RVG 142, RVG 5200 and RVG 6200 User & Installation Guide
RVG 142, RVG 5200 and RVG 6200
Installation Guide
 
 
User &
 
 
 
 1 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RVG 142 and is the answer not in the manual?

Questions and answers

Summary of Contents for Carestream DENTAL RVG 142

  • Page 1       RVG 142, RVG 5200 and RVG 6200   User & Installation Guide     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  1 ...
  • Page 2: Table Of Contents

    Post Installation Procedure (Optional) ....................16 Acquiring Single Images Using RVG 142, RVG 5200 and RVG 6200 ........17 Preparing the Acquisition of a Single Image Using RVG 142, RVG 5200 & RVG 6200 ......17 Single Image X-ray Exposure Times ...................... 19 Acquiring Single Image Using RVG 142, RVG 5200 &...
  • Page 3     Troubleshooting ......................63 Contact Information ......................64 RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  3 ...
  • Page 4: Notice

    The information contained in this guide may be subject to modification without notice, justification or notification to the persons concerned. No part of this guide may be reproduced without the express permission of Carestream Dental LLC. The U.S. Federal law restricts this device to sale by or on the order of a physician.
  • Page 5: Conventions In This Guide

    CAUTION: Alerts you to a condition that might cause serious damage. IMPORTANT: Alerts you to a condition that might cause problems. Note: Emphasizes important information. Tip: Provides extra information and hints.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  5 ...
  • Page 6: Rvg 142, Rvg 5200 And Rvg 6200 Overview

    The RVG sensor is radio-sensitive. The active surface of the RVG sensor is the flat surface marked with #1 or #2 indicating the size of the sensor. (For RVG 142, only one sensor size is available.) The following table summarizes RVG sensor type, size and typical usage:...
  • Page 7: Sharing The Rvg Sensor Between Workstations

    Apply the same rules for positioning the RVG sensor in the mouth that you use in classic radiology. You can use different systems for positioning the RVG sensor in the mouth. A starter kit is provided with each RVG sensor (except RVG 142 - positioners can be purchased separately). X-ray Source Compatibility The RVG sensor is compatible with all X-ray sources that meet the current standards of intraoral radiology.
  • Page 8: Imaging Software Overview

    See Pairing RVG Connect Units with a Workstation. RVG sensor is connected to the workstation but an error has occurred. An error message will be displayed in a popup window.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  8 ...
  • Page 9: Control Panels Of Rvg Sensors In Cs Imaging

    Pseudo color 2 Pseudo color 2   CS Adapt full version Perio/Endo/ DEJ + 6 filter levels Perio/Endo + one filter level Dose indicator Dose indicator Dose indicator     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  9 ...
  • Page 10: Full Mouth Series (Fms) Image Aquisition Overview

    • Blue highlight: Frame in the revue and retake mode. This mode interrupts the automatic acquisition sequence. The retake images are displayed in the Retake Image gallery.  Figure 4    FMS Preferences RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  10 ...
  • Page 11 If you try to change the FMS template after you have finished your image acquisitions you are prompted with a warning that indicates that you risk loosing some of the images. Figure 5  FMS Retake Image Gallery RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  11 ...
  • Page 12 The FMS retake image gallery (B) only displays the images acquired for the frame highlighted in blue in the FMS template (A).       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  12 ...
  • Page 13: Setting Up Rvg 142, Rvg 5200 And Rvg 6200

    The Welcome to the RVGDriver Setup Wizard dialog box is displayed: Click Next. The Choose Components dialog box is displayed. IMPORTANT: Do NOT uncheck any boxes that are already checked. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  13 ...
  • Page 14 Then restart your computer (recommended but not mandatory). Connect the RVG sensor to a USB 2.0 port that is connected directly to the motherboard (generally located on the back of the workstation). RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  14 ...
  • Page 15 The Found New Hardware Wizard is displayed. Click Install the software automatically (Recommended), then click Next. Verify the RVG sensor by running the post installation procedure.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  15 ...
  • Page 16: Post Installation Procedure (Optional)

     on your desktop to start the RVG 142/5200/6200 Service Tools. The Service Tools home page is displayed. Click Post Installation Procedure then follow the instructions on screen. Activate the relevant license key when prompted.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  16 ...
  • Page 17: Acquiring Single Images Using Rvg 142, Rvg 5200 And Rvg 6200

