Page 1
Programmable AC Power Supply ASD-1600 USER MANUAL ISO-9001 CERTIFIED MANUFACTURER...
Page 2
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company. The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
SAFETY INSTRUCTIONS Table of Contents SAFETY INSTRUCTIONS ..........5 GETTING STARTED ............9 Main Features ............10 Appearance .............. 12 INSTALLATION .............. 17 Preparation for Use ..........18 Requirements of Input Power ........18 Input Connection ............. 19 Output Connection ..........21 Remote Sense Connection ........
Page 4
ASD-1600 User Manual REMOTE OPERAION ............82 Set GPIB address and RS-232 Parameters ....83 Programming ............85 Command List ............88 Command for Remote Control ......... 91 APPENDIX ..............124 Protection .............. 125 Specifications ............127 Certificate Of Compliance ........131 TTL Pin Assignment ..........
SAFETY INSTRUCTIONS AFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition. Safety Symbols These safety symbols may appear in this manual or on the instrument.
Page 6
ASD-1600 User Manual Safety Guidelines Do not place any heavy object on the ASD-1600. General Guideline Avoid severe impact or rough handling that leads to damaging the ASD-1600. CAUTION Do not discharge static electricity to the ASD- 1600.
Page 7
SAFETY INSTRUCTIONS Location: Indoor, no direct sunlight, dust free, Operation almost non-conductive pollution (Note below) Environment Relative Humidity: 20%~ 85% (no condensation) Altitude: < 2000m Temperature: 0°C to 50°C (Pollution Degree) EN61010-1:2010 and EN61010-2-030 specifies the pollution degrees and their requirements as follows. The ASD- 1600 falls under degree 2.
Page 8
ASD-1600 User Manual Power cord for the United Kingdom When using the power supply in the United Kingdom, make sure the power cord meets the following safety instructions. NOTE: This lead/appliance must only be wired by competent persons WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code: Green/ Yellow:...
GETTING STARTED ETTING STARTED The ASD-1600 is a high-power programmable AC power supply that provides low-distortion sine wave output and power measurement. Accurate digital calculation is done through DSP chip to control stable voltage and frequency output. And can do remote programming via GPIB and RS232.
ASD-1600 User Manual Main Features Local operation from the front panel keypad. Configuration Protection against Over-power, Over-load, Over-temperature and Fan-fail. Built-in isolation output relays. Remote operation via GPIB or RS-232C interface. Selective output voltage ranges with full scale ...
GETTING STARTED Accessories Please check the following accessories are included and not damaged during shipment. If there are damaged or shortage of the accessories, please contact GW Instek electronics or sales agent for help. Standard Accessories Part number Description Qty.
ASD-1600 User Manual Appearance Front Panel Programmable AC Power Source 150V / 300V / 600V, 60A / 30A / 15A, 9kVA Menu/ Home SAVE RECALL Page up Page down METER INDIV POWER ESC/ Enter Cancel Unlock TRIGGER...
Page 13
GETTING STARTED Item Symbol Description LCD panel, display setting and measurement results. LED indicator: OUT(green)indicates AC source normal output Trigger TRIGGER(yellow)indicates LIST, PULSE, STEP output Press this key to control ON/OFF state of AC output voltage Digital numerical keypad group: Used for parameter setting At main menu page, press this key to enter Menu/...
Page 14
ASD-1600 User Manual At main menu page, press this key to enter “advanced parameter setting ”page; at Page up Page down other pages, press this key to return to main menu Main switch: Turn on/off the AC power input POWER Back to local control in remote mode, negative sign in panel control Unlock...
GETTING STARTED Rear Panel Part A TTL I/O RS-232 SYNC GPIB SYNC TTL I/O GPIB RS-232 Part B The input terminals are inside the machine OUTPUT AC OUTPUT Ø 1 Ø 2 Ø 3 REMOTE SENSING Ø 11 Ø 22 Ø...
Page 16
ASD-1600 User Manual Power Line input Input connector. Please refer to page 19 Connector for connection...
If any damage is found, please file a claim against the carrier immediately. Do not return the instrument to the factory without obtaining the prior RMA acceptance from GW Instek. Preparation for Use ..........18 Requirements of Input Power ........18 Rating ....................
