Programmable AC Power Supply ASD-1300 USER MANUAL ISO-9001 CERTIFIED MANUFACTURER...
Page 2
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company. The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
SAFETY INSTRUCTIONS Table of Contents SAFETY INSTRUCTIONS ..........5 GETTING STARTED ............9 Main Features ............10 Appearance .............. 12 INSTALLATION .............. 15 Preparation for Use ..........16 Requirements of Input Power ........16 Input Connection ............. 17 Output Connection ..........19 Power on ..............
Page 4
ASD-1300 User Manual REMOTE OPERAION ............70 Set GPIB address and RS-232 Parameters ....71 Programming ............73 Command List ............76 Command for Remote Control ......... 78 APPENDIX ..............106 Protection .............. 107 Specifications ............109 Certificate Of Compliance ........112 TTL Pin Assignment ..........
SAFETY INSTRUCTIONS AFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition. Safety Symbols These safety symbols may appear in this manual or on the instrument.
Page 6
ASD-1300 User Manual Safety Guidelines Do not place any heavy object on the ASD-1300. General Guideline Avoid severe impact or rough handling that leads to damaging the ASD-1300. CAUTION Do not discharge static electricity to the ASD- 1300.
Page 7
SAFETY INSTRUCTIONS Location: Indoor, no direct sunlight, dust free, Operation almost non-conductive pollution (Note below) Environment Relative Humidity: 20%~ 85% (no condensation) Altitude: < 2000m Temperature: 0°C to 50°C (Pollution Degree) EN61010-1:2010 and EN61010-2-030 specifies the pollution degrees and their requirements as follows. The ASD- 1300 falls under degree 2.
Page 8
ASD-1300 User Manual Power cord for the United Kingdom When using the power supply in the United Kingdom, make sure the power cord meets the following safety instructions. NOTE: This lead/appliance must only be wired by competent persons WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code: Green/ Yellow:...
GW Instek ASD-1300 programmable AC/DC Power Source. In this manual the term, AC/DC Source, stands for GW Instek ASD-1300 programmable AC/DC Source. The AC source is a high efficiency programmable AC power source, which provides low distortion sinusoid AC voltage output and measurement.
ASD-1300 User Manual Main Features Local operation from the front panel keypad. Configuration Protection against Over-power, Over-load, Over-temperature and Fan-fail. Fan speed controlled by output power. Built-in isolation output relays. Remote operation via GPIB or RS-232C ...
GETTING STARTED Accessories Please check the following accessories are included and not damaged during shipment. If there are damaged or shortage of the accessories, please contact GW Instek electronics or sales agent for help. Standard Accessories Part number Description Qty.
ASD-1300 User Manual Appearance Front Panel Menu/ Home SAVE RECALL Page up Page down ESC/ Enter Cancel Unlock TRIGGER POWER Item Symbol Description LCD panel, display setting and measurement results. LED indicator: OUT(green)indicates AC source normal output Trigger TRIGGER(yellow)indicates LIST, PULSE, STEP ...
Page 13
GETTING STARTED At main menu page, press this key to enter Menu/ “advanced parameter setting” page; at other pages, Home press this key to return to main menu. Short press: Cursor move upward or downward. SAVE Long press: long press at “main menu” “advanced parameters setting menu”...
ASD-1300 User Manual Rear Panel OUTPUT INPUT AC1-150V~/AC2-300V~ Max Power 3000W AC220-240V~ SERIAL NO. MODEL REVISION AND BARCODE LABEL Item Name Description TTL I/O port The 9-pin, D-type female connector conveys TTL signals (AC_ ON, FAULT_OUT and REMOTE_INHIBIT) Output connector Output connector. Outputs power to the loading device.
If any damage is found, please file a claim against the carrier immediately. Do not return the instrument to the factory without obtaining the prior RMA acceptance from GW Instek. Preparation for Use ..........16 Requirements of Input Power ........16 Rating ....................
ASD-1300 User Manual Preparation for Use Before start using, the instrument must be connected with an appropriate AC line input. As the instrument is cooled by fans, it must be installed in an area with sufficient space for air circulation. It should be used in an area where the environment temperature does not exceed 40°C.
INSTALLATION Input Connection The input terminal block is located on the instrument rear panel. The temperature rating of the power cord must be rated at least for 85°C. The current rating of input power line must be greater than or equal to the maximum current rating of the AC source. To put a 30A breaker (ABT A) (30A breaker (ABT B)) at AC input side.