    Cover the RVG sensor with a disposable hygienic sleeve specifically designed for each size of RVG sensor. IMPORTANT: Use a NEW hygienic sleeve for each new patient to prevent cross- contamination. Place the protected RVG sensor in the biteblock of the RVG sensor positioner. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  17 ...
  • Page 18 Move the X-ray source tube head close to the patient and align it with the tooth of the patient and the RVG sensor. IMPORTANT: Make sure that the tube head is not shaking. Select the X-ray exposure time according to the region of interest and the patient type.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  18 ...
  • Page 19: Single Image X-Ray Exposure Times

    IMPORTANT: These are suggested exposure times and need to be adjusted for your specific X-ray source. For dark images, reduce the exposure time and for grainy images, increase the exposure time.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  19 ...
  • Page 20: Acquiring Single Image Using Rvg 142, Rvg 5200 & Rvg 6200

    Use the Favorite function to select which filters are displayed in the Control Panel. Use the Acquisition Default function to automatically apply the selected filter at acquisition time. Import or export filter libraries.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  20 ...
  • Page 21: Acquiring Fms Images Using Rvg 142, Rvg 5200 And Rvg 6200

    Select an appropriate positioner for the region of interest and the size of the RVG sensor. Cover the RVG sensor with a disposable hygienic sleeve specifically designed for each size of RVG sensor. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  21 ...
  • Page 22 Move the X-ray source tube head to the patient and align it with the tooth of the patient and the RVG sensor. IMPORTANT: Make sure that the tube head is not shaking. Select the X-ray exposure time according to the region of interest and the patient type.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  22 ...
  • Page 23: Fms Images X-Ray Exposure Times

    IMPORTANT: These are suggested exposure times and need to be adjusted for your specific X-ray source. For dark images, reduce the exposure time and for grainy images, increase the exposure time.         RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  23 ...
  • Page 24: Acquiring Fms Images Using Rvg 142, Rvg 5200 & Rvg 6200

    Not satisfactory, for example, if the exposure quality indicator is red, retake the X-ray (see Retaking FMS Images with RVG 142, RVG 5200 and RVG 6200). Satisfactory, remove the X-ray source tube head. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  24 ...
  • Page 25 Use the Favorite function to select which filters are displayed in the Control Panel. • Use the Acquisition Default function to automatically apply the selected filter at acquisition time. • Import or export filter libraries.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  25 ...
  • Page 26: Retaking Fms Images With Rvg 142, Rvg 5200 And Rvg 6200

    To restart the automatic acquisition, click the next empty frame in the acquisition sequence. If you want to set your image adjustments to default, click (A). The CS Adapt Library window appears: RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  26 ...
  • Page 27 IMPORTANT: Do NOT pull the RVG sensor by its cable when you remove the hygienic protection. Clean and disinfect the RVG sensor after each patient (see the RVG 142, RVG 5200 and RVG 6200 Safety, Regulatory and Technical Specifications User Guide (SM847)).     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  27 ...
  • Page 28: Setting Up And Using Rvg Connect For Rvg 6200

    • Green LED indicates that the RVG Connect  unit is connected.     Ethernet cable connecting the PoE Injector to the LAN, or an Ethernet hub, or a workstation with two Ethernet cards. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  28 ...
  • Page 29     After three minutes of inactivity, the RVG Connect unit goes into standby mode. The RVG Connect unit is automatically  reactivated when you acquire an image. You can also touch any button to reactivate the RVG Connect unit.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  29 ...
  • Page 30: Rvg Connect Unit Overview

    • A short press of the button displays the menu and activates choices you make. 5 USB 2.0 port for the RVG sensor 6 RVG sensor holder  indicates that the patient name has not yet been received from the workstation. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  30 ...
  • Page 31 Once the patient name is received from the workstation, the screen displays the patient name at the top as in this example: 8 Ethernet cable connecting the RVG Connect unit to the  PoE Injector (power over Ethernet ). After three minutes of inactivity, the RVG Connect unit goes into standby mode. The RVG Connect unit is automatically  reactivated when you acquire an image. You can also touch any button to reactivate the RVG Connect unit.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  31 ...
  • Page 32: Rvg Connect Menu

    Manually Assigning Static IP Address. Displays the IPv6 address, prefix, and scope. This is useful if you are using IPv6 without Bonjour. By default, an RVG Connect unit has one IPv6 address (link-local). For an RVG Connect unit with several  IPv6 addresses, the information is displayed on several screens. Press the touch sensitive button under  to display the next screen.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  32 ...
  • Page 33: Imaging Software Overview

