet cap is leaking. is worn out. The auxiliary water inlet cap Install the auxiliary water inlet cap is incorrectly installed.
  • Page 119: Withdrawal Of The Endoscope With An Irregularity

    If the endoscope or EndoTherapy accessory cannot be withdrawn from the patient smoothly, do not attempt to forcibly withdraw it. Should any irregularity be observed, immediately contact Olympus. Forcibly withdrawing the endoscope or EndoTherapy accessory may cause patient injury, bleeding, and/or perforation.
  • Page 120 Chapter 5 Troubleshooting Turn the UP/DOWN and RIGHT/LEFT angulation locks in the “F ” direction to release them (the GIF-N260 has only the UP/DOWN angulation lock). Carefully withdraw the endoscope while observing the endoscopic image. When a splinting tube is used, withdraw both the endoscope and the splinting tube together from the patient’s anus (for CF/PCF models only).
  • Page 121 Chapter 5 Troubleshooting When using an endoscope with the flexibility adjustment function, set the insertion section to the most flexible condition (for endoscopes with flexibility adjustment only). Turn the UP/DOWN and RIGHT/LEFT angulation locks in the “F ” direction to release them (the GIF-N260 has only the UP/DOWN angulation lock).
  • Page 122: Returning The Endoscope For Repair

    Olympus. Before returning the endoscope for repair, contact Olympus. With the endoscope, include a description of the malfunction or damage and the name and telephone number of the individual at your location who is most familiar with the problem.
  • Page 123: Appendix