Page 18
ASD-1300 User Manual 100-240Vac input terminal To protect operators the wire connected to the GND Caution terminal must be connected to the ground. Under no circumstances shall this AC source be operated without an adequate ground connection. Installation of the AC input power cord must be done by professional personnel in accordance with local electrical codes.
INSTALLATION Output Connection The output terminal block is located at the rear of the AC source. Load connecting to the “N” and “L” is done at the output terminals. To meet the safety requirements, the safety cover must be fastened. The wires to the load must be sufficiently large gauges, so they will not overheat while carrying the output current.
ASD-1300 User Manual Power on Before turning on the instrument, all protective earth Caution terminals, extension cords and devices connected to the instrument must be connected to a protective ground. Any interruption of the protective grounding will cause a potential shock hazard that might injure people.
Page 21
INSTALLATION Error messages will show on the LCD panel if error Caution occurs during turn on process. Please refer to page 107 for error code. Digital circuit inside AC source may not operate properly if turn on AC source immediately after turn off.
Page 22
ASD-1300 User Manual OCAL OERATION The AC source can be configured to operate in local or remote mode. The operation in remote mode through GPIB controller or RS-232C will be described in next chapter. This chapter describes the operation in local mode through the keypad on the front panel for data entry and test.
ASD-1300 User Manual Operation through Keypad and Knob LCD panel will show the self-test message as below after AC switch is on. Then will enter main menu setting page as below. Operator can set output voltage and frequency for three phase. ASD-1300 provides a user-friendly programming interface by using the keypad on the front panel.
Page 25
LOCAL OERATION Press Enter key to enter individual parameter setting menu. After entering individual setting menu, press Up arrow or Down arrow key to move cursor to item which wants to modify. Use numerical key or knob to do the ...
ASD-1300 User Manual Menu Tree Main Menu Advanced parameter setting menu 1. SETUP 2. CONFIG 3. LIST 4. PULSE 5. STEP 6. SYNTHESIS Degree On Wave A Count Count Count Synthesis Waveform Degree Off Wave A CSIN CF List Degree Pulse V Step Vac, DV GAIN...
LOCAL OERATION Main Menu Setting After the AC source is on and initialization is finished, screen will show the main menu. Default output voltage and frequency is 110V and 60Hz. Voltage and frequency setting are shown on the left side of the screen, the right side of the screen shows the measurement value, example shown as below.
Page 28
ASD-1300 User Manual Ipeak/Irms. VA: Apparent power, calculation equation is Vrms x Irms. VAR: Reactive power. Calculation equation is √VA...
LOCAL OERATION Advanced Parameter Setting Menu Press Menu/ Home key at main menu to enter Operation step advanced parameter setting menu. User can select an item from item 1~6, then press Enter key to enter individual setting menu; or press Menu/ Home key to return main menu.
ASD-1300 User Manual SETUP Function Operation step 1. Press numerical key on the front panel to select SETUP option when the advanced parameter setting menu appears. 2. Press Enter key to enter SETUP setting menu shown as figure below. 3.
LOCAL OERATION Definitions for all parameters on SETUP setting menu Setting Range Parameter Definition Unit 150.0/ Maximum voltage setting V Limit 300.0 for all pages Maximum output Irms 32.0/ 16.0 A I Limit Delay time for over current OCP Delay protection Voltage Level Output voltage range 150 V/ 300 V...
ASD-1300 User Manual Voltage Level Setting ASD-1300 provides two output voltage levels used for low voltage and high voltage applications. At 300V level, AC source can provide 16.0A in total;At 150V level 32.0A in total can be provided, maximum output power is 3000VA. When switched from 300V to 150V level, AC source will check all voltage settings, and change the setting to 150V if any voltage setting checked is higher than 150V.
LOCAL OERATION Voltage Sense Voltage Sense determines output voltage measurement point is at AC output terminal (Vout) or input terminal at product under testing (Vsense). If it is set at Vsense, Remote Sense wire must be connected first, otherwise compensate error will occur. AC source will compensate the voltage drop of the output cable if Vsense is selected.
ASD-1300 User Manual 2. Rotate knob to change A to B, then press Enter key to confirm setting. Press OUT key when main menu appears, only A Note or B can be selected. If using LIST function, then A or B waveform can be output alternatively. Buzzer When user press numerical key or rotate knob, there is sound from buzzer to confirm the operation, it can be muted if the confirmation...
