Laser Frequency Calibration test
The Laser Frequency Calibration test measures the frequency (X-
axis) accuracy. The test is conducted by measuring a spectrum of a
polystyrene film. Absorbance frequencies from this spectrum are
compared to frequencies set by NIST SRM 1921 polystyrene film. The
software allows administrative-level users to use the test results to
set the laser calibration. Laser calibrations should only be set after
consultation with an Agilent technical support engineer. For this test,
a background is measured first. After the background collection, a
spectrum of polystyrene film is measured as instructed by the
software. The polystyrene test sample must be pressed firmly against
the diamond sample interface during the sample collection.