ASD-1600 User Manual Preparation for Use Before start using, the instrument must be connected with an appropriate AC line input. As the instrument is cooled by fans, it must be installed in an area with sufficient space for air circulation. It should be used in an area where the environment temperature does not exceed 40°C.
INSTALLATION Input Connection The input terminal block is located on the instrument rear panel. The temperature rating of the power cord must be rated at least for 85°C. The current rating of input power line must be greater than or equal to the maximum current rating of the AC source. It is recommended to put a 40A breaker (ABT A) (30A breaker (ABT B)) on AC input side.
Page 20
ASD-1600 User Manual 3. Make sure the removable safety cover does fully cover the AC input terminal. 380Vac 3 phase 5 INPUT wires Y Input Connection (ASD-1600 380V) 380V input OUTPUT AC OUTPUT terminal Ø 1 Ø 2 Ø 3 REMOTE SENSING Ø...
INSTALLATION Output Connection Output connection is different for single phase or three phase output, choose the correct one before connecting. Connect the load wire to “COM” and “φ 1/φ 2/φ 3“ terminal as figures on next page. To meet the safety requirements the wires to the load must use gauges large enough to prevent from overheating while carrying the output current.
ASD-1600 User Manual Remote Sense Connection The remote detection function of the AC power supply can monitor the load voltage and automatically compensate the voltage drop on the connection line to ensure that the voltage transmitted to the load terminal is the set voltage value. The user must set the remote sensing to ON to enable this function.
Page 23
INSTALLATION Parallel output and remote sense connection...
ASD-1600 User Manual Power on Before turning on the instrument, all protective earth Caution terminals, extension cords and devices connected to the instrument must be connected to a protective ground. Any interruption of the protective grounding will cause a potential shock hazard that might injure people.
Page 25
INSTALLATION Error messages will show on the LCD panel if error Caution occurs during turn on process. Please refer to page 125 for error code. Digital circuit inside AC source may not operate properly if turn on AC source immediately after turn off.
Page 26
ASD-1600 User Manual OCAL OERATION The AC source can be configured to operate in local or remote mode. The operation in remote mode through GPIB controller or RS-232C will be described in next chapter. This chapter describes the operation in local mode through the keypad on the front panel for data entry and test.
LOCAL OERATION GPIB, RS232 Communication Setting ........... 57 Switch menu to remote control........57 Switch from remote to local control ....... 57 Switch remote control between GPIB or RS232 ..57 Parameters setting for two interfaces ......58 Inhibit Output by Remote Control Signal (Remote Inhibit) ..60 Power On State .................
ASD-1600 User Manual Operation through Keypad and Knob LCD panel will show the self-test message as below after AC switch is on. Then will enter main menu setting page as below. Default SET MODE is “Parallel” means parallel output operation. Operator can set output voltage and frequency.
Page 29
LOCAL OERATION Advanced parameter setting menu User can make more detail setting from advanced parameter setting menu, or press Menu/ Home key again to return to main menu. On advanced parameter setting menu, user can press numerical key to do a selection. Press Enter key to enter individual parameter ...
ASD-1600 User Manual Menu Tree Main Menu Advanced parameter setting menu 1. SETUP 2. CONFIG 3. LIST 4. PULSE 5. STEP 6. SYNTHESIS Degree On Wave A Count Count Count Synthesis Waveform Degree Off Wave A CSIN CF List Degree Pulse V Step Vac, DV GAIN...
LOCAL OERATION Main Menu Setting After the AC source is on and initialization is finished, screen will show the main menu. Default output voltage and frequency is 110V and 60Hz. Voltage and frequency setting shown on upper place on the screen, lower place of the screen shows the measurement value, example shown as below.
Page 32
ASD-1600 User Manual Set V=220V and F=50Hz Example 1. Press Up arrow or Down arrow key to Operation step move cursor to V (default setting). 2. Change parameter value to 220 by numerical key or knob. Then press Enter key to confirm setting.
LOCAL OERATION Series/Parallel output mode Display At main menu, press OUT key to change editing (setting) or output mode. At output mode, AC source outputs according the voltage and frequency set at main menu. Measurement values are shown at the lower position of the screen. Following explains the meaning of measurement information at different output mode Voltage output on series mode Voltage output on...
ASD-1600 User Manual Meter Report display will keep unchanged until user switch display pages, for example users do operation as below. At OUT MODE switch to Meter Report Φ1 then stop output and start again after set new parameters. During the process described above r Report is stayed at Φ1 until user Enter key longer than , Mete...