LOCAL OERATION Disable (Lock) the Keys at Front Panel ASD-1300 allows the user to lock the parameters at front panel to avoid being changed accidently. After the parameters are locked, only Menu/ Home key and OUT key are enabled. Menu/ Home key allows user to return to SETUP menu to enable or disable lock.
ASD-1300 User Manual Operation step 1. Press Up arrow or Down arrow key to move cursor to V Limit. 2. Change value to 200.0V by numerical key or knob, then press Enter key to confirm setting. I Limit, OCP Delay I Limit represents current limit, it is used to set the maximum output current (rms).
Page 37
LOCAL OERATION Current Output Time Set I Limit is to 3A, OCP delay time to 1.5s. Example Operation step 1. Press Up arrow or Down arrow key to move cursor to I Limit. 2. Change value to 3A by numerical key or knob. Then press Enter key to confirm setting.
ASD-1300 User Manual Degree On, Off ASD-1300 can control the start phase angle (Degree On) or stop phase angle (Degree Off) of the output voltage waveform. Range for Degree On is 0.0~359.9, and is 0.0~360.0 for Degree Off. When Degree Off parameter is 0, the output stopped immediately no matter which output status is at.
LOCAL OERATION Is Start, Is Interval Both the parameter ranges for Is Start and Is Intervals are 0.0~1000.0 msec, these parameters are used for the surge current measurement. Figure shown below indicates the measurement of the surge current starting at Is Start after output enabled, and measurement time is Is Interval.
Page 40
ASD-1300 User Manual 3. Press Up arrow or Down arrow key to move cursor to Is Interval. 4. Change parameter to 200 by numerical key or knob. Then press Enter key to confirm setting.
LOCAL OERATION CONFIG Function Operation step 1. Press numerical key on the front panel to select CONFIG option when the advanced parameter setting menu appears. 2. Press Enter key to enter CONFIG setting menu shown as figure below. 3. Press ► Page down key to enter page 2.
ASD-1300 User Manual Definitions for all parameters on CONFIG setting menu Setting Range Parameter Definition Minimum Maximum Stored group A waveform Wave A Sin, CSin, DST0 ~ DST31 Stored group B waveform Wave B Address of GPIB GPIB ADDR RS232 Odd/Even parity check NONE, EVEN, ODD RS232 Parity RS232 Baud Rate Baud Rate of RS232 9600/ 19200...
Page 43
LOCAL OERATION 2. Rotate Knob to change SIN to DST20. Then press Enter key to confirm setting. The processes to set group B waveform as cut Example sinusoidal waveform (Cut Sine) with CF 1.300. Operation step 1. Press Up arrow or Down arrow key to move cursor to Wave B.
ASD-1300 User Manual GPIB, RS232 Communication Setting ASD-1300 provides remote control operation, there are GPIB or RS232 interface for selection. Only one interface can be activated at the same time. Controller Menu Remote (GPIB) Remote (RS232) (front panel) Operation Mode Menu (front panel) Remote (GPIB) Remote (RS232)
Page 45
LOCAL OERATION Set GPIB address to 20 Example Operation step 1. Press Up arrow or Down arrow key to move cursor to GPIB ADDR. 2. Change parameter to 20 by using numerical key or knob. Then press Enter key to confirm setting.
Page 46
ASD-1300 User Manual 4. Rotate Knob to change parameter from NONE to ODD. Then press Enter key to confirm setting. Inhibit Output by Remote Control Signal (Remote Inhibit) Remote Inhibit is an “Active HIGH” TTL signal, high level is 5V, low level is 0V.
Page 47
LOCAL OERATION Output Time TTL signal changes from LOW to HIGH, and stays high longer than 1ms, ASD-1300 resume output. Output Time TRIG: TTL signal changes from HIGH to LOW, and keeps low longer than 1ms, ASD-1300 turn off output, and stop detecting TTL signal, output stays at off state even TTL signal changes to high.
Page 48
ASD-1300 User Manual The processes to change Remote Inhibit setting Example from OFF to TRIG. Operation step 1. Press Up arrow or Down arrow key to move cursor to Remote Inhibit. 2. Rotate Knob to change parameter value from OFF to TRIG.