    New products released after the introduction of the endoscopes may also be compatible for use in combination with the endoscopes. For further details, contact Olympus. If combinations of equipment other than those shown below are used, full responsibility is assumed by the medical treatment facility.
  • Page 124 Appendix Endoscope position detecting unit Magnification controller (for CF-Q260DL/I, CF-H260DL/I only) (for CF-H260AZL/I only) UPD cable Magnification (MAJ-966) control cable (MAJ-572, Magnification MAJ-573) controller (MAJ-570) Endoscope position detecting unit (UPD) CF-H260AZL/I CF-Q260DL/I, CF-H260DL/I Water pump (OFP) (for endoscopes Auxiliary water tube (MAJ-855) with auxiliary water (for endoscopes with auxiliary feeding only)
  • Page 125 Appendix Endoscope position detecting unit (UPD) Endoscopes Biopsy valve (MB-358) GIF-N260, GIF-XP260, GIF-PQ260, GIF-XQ260, GIF-Q260, GIF-H260, 4 Electrosurgical unit GIF-H260Z, CF-Q260AL/I, CF-Q260DL/I, CF-H260AL/I, CF-H260AZL/I, CF-H260DL/I, PCF-PQ260L/I, PCF-Q260AL/I, PCF-Q260AZI, PCF-Q260JL/I ESG-100 Position detecting probe 4 Electrosurgical unit (MAJ-1300) (for CF/PCF models except CF-Q260DL/I, CF-H260DL/I) PSD-30...
  • Page 126 Appendix Reprocessing equipment Channel cleaning Channel-opening brush (BW-20T) cleaning brush (MH-507) Single use channel-opening cleaning brush (MAJ-1339) Single use channel cleaning Endoscope washer (EW-30) brush (BW-201T) Endoscope reprocessor (OER, OER-A, OER-AW) AW channel cleaning adapter (MH-948) Channel plug Injection tube Ultrasonic cleaner (MH-944) (MH-946)
  • Page 127 Appendix  Video system center Endoscope CV-200 CV-230 CV-240 CV-260 CV-260SL      GIF-N260      GIF-XP260 – –    GIF-PQ260      GIF-XQ260 – –    GIF-Q260 –...
  • Page 128 Appendix  Accessories for EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE only Mouthpiece Endoscope MA-474 MB-142   GIF-N260   GIF-XP260   GIF-PQ260 –  GIF-XQ260 –  GIF-Q260 –  GIF-H260 –  GIF-H260Z  compatible – not compatible  Accessories for EVIS LUCERA COLONOVIDEOSCOPE only Splinting tube Endoscope...
  • Page 129 Appendix  EndoTherapy accessories Please note that some of the accessories may not be available in some areas. Biopsy forceps Biopsy forceps (fenestrated) Single side open With needle Alligator jaws type Standard type type Endoscope GIF-N260 – – FB-15K-1 FB-19K-1 GIF-XP260 –...
  • Page 130 Appendix Biopsy forceps (fenestrated) Elongated cups with Elongated cups With needle Alligator jaws type needle Endoscope GIF-N260 FB-21K-1 FB-34K-1 – – GIF-XP260 FB-21K-1 FB-34K-1 – – GIF-PQ260 FB-21K-1 FB-34K-1 – – GIF-XQ260 FB-21K-1 FB-23K-1 FB-24K-1 FB-36K-1 GIF-Q260 FB-21K-1 FB-23K-1 FB-24K-1 FB-36K-1 GIF-H260 FB-21K-1...
  • Page 131 Appendix Biopsy forceps (fenestrated) Swing jaws alligator Swing jaws alligator Swing jaws alligator type with tooth Rat tooth type type with tooth type with tooth (elongated cups with (elongated cups) needle) Endoscope GIF-N260 – FB-52K-1 – – GIF-XP260 – FB-52K-1 –...
  • Page 132 Appendix Disposable biopsy forceps (fenestrated) Swing jaws alligator Swing jaws Swing jaws alligator type with tooth Swing jaws (elongated cups with type with tooth (with needle) needle) Endoscope GIF-N260 FB-211K FB-221K FB-231K FB-241K GIF-XP260 FB-211K FB-221K FB-231K FB-241K GIF-PQ260 FB-211K FB-221K FB-231K FB-241K...
  • Page 133 Appendix Rotatable biopsy Rotatable biopsy forceps (fenestrated) forceps (fenestrated) Swing jaws alligator Elongated cups with Swing jaws alligator Standard type type with tooth needle type with tooth (elongated cups) Endoscope GIF-N260 FB-19KR-1 – – – GIF-XP260 FB-19KR-1 – – – GIF-PQ260 FB-19KR-1 –...
  • Page 134 Appendix Rotatable biopsy Grasping forceps forceps (fenestrated) Swing jaws alligator type with tooth W shape jaw Alligator jaws Rat tooth (elongated cups with needle) Endoscope GIF-N260 – FG-4L-1 – FG-14P-1 GIF-XP260 – FG-4L-1 – FG-14P-1 GIF-PQ260 – FG-4L-1 – FG-14P-1 FG-8L-1, GIF-XQ260 FB-55KR-1...
  • Page 135 Appendix Grasping forceps Rat tooth with Covered tips Sharp tooth Tripod type alligator jaws Endoscope GIF-N260 FG-20P-1 – – – GIF-XP260 FG-20P-1 – – – GIF-PQ260 FG-20P-1 – – – FG-42L-1, GIF-XQ260 FG-21L-1 FG-32L-1 FG-47L-1, FG-45L-1 FG-49L-1 FG-42L-1, GIF-Q260 FG-21L-1 FG-32L-1 FG-47L-1, FG-45L-1...
  • Page 136 Appendix Single use Grasping forceps grasping forceps Surgical scissors Pentapod type Basket type Tripod type Endoscope GIF-N260 – FG-17K-1 – – GIF-XP260 – FG-17K-1 – – GIF-PQ260 – FG-17K-1 – – GIF-XQ260 FG-46L-1 FG-16L-1 FG-600U FS-3L-1 GIF-Q260 FG-46L-1 FG-16L-1 FG-600U FS-3L-1 GIF-H260 FG-46L-1...
  • Page 137 Appendix Single use rotatable Rotatable clip fixing clip fixing device Loop cutter Heat probe device Standard type Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 – – – – GIF-XQ260 FS-5L-1 CD-21Z, CD-120U HX-110LR HX-201LR-135 GIF-Q260 FS-5L-1 CD-21Z, CD-120U HX-110LR...
  • Page 138 Appendix Single use rotatable Single use clip fixing device Ligating device ligating device Long type Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 – – – – GIF-XQ260 HX-201LR-135L HX-20L-1 HX-21L-1 HX-400U-30 GIF-Q260 HX-201LR-135L HX-20L-1 HX-21L-1 HX-400U-30 GIF-H260 HX-201LR-135L...
  • Page 139 Appendix Disposable Washing pipe Injection needle Standard type Standard type injection needle Endoscope GIF-N260 NM-8L-1, NM-9L-1 NM-201L series – PW-2L-1 GIF-XP260 NM-8L-1, NM-9L-1 NM-201L series – PW-2L-1 GIF-PQ260 NM-8L-1, NM-9L-1 NM-201L series – PW-2L-1 NM-200L series, GIF-XQ260 NM-4L-1 to 7L-1 PW-1L-1 –...
  • Page 140 Appendix Washing pipe Distal attachment Spray type Straight Oblique With rim Endoscope GIF-N260 PW-6P-1 – – – GIF-XP260 PW-6P-1 – – – GIF-PQ260 PW-6P-1 – – – GIF-XQ260 PW-5L-1 MH-462 MH-587 MH-593 GIF-Q260 PW-5L-1 MH-463 MH-588 MH-594 GIF-H260 PW-5L-1 MH-463 MH-588 MH-594 GIF-H260Z...
  • Page 141 Appendix Distal attachment Disposable distal attachment Wide opening Oblique with rim oblique with rim Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 – – – – GIF-XQ260 MAJ-289 MAJ-295 D-201-10704 D-206-02 GIF-Q260 MAJ-290 MAJ-296 D-201-10704 D-206-03 GIF-H260 MAJ-290 MAJ-296...
  • Page 142 Appendix  Electrosurgical accessories Please note that some of the accessories may not be available in some areas. Electrosurgical Disposable hot biopsy forceps snare Hot biopsy forceps Alligator-jaw step Standard oval Crescent fenestrated fenestrated Endoscope GIF-N260 – – – – GIF-XP260 –...
  • Page 143 Appendix Electrosurgical snare Hexagonal Oval Mini oval Oval with thorns Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 SD-8P-1 – – – GIF-XQ260 SD-6L-1 SD-9L-1, SD-11L-1 SD-12L-1, SD-13L-1 SD-16L-1 GIF-Q260 SD-6L-1 SD-9L-1, SD-11L-1 SD-12L-1, SD-13L-1 SD-16L-1 GIF-H260 SD-6L-1 SD-9L-1, SD-11L-1...
  • Page 144 Appendix Electrosurgical Disposable electrosurgical snare snare Mini oval with thorns Crescent Oval Mini oval Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 – SD-221L-25 – – GIF-XQ260 SD-17L-1 SD-221L-25 SD-210U-25 SD-210U-15 GIF-Q260 SD-17L-1 SD-221L-25 SD-210U-25 SD-210U-15 GIF-H260 SD-17L-1 SD-221L-25...
  • Page 145 Appendix Disposable electrosurgical snare Mini oval with soft Extra mini oval Oval with spiral Oval with soft wire wire Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 – – – – GIF-XQ260 SD-210U-10 SD-230U-20 SD-240U-25 SD-240U-15 GIF-Q260 SD-210U-10 SD-230U-20...
  • Page 146 Appendix Extra mini oval with Electrosurgical knife Single use electrosurgical knife soft wire Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 – – – – GIF-XQ260 SD-240U-10 KD-1L-1 KD-610L KD-611L GIF-Q260 SD-240U-10 KD-1L-1 KD-610L KD-611L GIF-H260 SD-240U-10 KD-1L-1 KD-610L...
  • Page 147 Appendix Single use electrosurgical knife Endoscope GIF-N260 – – – – GIF-XP260 – – – – GIF-PQ260 – – – – GIF-XQ260 KD-620LR KD-630L KD-640L KD-650L GIF-Q260 KD-620LR KD-630L KD-640L KD-650L GIF-H260 KD-620LR KD-630L KD-640L KD-650L GIF-H260Z KD-620LR KD-630L KD-640L KD-650L CF-Q260AL KD-620UR...
  • Page 148 Appendix Single use Single use electrosurgical incision forceps electrosurgical hemostatic forceps Endoscope GIF-N260 – – – GIF-XP260 – – – GIF-PQ260 – – – GIF-XQ260 FD-430L FD-420LR FD-410LR GIF-Q260 FD-430L FD-420LR FD-410LR GIF-H260 FD-430L FD-420LR FD-410LR GIF-H260Z FD-430L FD-420LR FD-410LR CF-Q260AL –...
  • Page 149: Emc Information