Page 35
LOCAL OERATION V: Voltage (Vrms), voltage measurement value Parameter (root mean square), unit is Volt. definition of measurement VA: Apparent power, calculation equation is Vrms x Irms. VAR: Reactive power. Calculation equation is √VA F: Frequency, Hz is used as measurement ...
ASD-1600 User Manual Advanced Parameter Setting Menu Press Menu/ Home key at main menu to enter Operation step advanced parameter setting menu. User can select an item from item 1~6, then press Enter key to enter individual setting menu;...
LOCAL OERATION SETUP Function Operation step 1. Press numerical key on the front panel to select SETUP option when the advanced parameter setting menu appears. 2. Press Enter key to enter SETUP setting menu shown as figure below. 3. Press ► Page down key to move to next page. 4.
LOCAL OERATION Definitions for all parameters on SETUP setting menu Setting Range Parameter Definition Unit 150.0/ Maximum voltage setting V Limit 300.0 for all pages Maximum output Irms 64.0/32.0 A I Limit Delay time for over current OCP Delay protection Voltage Level Output voltage range 150 V/ 300 V Output voltage...
ASD-1600 User Manual Voltage Limit To limit the maximum output voltage for all pages, It is related to Voltage Level parameter, resolution is 0.1V. Voltage level V Limit 300V 0.0~300.0V I50V 0.0~150.0V It is firmware to set the limitation instead of by hardware. The process to set Voltage Limit = 200V shows as Example below.
LOCAL OERATION I Limit, OCP Delay I Limit represents current limit, it is used to set the maximum output current (rms). Its range is related to Voltage Level setting. Voltage Level I Limit 300V 0.0~32.0 A I50V 0.0~64.0 A OCP Delay represents over current protection (OCP) delay time. It is the delay time to shut down output voltage after output current reaches OCP point.
Page 42
ASD-1600 User Manual 2. Change value to 3A by numerical key or knob. Then press Enter key to confirm setting. 3. Press Up arrow or Down arrow key to move cursor to OCP Delay. 4. Change value to 1.5S by numerical key or knob.
LOCAL OERATION Voltage Level Setting ASD-1600 provides two output voltage levels used for low voltage and high voltage applications. At 300V level, AC source can provide 32.0A in total;At 150V level 64.0A in total can be provided, maximum output power is 6000VA. When switched from 300V to 150V level, AC source will check all voltage settings, and change the setting to 150V if any voltage setting checked is higher than 150V.
ASD-1600 User Manual Voltage Sense Voltage Sense determines output voltage measurement point is at AC output terminal (Vout) or input terminal at product under testing (Vsense). If it is set at Vsense, Remote Sense wire must be connected first, otherwise compensate error will occur. AC source will compensate the voltage drop of the output cable if Vsense is selected.
LOCAL OERATION Degree On, Off ASD-1600 can control the start phase angle (Degree On) or stop phase angle (Degree Off) of the output voltage waveform. Range for Degree On is 0.0~359.9, and is 0.0~360.0 for Degree Off. When Degree Off parameter is 0, the output stopped immediately no matter which output status is at.
ASD-1600 User Manual 4. Change parameter to 200 by numerical key or knob. Then press Enter key to confirm setting. Is Start, Is Interval Both the parameter ranges for Is Start and Is Intervals are 0.0~1000.0 msec, these parameters are used for the surge current measurement.
Page 47
LOCAL OERATION Set Is Start = 500ms and Is Interval = 200ms Example Operation step 1. Press Up arrow or Down arrow key to move cursor to Is Start. 2. Change parameter to 500.0 by numerical key or knob. Then press Enter key to confirm setting.
LOCAL OERATION Waveform Selection ASD-1600 provides two sets built in waveforms A and B. User can select Sin, CSin, Squa or other self-defined waveform from A or B set at CONFIG page, then go to SETUP page to set output voltage waveform for main menu.
ASD-1600 User Manual Buzzer When user press numerical key or rotate knob, there is sound from buzzer to confirm the operation, it can be muted if the confirmation is not needed. Mute the buzzer Example Operation Step 1. Press Up arrow or Down arrow key to move cursor to Buzzer.
