Once AC power is connected, the power button located next to the
power and USB connectors on the left side of the 4500 FTIR, will
appear a solid red color. This means the battery pack is being
charged, but the instrument is off.
The 4500t FTIR contains a Molicel lithium-ion rechargeable battery pack. Agilent
NOTE
have fully charged the 4500t system prior to shipment, but the system should
also be plugged into a power outlet when not in use. The 4500t battery life under
continuous use is approximately 3.5 hours. It is recommended to allow the
battery pack to be re-charged at initial setup prior to turning on the power to the
instrument.
Molicel lithium-ion rechargeable battery pack specifications:
Battery operating specifications:
Agilent 4500 Series FTIR Operation Manual
Capacity (nominal):
Nominal voltage:
Energy:
Size:
Weight:
Operating voltage:
Charge voltage:
Cutoff voltage:
Temperature range
Discharge:
Charge:
Maximum discharge current:
(continuous)
Maximum charge current:
Getting Started
6600 mAh
11.1 V
73.2 Wh
149.5 x 89.3 x 20 mm
(5.8 x 3.5 x 0.8 in)
480 g (1.1 lb)
12.6 to 9.0 V
12.6 V
7.5 V
-20 to 60 °C (-4 to 140 °F)
0 to 45 °C (32 to 113 °F)
3.5 A
3.5 A
11