    Appendix EMC information  Guidance and manufacturer’s declaration — Electromagnetic emissions This model is intended for use by medical personnel in hospital environments and for use in the electromagnetic environment specified below. The customer or the user of this model should assure that it is used in such an environment.
  • Page 150 The customer or the user of this model should assure that it is used in such an environment. This instrument can be used with the high-frequency electrosurgical equipment that designated by Olympus. IEC 60601-1-2 IEC 60601-1-2 IEC 60601-1-2 (2007, 2001)
  • Page 151 Appendix IEC 60601-1-2 IEC 60601-1-2 IEC 60601-1-2 (2007, 2001) Compliance Immunity test (2014) (2007, 2001) Electromagnetic environment — level test level test level Guidance Voltage dips, 0% U < 5% U Same as left Mains power quality should be that of a short typical commercial or hospital (100% dip in U...
  • Page 152 Appendix IEC 60601-1-2 IEC 60601-1-2 IEC 60601-1-2 (2007, 2001) Compliance Immunity test (2014) (2007, 2001) Electromagnetic environment — level test level test level Guidance Radiated RF 3V/m (80 MHz – 3V/m (E Same as left Recommended separation distance IEC 61000-4-3 2.7 GHz) (80 MHz –...
  • Page 153 30 cm (12 inches) to any part of the video system center, including cables specified by Olympus. Otherwise, degradation of the performance of this equipment could result.  Guidance and manufacturer’s declaration — Cables used for EMC compliance testing Refer to the instruction manuals for each piece of equipment.
  • Page 154 Appendix  Recommended separation distances between portable and mobile RF communications equipment and this model This model is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this model can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and this model as recommended below, according to the maximum output power of the communications equipment.
  • Page 155 ©2010 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. No part of this publication may be reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP. OLYMPUS is a registered trademark of OLYMPUS CORPORATION. Trademarks, product names, logos, or trade names used in this document are generally registered trademarks or trademarks of each company.
  • Page 156 A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road, Chaoyang District, Beijing, 100027 P.R.C. Fax: (86)10-5976-1299 Telephone: (86)10-5819-9000 Olympus Tower 9F, 446, Bongeunsa-ro, Gangnam-gu, Seoul, Korea 135-509 Fax: (02)6255-3494 Telephone: (02)6255-3210 491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373...

Table of Contents