LOCAL OERATION Disable (Lock) the Keys at Front Panel ASD-1600 allows the user to lock the parameters at front panel to avoid being changed accidently. After the parameters are locked, only Menu/ Home key and OUT key are enabled. Menu/ Home key allows user to return to SETUP menu to enable or disable lock.
ASD-1600 User Manual Set Series / Parallel Mode The ASD-1600 provides two output modes. Series: Series output mode. Simultaneously set the voltage and frequency of two phases. Parallel: Parallel output mode. Rotate knob to set this parameter, then press Example Enter key to confirm, about 800msec is needed to do the change.
ASD-1600 User Manual CONFIG Function Operation step 1. Press numerical key on the front panel to select CONFIG option when the advanced parameter setting menu appears. 2. Press Enter key to enter CONFIG setting menu shown as figure below. Definitions for all parameters on CONFIG setting menu Setting Range Parameter...
LOCAL OERATION Output status for next time ON/ OFF Output after AC on Output mode for next time Series, Parallel Phase Sel. after AC on Waveform A, B Generator ASD-1600 provides memories to store group A and B waveforms, both A and B groups have waveforms as below Sinusoidal (Sine) Waveforms ...
Page 56
ASD-1600 User Manual The processes to set group B waveform as cut Example sinusoidal waveform (Cut Sine) with CF 1.300. Operation step 1. Press Up arrow or Down arrow key to move cursor to Wave B. 2. Rotate Knob to change SIN to CSIN. Press Enter key to confirm setting.
Page 57
LOCAL OERATION GPIB, RS232 Communication Setting ASD-1600 provides remote control operation, there are GPIB or RS232 interface for selection. Only one interface can be activated at the same time. Controller Menu Remote (GPIB) Remote (RS232) (front panel) Operation Mode Menu (front panel) Remote (GPIB) Remote (RS232) Switch menu to remote control...
Page 58
ASD-1600 User Manual Parameters setting for two interfaces GPIB RS232 Address Odd/Even parity check(Parity) Baud rate (Baud Rate) 1-30 Even parity check (EVEN) 9600 Odd parity check (ODD) 19200 No parity check (NONE) Set GPIB address to 20 Example Operation step 1.
Page 59
LOCAL OERATION 2. Rotate Knob to change parameter value from 9600 to 19200. The press Enter key to confirm setting. 3. Press Up arrow or Down arrow key to move cursor to RS232 Parity. 4. Rotate Knob to change parameter from NONE to ODD.
Page 60
ASD-1600 User Manual Inhibit Output by Remote Control Signal (Remote Inhibit) Remote Inhibit is an “Active HIGH” TTL signal, high level is 5V, low level is 0V. Its signals send to 9 pin TTL signal connector (D- SUB) at back panel of the AC source. Please refer to appendix “TTL Pin Assignment”...
Page 61
LOCAL OERATION TRIG: TTL signal changes from HIGH to LOW, and keeps low longer than 1ms, ASD-1600 turn off output, and stop detecting TTL signal, output stays at off state even TTL signal changes to high. User must press OUT key to resume output and enable TTL signal detecting.
Page 62
ASD-1600 User Manual 2. Rotate Knob to change parameter value from OFF to TRIG. Then press Enter key to confirm setting. Power On State The ASD-1600 may set different operation modes or enable turn on output immediately next time after AC is on. Output voltage and frequency depend on stored parameter before AC input is off.
Page 63
LOCAL OERATION 3. Press Up arrow or Down arrow key to move cursor to Phase Sel. 4. By pressing numeric key or rotating knob to change parameter to Single. Then press Enter key to confirm setting.
ASD-1600 User Manual PPLICAION ASD-1600 provides not only stable AC voltage but also three kinds of programmable dynamic output. Following paragraphs explain the definition of parameters and how to do the setting. LIST Mode ............... 65 Definitions for all parameters on LIST setting menu ....66 PULSE Mode ............
APPLICAION LIST Mode Let user to set 10 sets of user-defined voltage, frequency. This mode can make 10 different outputs consecutively. Operation step 1. Press numerical key on the front panel to select LIST option when the advanced parameter setting menu appears. 2.
ASD-1600 User Manual LIST mode provides 10 sets of programmable waveform (SEQ-0 ~ SEQ-9). Output waveforms start from SEQ-0, then SEQ-1, SEQ-2… in orders until time parameter (Time) of SEQ-n being processed is 0 or SEQ-9 is finished outputting, then one List waveform output is completed.