Page 49
LOCAL OERATION 4. Press Up arrow or Down arrow key to move cursor to Vdc . 5. Change parameter value to 100.0V by using numerical key or knob. Then press Enter key to confirm setting. 6. Press Up arrow or Down arrow key to move cursor to F.
ASD-1300 User Manual PPLICAION ASD-1300 provides not only stable AC voltage but also three kinds of programmable dynamic output. Following paragraphs explain the definition of parameters and how to do the setting. LIST Mode ............... 51 Definitions for all parameters on LIST setting menu ....52 PULSE Mode ............
APPLICAION LIST Mode Let user to set 10 sets of user-defined voltage, frequency. This mode can make 10 different outputs consecutively. Operation step 1. Press numerical key on the front panel to select LIST option when the advanced parameter setting menu appears. 2.
ASD-1300 User Manual Definitions for all parameters on LIST setting menu Setting Range Parameter Definition Min. Max. Execution times for one complete List waveform. When 0 is setting, output 10000 Count recycles endless until user presses Enter key to stop. Start voltage(V): (high voltage range) 300.0 V Start...
Page 53
APPLICAION Trigger ON ⁄ OFF is output switch for List waveform, output status is current output status. At this page, there are two modes. 1. To output List waveform Press Enter key to start output List waveform. Screen will show information as below during output is in process.
Page 54
ASD-1300 User Manual programmed in LIST mode. There are two ways to leave Trigger page Under output is off, press ESC/ Cancel key will return to LIST initialization page. Press Menu/ Home key return to main menu. Example: Parameter setting Output waveform SEQ - 0...
APPLICAION PULSE Mode Simulate repeatedly changed output voltage. Operation step 1. Press numerical key on the front panel to select PULSE option when the advanced parameter setting menu appears. 2. Press Enter key to enter PULSE setting menu shown as figure below. PULSE mode lets user insert fixed period, programmed special waveform to output waveform set at main menu.
Page 56
ASD-1300 User Manual Degree On Start phase angle of Pulse output 359.9 Select Go to Trigger Mode, press Enter key to Operation step do the setting. Screen shows information as below at trigger page. “Trigger On” represents that the trigger is ...
ASD-1300 User Manual STEP Mode Simulate gradually changed output voltage. Operation step 1. Press numerical key on the front panel to select STEP option when the advanced parameter setting menu appears. 2. Press Enter key to enter STEP setting menu shown as figure below.
Page 59
APPLICAION For each STEP, voltage waveform changes satisfy following equation. + dV Voltage amplitude of next STEP =current voltage amplitude + STEP amplitude change. + dF Voltage frequency of next STEP =current voltage frequency + STEP frequency change. Select Go to Trigger Mode, press Enter key to Operation step ...
Page 60
ASD-1300 User Manual When the output state of AC source is triggered, press OUT key can stop the output. Press OUT key again, the AC source will output the waveform according to the main page setting. Press Enter key, then output state can be ...
APPLICAION SYNTHESIS Self-Defined Waveform Mode ASD-1300 also provides 39 orders of amplitude and phase angle settings for user to make self-defined waveforms. It can be used at all output modes once the self-defined waveforms are created. Operation step 1. Press numerical key on the front panel to select STEP option when the advanced parameter setting menu appears.
ASD-1300 User Manual Definitions for all parameters on STEP setting menu Setting Range Parameter Definition Min. Max. Name for two sets of SYNTHESIS DST30/ DST31 Waveform waveform Order 2~20 33.33 Voltage amplitude ratio of Order 21~30 Harmonic to fundamental wave Gain 30.00 for each order.
ASD-1300 User Manual AVE AND RECALL ASD-1300 provides two modes for users to save and recall the output setting or the system information. Save and recall the main page setting ...... 65 Save the main page setting .............. 65 Recall the main page setting ............66 Save and recall the system setting ......
SAVE AND RECALL Save and recall the main page setting AC power supply provides two modes for users to save and recall the output setting or the system information. Save the main page setting ASD-1300 provides 9 channels for users to save V and F. Users can save or recall the specific parameters at any time.
ASD-1300 User Manual Recall the main page setting The process to recall the setting stored in Channel Example 2 Memory (V=150V, F=80Hz) to the main page. Operation step 1. Press Recall key about 2 seconds in the main page, then it will enter the storage page as shown below.