    See Pairing RVG Connect Units with a Workstation. RVG sensor is connected to the workstation but an error has occurred. An error message will be displayed in a popup window.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  33 ...
  • Page 34: Setting Up Rvg Connect

    The Welcome to the RVG Driver Setup Wizard dialog box is displayed. Click Next. The Choose Components dialog box is displayed. IMPORTANT: Do NOT uncheck any boxes that are already checked. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  34 ...
  • Page 35 Manual workstation configuration with a static IP address using the RVG Service Tools (used when Bonjour is not installed). Assigning IP Addresses to RVG Connect Unit Using RVG Service Tools without Bonjour. Click Install. The Installing dialog box is displayed. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  35 ...
  • Page 36     When the installation is complete, the Completing the RVG Setup Wizard dialog box is displayed. Click Finish.   RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  36 ...
  • Page 37: Installing Rvg Connect Hardware

    Ethernet cable to the side of the arm instead of the top. However, you must leave enough slack in the cable at each moving joint of the arm to allow free movement. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  37 ...
  • Page 38 IMPORTANT: Leave enough slack in the cable at each moving joint of the arm to allow free movement of the X-ray source. Check that the workstation is connected to the local area network with an ethernet cable.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  38 ...
  • Page 39: Network Configuration

    Manual workstation configuration with a static IP address using the RVG Service Tools; this is used when Bonjour is not installed (see Assigning IP Addresses to RVG Connect Unit Using RVG Service Tools without Bonjour). Alternatively you can manually enter an available IP address directly on the RVG Connect  unit (see Manually Assigning Static IP Address).   RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  39 ...
  • Page 40: Checking Dynamic Attribution Of Ip Address

    At first the IP address is not displayed as in the example on the left then after a few moments the IP address is displayed as in the example on the right.               RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  40 ...
  • Page 41: Activating Dhcp Configuration

    Briefly press the menu button on the side of the RVG Connect unit.  above the left touch sensitive button cancels the procedure. Press the right touch sensitive button under When the processing is completed the following information is displayed at the top of the screen.   RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  41 ...
  • Page 42: Assigning Default Static Ip Address With Automatic Device Detection

      displayed on the screen. Briefly press the menu button on the side of the RVG Connect unit.  above the left touch sensitive button cancels the procedure. Press the right touch sensitive button under The RVG Connect unit automatically restarts. Briefly press the menu button on the side of the RVG Connect unit. The following information is displayed at the top of the screen:   RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  42 ...
  • Page 43: Manually Assigning Static Ip Address

    Press the right touch sensitive button below The RVG Connect unit automatically restarts. Briefly press the menu button on the side of the RVG Connect unit. The manually entered IP address is displayed at the top of the screen.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  43 ...
  • Page 44: Assigning Ip Address To Rvg Connect Unit Using Rvg Service Tools With Bonjour

    Under Network Configuration, either: Click Obtain an IP address automatically, then click Apply. Manually enter the IP Address, Subnet Mask, and Default Gateway, then click Apply.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  44 ...
  • Page 45: Assigning Ip Addresses To Rvg Connect Unit Using Rvg Service Tools Without Bonjour

    Click Configure. The first page explains how to configure the network of the workstation using Microsoft Windows tools. The second page explains the hardware connections and how to put the RVG Connect unit into IP static  mode. The third page allows you to assign a specific IP address to the RVG Connect unit. Follow the instructions on screen.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  45 ...
  • Page 46: Managing List Of Ip Addresses For Use On Several Workstations

    Allows you to import or export the list as a file. This option is useful for a network manager who wants to manage the list for several workstations.         RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  46 ...
  • Page 47: Pairing Rvg Connect Units With A Workstation

    On the RVG Connect unit, click (A) to confirm that you want to unlock this unit. In the Sensor list, click RVG Connect unit with your workstation. to pair the selected The Sensor list is updated and displays the status Locked on this computer. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  47 ...
  • Page 48     This indicates that the RVG Connect unit and RVG Sensor are paired with  this workstation.                             RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  48 ...
  • Page 49: Associating Buttons On Rvg Connect Unit With Workstation (Optional)

        Associating Buttons on RVG Connect Unit with Workstation (Optional) After you have paired an RVG Connect unit with a workstation, you can optionally associate one or two buttons on the  RVG Connect unit with that workstation. Press either button or button  on the front of the RVG Connect unit for a few seconds to pair the workstation  with that button.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  49 ...
  • Page 50: Initializing The Rvg Sensor With Rvg Connect For First Time