Page 67
APPLICAION Press Enter key to enter Trigger setting page shown as below. Trigger ON⁄ OFF is output switch for List waveform, output status is current output status. At this page, there are two modes. 1. To output List waveform Press Enter key to start output List waveform.
Page 68
ASD-1600 User Manual Output will stop when List waveform output is completed and stay at Trigger page. 2. To output voltage set at main menu Press OUT key will start output waveform set at main menu. Screen will show information as below during output is in process.
ASD-1600 User Manual PULSE Mode Simulate repeatedly changed output voltage. Operation step 1. Press numerical key on the front panel to select PULSE option when the advanced parameter setting menu appears. 2. Press Enter key to enter PULSE setting menu shown as figure below.
Page 71
APPLICAION 300.0 Voltage of Pulse: Frequency of Pulse 30.0 1000.0 Duration of total period 16000000.0 Period Duration time of Pulse, its value can’t 15999999.0 Duty be greater than the Period. Degree On Start phase angle of Pulse output 359.0 Select Go to Trigger Mode, press Enter key to Operation step ...
APPLICAION STEP Mode Simulate gradually changed output voltage. Operation step 1. Press numerical key on the front panel to select STEP option when the advanced parameter setting menu appears. 2. Press Enter key to enter STEP setting menu shown as figure below. Voltage and frequency are same for each step in STEP mode, but may be different in different steps.
Page 74
ASD-1600 User Manual Frequency change for each step. May be -150.0 150.0 positive or negative Output duration for each step 16000000 Dwell Degree On Initial phase angle for each STEP For each STEP, voltage waveform changes satisfy following equation. + dV ...
ASD-1600 User Manual SYNTHESIS Self-Defined Waveform Mode ASD-1600 also provides 39 orders of amplitude and phase angle settings for user to make self-defined waveforms. It can be used at all output modes once the self-defined waveforms are created. Operation step 1.
APPLICAION Definitions for all parameters on STEP setting menu Setting Range Parameter Definition Min. Max. Name for two sets of SYNTHESIS DST30/ DST31 Waveform waveform Order 2~20 33.33 Voltage amplitude ratio of Order 21~30 Harmonic to fundamental wave Gain 30.00 for each order.
Page 78
ASD-1600 User Manual It takes about 8 seconds to do the calculation and saving. It will return to first page of SYNTHESIS when saving is completed. For protection purpose, amplitude for each order must not exceed limitation as below. Orders Gain limit 33.33% 2~20...
SAVE AND RECALL AVE AND RECALL ASD-1600 provides two modes for users to save and recall the output setting or the system information. System parameter Save and Recall ......80 Save the main page setting ............. 80 Recall the main page setting ............81...
ASD-1600 User Manual System parameter Save and Recall The ASD-1600 provides 4 set of memories for user to save and recall modified system information. User can back to default system setting easily any time by selecting default setting. Save the main page setting How to save current system settings to first memory No.1.
Page 81
SAVE AND RECALL Recall the main page setting How to back to default factory settings Operation step 1. At advanced parameters setting manual, long press Recall key for two seconds will enter recall system setting page. DEFAULT is the default factory settings.
ASD-1600 User Manual EMOTE OPERAION ASD-1600 can be remotely controlled by GPIB or RS-232. GPIB interface is an 8-bit parallel data bus with other control lines to manage communication. RS-232 transmits data in series so communication speed is slower. If transmission speed is faster than data processing speed of micro-processor, micro- processor will neglect new received command when there is command under processing.
REMOTE OPERAION Set GPIB address and RS-232 Parameters GPIB and RS-232 parameters can be set at CONFIG page. Default address of GPIB is 30, baud rate of RS232 is 9600, and Even/odd parity check of RS232 is NONE. RS-232 pin definition RS-232 interface uses RXD and TXD signals only.
Page 84
ASD-1600 User Manual remotely. Press - / Unlock key to switch from remote control back to local control, other keys are inactive during remote control. Screen shows >REMOTE< when AC source is Operation step in remote control as below...
REMOTE OPERAION Programming All commands and responses are transferred in the form of ASCII code. Response must be received completely before sending a new command. Otherwise, error will occur due to message disorder. Conventions <> Items in angle brackets are abbreviations for Angle bracket parameter.