SAVE AND RECALL Save and recall the system setting ASD-1300 provides 3 groups of memory for users to save the system data. System data contains the parameters in SETUP and CONFIG pages. Enter the PAGE SELECT screen, as shown below. Operation step Save the System Setting Operation step...
ASD-1300 User Manual Recall the System Setting Operation step 1. Press Recall key about 2 seconds in PAGE SELECT page, then it will enter the system data recalling mode, as shown below. 2. Choose a group of Group 1 to Group 3. 3.
ASD-1300 User Manual EMOTE OPERAION ASD-1300 can be remotely controlled by GPIB or RS-232. GPIB interface is an 8-bit parallel data bus with other control lines to manage communication. RS-232 transmits data in series so communication speed is slower. If transmission speed is faster than data processing speed of micro-processor, micro- processor will neglect new received command when there is command under processing.
REMOTE OPERAION Set GPIB address and RS-232 Parameters GPIB and RS-232 parameters can be set at CONFIG page. Default address of GPIB is 30, baud rate of RS232 is 9600, and Even/odd parity check of RS232 is NONE. RS-232 pin definition RS-232 interface uses RXD and TXD signals only.
Page 72
ASD-1300 User Manual remotely. Press - / Unlock key to switch from remote control back to local control, other keys are inactive during remote control. Screen shows >REMOTE< when AC source is Operation step in remote control as below...
REMOTE OPERAION Programming All commands and responses are transferred in the form of ASCII code. Response must be received completely before sending a new command. Otherwise, error will occur due to message disorder. Conventions <> Items in angle brackets are abbreviations for Angle bracket parameter.
Page 74
ASD-1300 User Manual higher level commands are more close to left side, and the more detailed commands are more close to right side. ASD-1300 includes two distinctive types of command. The first one is command described in IEEE 488.2, it is common commands for GPIB, represented by “*”...
Page 75
REMOTE OPERAION VOLT:AC Example 100;:FETCh:CURRent:AC? Explanation The above instruction has two commands which are VOLT:AC 100 and FETCh:CURRent:AC?. Using (;:) can connect commands more than two. Space( ): Space is used to separate command and number or string after it. Instrument differentiates command and data by space.
ASD-1300 User Manual Command for Remote Control IEEE 488.2 Standard Commands *ESE ..................78 *ESR ..................79 *IDN ..................79 *RCL ..................79 *SAV ..................80 *SRE ..................80 *STB ..................81 *CLS ..................81 *ESE Query Set and query flag mask for standard event status Description register.
ASD-1300 User Manual *SAV Save system setting Description Syntax *SAV<NR1> Return parameter 1 *SRE Query Set and return status of flag mask status for Description service events. This command is for a register used for IEE488.2 and instrument defined service events. It is used for flag mask or control enable.
ASD-1300 User Manual FETCH and MEAURE Commands Differences between FETCH and MEASURE are: FETCH reads existing measurements value. Execution time is shorter because data is available immediately but the data may not be the newest. Instrument will send the next sampling data to the computer after receiving MEASURE command.
ASD-1300 User Manual OUTPUT Commands :OUTPut ................. 86 :OUTPut:MODE ..............86 :OUTPut Query Set or query status of turning on or turning off Description output. Syntax OUTPut Query Syntax OUTPut? Parameter/Return parameter Example OUTPut? OUTPut ON Query output status. Turn on output. :OUTPut:MODE Query Set or query output mode.
Page 87
REMOTE OPERAION Example OUTPut:MODE? OUTPut:MODE LIST Query output mode. Set output as LIST mode.
ASD-1300 User Manual Example [:SOURce:]CURRent:INRush:STARt? 200.3 [:SOURce]:CURRent:INRush:INTerval Query Set or query inrush current measurement time. Description Syntax [:SOURce:]CURRent:INRush:INTerval Query Syntax [:SOURce:]CURRent:INRush:INTerval? Effective rang: 0.0 ~ 1000.0 (unit: msec) Parameter/ <NR2> Return parameter Example [:SOURce:]CURRent:INRush:INTerval? 400.8 [:SOURce]:FREQuency Query Set or query frequency of output waveform. Description Syntax [:SOURce:]FREQuency...