    Wait for the initialization process to complete. Note: The first time you plug an RVG sensor into the RVG Connect unit, it can take awhile to initialize. The RVG Connect screen displays:  indicates that the RVG sensor is initialized.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  50 ...
  • Page 51: Preparing Acquisition Of Single Image Using Rvg Connect

    Cover the RVG sensor with a disposable hygienic sleeve specifically designed for each size of RVG sensor.   IMPORTANT: Use a NEW hygienic sleeve for each new patient to prevent cross- contamination. 8.    Place the protected RVG sensor in the biteblock of the RVG sensor positioner. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  51 ...
  • Page 52 IMPORTANT: Always insert the RVG sensor holding it horizontally for the comfort of the patient. 10.    Move the X-ray source tube head close to the patient and align it with the tooth of the patient and the RVG  sensor. IMPORTANT: Make sure that the tube head is not shaking. 11.    Select the X-ray exposure time according to the region of interest and the patient type (see Single Image X-ray Exposure Times).     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  52 ...
  • Page 53: Single Image X-Ray Exposure Times

    IMPORTANT: These are suggested exposure times and need to be adjusted for your specific X-ray source. For dark images, reduce the exposure time and for grainy images, increase the exposure time.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  53 ...
  • Page 54: Acquiring Single Image Using Rvg Connect

    Clean and disinfect the RVG sensor after each patient (see the RVG 142, RVG 5200 and RVG 6200 Safety, Regulatory, and Technical Specifications User Guide (SM847)). Use the CS Adapt Library software to manage the brightness/contrast filters. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  54 ...
  • Page 55 Use the Favorite function to select which filters are displayed in the Control Panel. • Use the Acquisition Default function to automatically apply the selected filter at acquisition time. • Import or export filter libraries.         RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  55 ...
  • Page 56: Preparing Acquisition Of Fms Images Using Rvg Connect

    Select an appropriate positioner for the region of interest and the size of the RVG sensor. Cover the RVG sensor with a disposable hygienic sleeve specifically designed for each size of RVG sensor. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  56 ...
  • Page 57 IMPORTANT: Make sure that the tube head is not shaking. Select the X-ray exposure time according to the region of interest and the patient type (see FMS Images X-ray Exposure Times).       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  57 ...
  • Page 58: Fms Images X-Ray Exposure Times

    IMPORTANT: These are suggested exposure times and need to be adjusted for your specific X-ray source. For dark images, reduce the exposure time and for grainy images, increase the exposure time.         RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  58 ...
  • Page 59: Acquiring Fms Images Using Rvg Connect

    Continue acquiring images until the whole FMS template is completed. Check the image and if the quality is: Not satisfactory, for example, if the exposure quality indicator is red, retake the X-ray (see Retaking FMS Images with RVG 142, RVG 5200 and RVG 6200). Satisfactory, remove the X-ray source tube head. RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  59 ...
  • Page 60 Use the Favorite function to select which filters are displayed in the Control Panel. • Use the Acquisition Default function to automatically apply the selected filter at acquisition time. • Import or export filter libraries.       RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  60 ...
  • Page 61: Retaking Fms Images With Rvg 142, Rvg 5200 And Rvg 6200

    To restart the automatic acquisition, click the next empty frame in the acquisition sequence. If you want to set your image adjustments to default, click (A). The CS Adapt Library window appears: RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  61 ...
  • Page 62 IMPORTANT: Do NOT pull the RVG sensor by its cable when you remove the hygienic protection. Clean and disinfect the RVG sensor after each patient (see the RVG 142, RVG 5200 and RVG 6200 Safety, Regulatory and Technical Specifications User Guide (SM847)).     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  62 ...
  • Page 63 Overview). the color of the icon shows the connection status. When the icon is red, an error has occurred and an error message will be displayed in a popup window.     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  63 ...
  • Page 64 List of Importers for European Union According to MDR 2017/745 Carestream Dental France SAS 4 rue F. Pelloutier, Croissy-Beaubourg, 77435 Marne-la-Vallée Cedex 2, France Carestream Dental Germany GmbH RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  64 ...
  • Page 65 70327 Stuttgart, Germany Carestream Dental Spain S.L.U. Paseo de la Castellana, 79 Madrid 28046, España Carestream Dental Italy S.r.l. Via Mario Idiojmi 3/3, Assago 20090 (MI), Italia     RVG 142, RVG 5200 and RVG 6200 User & Installation Guide  65 ...