Page 86
ASD-1600 User Manual higher level commands are more close to left side, and the more detailed commands are more close to right side. ASD-1600 includes two distinctive types of command. The first one is command described in IEEE 488.2, it is common commands for GPIB, represented by “*”...
Page 87
REMOTE OPERAION VOLT:AC Example 100;:FETCh:CURRent:AC? Explanation The above instruction has two commands which are VOLT:AC 100 and FETCh:CURRent:AC?. Using (;:) can connect commands more than two. Space( ): Space is used to separate command and number or string after it. Instrument differentiates command and data by space.
REMOTE OPERAION Command for Remote Control IEEE 488.2 Standard Commands *ESE ..................91 *ESR ..................92 *IDN ..................92 *RCL ..................92 *SAV ..................93 *SRE ..................93 *STB ..................94 *CLS ..................94 *ESE Query Set and query flag mask for standard event status Description register.
ASD-1600 User Manual Power on Syntax *ESE<NR1> Query Syntax *ESE? Parameter/ <NR1> Return parameter *ESR Query Query value stored in standard events status Description register, 1 represents event occurs. Query Syntax *ESR? *IDN Query Return string of AC source identification. Description Query Syntax *IDN?
REMOTE OPERAION *SAV Save system setting Description Syntax *SAV<n> Return parameter 1 *SRE Query Set and return status of flag mask status for Description service events. This command is for a register used for IEE488.2 and instrument defined service events. It is used for flag mask or control enable.
ASD-1600 User Manual Syntax *SRE <n> Query Syntax *SRE? Parameter/ <NR1> Return parameter *STB Query Query value stored in service event register, 1 Description represents event occurs. Query Syntax *STB? Return parameter <NR1> *CLS Clear status. Following operation may execute Description depends on different conditions.
Page 95
REMOTE OPERAION FETCH and MEAURE Commands Differences between FETCH and MEASURE are: FETCH reads existing measurements value. Execution time is shorter because data is available immediately but the data may not be the newest. Instrument will send the next sampling data to the computer after receiving MEASURE command.
ASD-1600 User Manual OUTPUT Commands :OUTPut ................102 :OUTPut:MODE ..............102 :OUTPut Query Set or query status of turning on or turning off Description output. Syntax OUTPut Query Syntax OUTPut? Parameter/Return parameter Example OUTPut? OUTPut ON Query output status. Turn on output. :OUTPut:MODE Query Set or query output mode.
Page 103
REMOTE OPERAION Example OUTPut:MODE? OUTPut:MODE LIST Query output mode. Set output as LIST mode.
REMOTE OPERAION Parameter/Return parameter Example [:SOURce:]FUNCtion:SHAPe? [:SOURce]:FUNCtion:SHAPe:A Query Set or query waveform of waveform buffer A. Description Syntax [:SOURce:]FUNCtion:SHAPe:A Query Syntax [:SOURce:]FUNCtion:SHAPe:A? Parameter/ SINE Return parameter CSIN DST<00~31> Example [:SOURce:]FUNCtion:SHAPe:A? CSIN [:SOURce]:FUNCtion:SHAPe:A:CF Query Set or query CF of the cut sin waveform in Description waveform buffer A.
ASD-1600 User Manual Syntax [:SOURce:]FUNCtion:SHAPe:B Query Syntax [:SOURce:]FUNCtion:SHAPe:B? Parameter/ SINE Return parameter CSIN DST<00~31> Example [:SOURce:]FUNCtion:SHAPe:B? CSIN [:SOURce]:FUNCtion:SHAPe:B:CF Query Set or query CF of the cut sin waveform in Description waveform buffer B. It is effective only when CSIN is selected for waveform buffer B. Syntax [:SOURce:]FUNCtion:SHAPe:B:CF Query Syntax...
REMOTE OPERAION [:SOURce]:VOLTage:AC:1 Query Set or query series output AC voltage setting Description value; query the 1st phase AC voltage setting value of series output. Syntax [:SOURce:]VOLTage:AC:1 Query Syntax [:SOURce:]VOLTage:AC:1? Effective range: Parameter/ <NR2> 0.0 ~ 150.0 (150V voltage range, Return parameter 0.0 ~ 300.0 (300V voltage range) Example...
