Chapter8 Troubleshooting Guide
Code
Description in Report Editor
1
No signal
2
Overreach movement
4
Measurement timeout
85
Airway obstructed
86
Measurement cancelled
87
Cuff leak
89
cuff overpressure
102
Self-check failed
110
Out of range
115
System error
Solution
Check position of cuff, tighten cuff
Remain still during BP reading
Check air hose connections and make certain
cuff is tight
Check air hose connections and make certain
air tubing is not crimped.
Push START/STOP button to start reading.
Check air hose and cuff
Check for blocked or kinked air hose.
Return for servicing.
Retry again, if problem persists return for
servicing.
Return for servicing.