REMOTE OPERAION Example [:SOURce:]FUNCtion:SHAPe? [:SOURce]:FUNCtion:SHAPe:A Query Set or query waveform of waveform buffer A. Description Syntax [:SOURce:]FUNCtion:SHAPe:A Query Syntax [:SOURce:]FUNCtion:SHAPe:A? Parameter/ SINE Return parameter CSIN DST<00~31> Example [:SOURce:]FUNCtion:SHAPe:A? CSIN [:SOURce]:FUNCtion:SHAPe:A:CF Query Set or query CF of the cut sin waveform in Description waveform buffer A.
ASD-1300 User Manual Parameter/ SINE Return parameter CSIN DST<00~31> Example [:SOURce:]FUNCtion:SHAPe:B? CSIN [:SOURce]:FUNCtion:SHAPe:B:CF Query Set or query CF of the cut sin waveform in Description waveform buffer B. It is effective only when CSIN is selected for waveform buffer B. Syntax [:SOURce:]FUNCtion:SHAPe:B:CF Query Syntax...
ASD-1300 User Manual [:SOURce]:CONFigure:INHibit Query Set or query remote TTL inhibition operation. Description Refer to page 46 for details. Syntax [:SOURce:]CONFigure:INHibit Query Syntax [:SOURce:]CONFigure:INHibit? Parameter/ Return parameter LIVE TRIG EXCITE Example [:SOURce:]CONFigure:INHibit? LIVE [:SOURce]:PHASe:ON Query Set or query waveform start phase angle set at Description main menu.
REMOTE OPERAION Effective rang: 0 ~ 360.0, 360.0 Parameter/ <NR2> Return parameter representing turn off immediately Example [:SOURce:]PHASe:OFF? 250.5 [:SOURce]:LIST:COUNt Query Set or query how many times to execute effective Description sequences in LIST mode. Syntax [:SOURce:]LIST:COUNt Query Syntax [:SOURce:]LIST:COUNt? Effective rang: 0 ~ 10000 Parameter/ <NR1>...
ASD-1300 User Manual Syntax [:SOURce:]LIST:SHAPe Query Syntax [:SOURce:]LIST:SHAPe? Parameter/ A|B… Return parameter Example [:SOURce:]LIST:SHAPe? A B A A A [:SOURce]:LIST:VOLTage:AC:STARt Query Set or query start voltage of 10 individual Description sequences in LIST mode. Syntax [:SOURce:]LIST:VOLTage:AC:STARt Query Syntax [:SOURce:]LIST:VOLTage:AC:STARt? Effective rang: Parameter/ <NR2>…...
REMOTE OPERAION [:SOURce]:LIST:FREQency:STARt Query Set or query start frequency of 10 individual Description sequences in LIST mode. Syntax [:SOURce:]LIST:FREQency:STARt Query Syntax [:SOURce:]LIST:FREQency:STARt? Effective rang: 30.0 ~ 1000.0 (unit: Hz) Parameter/ <NR2>… Return parameter <NR2> Example [:SOURce:]LIST:FREQency:STARt? 50.8 80.5 2.2 [:SOURce]:LIST:FREQency:END Query Set or query stop frequency of 10 individual Description...
ASD-1300 User Manual Example [:SOURce:]LIST:DEGRee:END? 30.6 96.5 88.0 71 [:SOURce]:PULSe:VOLTage:AC Query Set or query AC voltage of PULSE waveform. Description Syntax [:SOURce:]PULSe:VOLTage:AC Query Syntax [:SOURce:]PULSe:VOLTage:AC? Effective rang: Parameter/ <NR2>… 0.0 ~ 150.0 (low voltage range), Return parameter <NR2> 0.0 ~ 300.0 (high voltage range) Example [:SOURce:]PULSe:VOLTage:AC? 250.1...
REMOTE OPERAION Example [:SOURce:]PULSe:SPHase? 60.0 [:SOURce]:PULSe:COUNt Query Set or query how many times to execute PULSE Description voltage. Syntax [:SOURce:]PULSe:COUNt Query Syntax [:SOURce:]PULSE:COUNt? Effective rang: 0 ~ 10000 Parameter/ <NR1> Return parameter Example [:SOURce:]PULSe:COUNt? [:SOURce]:PULSe:DCYCle Query Set or query execution time of PULSE waveform, Description it must be no longer than total period in PULSE mode.