ASD-1600 User Manual Syntax [:SOURce:]VOLTage:RANGe Query Syntax [:SOURce:]VOLTage:RANGe? Parameter/ Return parameter HIGH Example [:SOURce:]VOLTage:RANGe? HIGH [:SOURce]:VOLTage:SENSe Query Set or query remote sense function. REMOTE is in Description enable, VOUT is in disable Syntax [:SOURce:]VOLTage:SENSe Query Syntax [:SOURce:]VOLTage:SENSe? Parameter/ VOUT Return parameter REMOTE Example [:SOURce:]VOLTage:SENSe?
REMOTE OPERAION [:SOURce]:PHASe:ON Query Set or query waveform start phase angle set at Description main menu. Syntax [:SOURce:]PHASe:ON Query Syntax [:SOURce:]PHASe:ON? Effective rang: 0.0 ~ 359.9 Parameter/ <NR2> Return parameter Example [:SOURce:]PHASe:ON? 200.5 [:SOURce]:PHASe:OFF Query Set or query waveform stop phase angle set at Description main menu.
ASD-1600 User Manual [:SOURce]:LIST:COUNt Query Set or query how many times to execute effective Description sequences in LIST mode. Syntax [:SOURce:]LIST:COUNt Query Syntax [:SOURce:]LIST:COUNt? Effective rang: 0 ~ 10000 Parameter/ <NR1> Return parameter Example [:SOURce:]LIST:COUNt? [:SOURce]:LIST:DWELl Query Set or query execution time of 10 individual Description sequences in LIST mode.
REMOTE OPERAION Example [:SOURce:]LIST:SHAPe? A B A A A [:SOURce]:LIST:VOLTage:AC:STARt Query Set or query start voltage of 10 individual Description sequences in LIST mode. Syntax [:SOURce:]LIST:VOLTage:AC:STARt Query Syntax [:SOURce:]LIST:VOLTage:AC:STARt? Effective rang: Parameter/ <NR2>… Return parameter 0.0 ~ 150.0 (low voltage range), <NR2>...
ASD-1600 User Manual Syntax [:SOURce:]PULSe:COUNt Query Syntax [:SOURce:]PULSE:COUNt? Effective rang: 0 ~ 10000 Parameter/ <NR1> Return parameter Example [:SOURce:]PULSe:COUNt? [:SOURce]:PULSe:DCYCle Query Set or query execution time of PULSE waveform, Description it must be no longer than total period in PULSE mode.
ASD-1600 User Manual [:SOURce]:SYNThesis:PHASe Query Set or query phase for each order of harmonic Description waveforms. Syntax [:SOURce:]SYNThesis:PHASe Query Syntax [:SOURce:]SYNThesis:PHASe? Effective rang: 0.0 ~ 359.9 <NR2>… <NR2> [:SOURce:]SYNThesis:PHASe? 100.5 20.8 60.5 77.8...
ASD-1600 User Manual Example :NPHase? :NPHase SERIES Query output mode set output as series mode It needs 800msec to make this command take effect, Note please reserve delay time to make sure operation normally. :INSTrument:NSELect 1|2 Query Set or query specific output phase for responding Description queries Syntax...
REMOTE OPERAION Return Parameter S00E02 Example :VERion:LCM? S00E02 :VERion:UI? Query Query firmware version of UI. Description Syntax :VERion:UI? Return Parameter S00E02 Example :VERion:UI? S00E02 :SYSTem:ERRor Query Query error status. It must be used together with Description command “INSTrument: NSELect” Syntax :SYSTem: ERRor? Return Parameter Return NORMAL for no errors.
APPENDIX Protection The ASD-1600 provide software and hardware protections. The ASD-1600 will cut off output relay to turn off output and display protection status on screen when protection happens. To resume the output, following three errors can be released by long press any key.
Page 126
ASD-1600 User Manual Fan malfunction FAN Fault Input current reaches trip point of built in no fuse Input OCP breaker** AC mains voltage higher or lower than rating ranges PFC AC Fault PFC OVP Fault Over voltage for PFC Board PFC OTP Fault Over temperature for PFC Board Output abnormal for PFC Board PFC PG Fault...
APPENDIX Specifications The operation specifications of ASD-1600 programmable AC Power Source are listed below. All specifications have been tested according to the standard of Delta test procedures. All specifications are based on resistor load with remote sense connection, under 25±2°C unless specified otherwise. Input ratings (AC rms) 220Vac to 240 Vac (L-L) 3 φ...
Page 128
ASD-1600 User Manual Single phase, Three-wire Output phase 90A (150V) 150V Maximum 30A (300V) 300V current 15A (600V) 600V 180A (30Hz~100Hz) 150V 150A (>100Hz~1kHz) 90A (30Hz~100Hz) Maximum 300V 75A (>100Hz~1kHz) peak current 45A (30Hz~100Hz) 600V 38A (>100Hz~1kHz) 0 to 1(leading phase or lagging phase) Load power factor 6000 VA Power capacity...
Page 129
APPENDIX 0.1W Active (W) Resolution 0.1 VA Power Apparent (VA) Resolution 0.1 VAR Reactive (VAR) Resolution 0.000 to 1.000 Range Load power factor 0.001 Resolution 1.2 to 1.414 Range Load crest factor 0.001 Resolution Up to 39th order of the Harmonic voltage Range fundamental wave...
Page 130
ASD-1600 User Manual Operating Indoor use, Overvoltage Category II environment Operating 0 °C to 40 °C temperature range Environment Storage -40 °C to 60 °C temperature range Storage humidity 90 % RH or less (no condensation) range Up to 2000 m Altitude 546(W) ×...
APPENDIX Certificate Of Compliance GOOD WILL INSTRUMENT CO., LTD. declare that the CE marking mentioned product satisfies all the technical relations application to the product within the scope of council: Directive: EMC; LVD; WEEE; RoHS The product is in conformity with the following standards or other normative documents: ◎...
APPENDIX Built In Waveforms DST 0 Order Gain Phase 2.07 9.80 15.80 0.0 2.16 DST 1 Order Gain Phase 1.50 1.50 2.00 DST 2 Order Gain Phase 2.00 1.40 2.00 1.40 1.00...
Page 134
ASD-1600 User Manual DST 3 Order Gain Phase 2.50 1.90 2.50 1.90 1.10 1.50 1.10 DST 4 Order Gain Phase 1.10 2.80 1.40 2.30 1.50 DST 5 Order Gain Phase 1.65 4.20 3.45 1.05 3.00...
Page 135
APPENDIX DST 6 Order Gain Phase 2.20 5.60 2.80 4.60 3.00 1.40 1.00 DST 7 Order Gain Phase 4.90 1.60 2.70 1.40 2.00 1.10 DST 8 Order Gain Phase 7.35 2.40 4.05 2.10 1.05 3.00 1.65 1.05 1.05 1.20 1.05...
Page 136
ASD-1600 User Manual DST 9 Order Gain Phase 9.80 3.20 5.40 1.20 2.80 1.40 4.00 2.20 1.40 1.40 1.60 1.40 DST 10 Order Gain Phase 17.75 0.0 DST 11 Order Gain Phase 21.25 0.0...
Page 137
APPENDIX DST 12 Order Gain Phase 24.50 0.0 DST 13 Order Gain Phase 2.30 9.80 15.80 0.0 2.50 DST 14 Order Gain Phase 1.15 4.90 7.90 1.25...
Page 138
ASD-1600 User Manual DST 15 Order Gain Phase 1.15 4.90 DST 16 Order Gain Phase 11.00 180.0 4.05 2.00 180.0 1.30 DST 17 Order Gain Phase 7.17 3.42 180.0 0.80...
Page 139
APPENDIX DST 18 Order Gain Phase 8.11 3.48 180.0 1.00 DST 19 Order Gain Phase 9.38 3.44 180.0 1.15 DST 20 Order Gain Phase 2.06 180.0 1.77 1.62 180.0 1.23 0.91 180.0 0.54 0.51 0.53 180.0...
Page 140
ASD-1600 User Manual DST 21 Order Gain Phase 3.08 180.0 2.72 2.43 180.0 1.97 1.41 180.0 0.86 0.62 180.0 0.73 0.77 180.0 0.69 0.56 180.0 DST 22 Order Gain Phase 0.13 180.0 4.28 180.0 3.77 3.27 180.0 2.57 1.93 180.0 1.22 0.55 180.0...
Need help?
Do you have a question about the ASD Series and is the answer not in the manual?
Questions and answers