REMOTE OPERAION Syntax [:SOURce:]SYNThesis:AMPLitude Query Syntax [:SOURce:]SYNThesis:AMPLitude? Effective rang <NR2>… Order N=2~N=20, Gain limit 33.33% <NR2> Order N=21~N=30, Gain limit 30.00% Order N=31~N=39, Gain limit 15.00% [:SOURce:]SYNThesis:AMPLitude? 20.55 33.10 2.55 [:SOURce]:SYNThesis:PHASe Query Set or query phase for each order of harmonic Description waveforms.
APPENDIX Protection ASD-1300 provide software and hardware protections. A1500 SERIES will disconnect output relay to turn off output and display protection status on screen when protection happens. To resume the output, following three errors can be released by long press any key.
Page 108
ASD-1300 User Manual breaker** AC mains voltage higher or lower than rating ranges PFC AC Fault PFC OVP Fault Over voltage for PFC Board PFC OTP Fault Over temperature for PFC Board Output abnormal for PFC Board PFC PG Fault * D2A OCP ranges: When Voltage Level is 300V, OCP trigger point is 16A.
APPENDIX Specifications The operation specifications of ASD-1300 programmable AC Power Source are listed below. All specifications have been tested according to the standard of Delta test procedures. All specifications are based on resistor load with remote sense connection, under 25±2°C unless specified otherwise. Input ratings (AC rms) 220Vac to 240 Vac (single phase) Nominal input voltage...
Page 110
ASD-1300 User Manual 0.15% Accuracy 0.0° to 359.9° variable (setting Output on phase resolution 0.1°) Output voltage stability 0.1% Line regulation 0.1% Load regulation Output voltage waveform distortion ratio, Output voltage response time, Efficiency Output voltage waveform ≤ 1% distortion ratio Output voltage response 100 us (TYP) time...
Page 111
APPENDIX Store and recall settings, Basic settings: 4 (0~3 Memory Function numeric keys) Arbitrary Wave Number 30 (nonvolatile) of memories General Specifications IEEE 488.2 compliant interface GPIB Complies with the EIA-RS-232 Interface RS-232C specifications External Control I/O EXT Control Between input and Withstand 2210 Vac, 1 minute chassis, output and...
ASD-1300 User Manual Certificate Of Compliance GOOD WILL INSTRUMENT CO., LTD. declare that the CE marking mentioned product satisfies all the technical relations application to the product within the scope of council: Directive: EMC; LVD; WEEE; RoHS The product is in conformity with the following standards or other normative documents: ◎...
ASD-1300 User Manual Built In Waveforms DST 0 Order Gain Phase 2.07 9.80 15.80 0.0 2.16 DST 1 Order Gain Phase 1.50 1.50 2.00 DST 2 Order Gain Phase 2.00 1.40 2.00 1.40 1.00...
Page 115
APPENDIX DST 3 Order Gain Phase 2.50 1.90 2.50 1.90 1.10 1.50 1.10 DST 4 Order Gain Phase 1.10 2.80 1.40 2.30 1.50 DST 5 Order Gain Phase 1.65 4.20 3.45 1.05 3.00...
Page 116
ASD-1300 User Manual DST 6 Order Gain Phase 2.20 5.60 2.80 4.60 3.00 1.40 1.00 DST 7 Order Gain Phase 4.90 1.60 2.70 1.40 2.00 1.10 DST 8 Order Gain Phase 7.35 2.40 4.05 2.10 1.05 3.00 1.65 1.05 1.05 1.20 1.05...
Page 117
APPENDIX DST 9 Order Gain Phase 9.80 3.20 5.40 1.20 2.80 1.40 4.00 2.20 1.40 1.40 1.60 1.40 DST 10 Order Gain Phase 17.75 0.0 DST 11 Order Gain Phase 21.25 0.0...
Page 118
ASD-1300 User Manual DST 12 Order Gain Phase 24.50 0.0 DST 13 Order Gain Phase 2.30 9.80 15.80 0.0 2.50 DST 14 Order Gain Phase 1.15 4.90 7.90 1.25...
Page 119
APPENDIX DST 15 Order Gain Phase 1.15 4.90 DST 16 Order Gain Phase 11.00 180.0 4.05 2.00 180.0 1.30 DST 17 Order Gain Phase 7.17 3.42 180.0 0.80...
Page 120
ASD-1300 User Manual DST 18 Order Gain Phase 8.11 3.48 180.0 1.00 DST 19 Order Gain Phase 9.38 3.44 180.0 1.15 DST 20 Order Gain Phase 2.06 180.0 1.77 1.62 180.0 1.23 0.91 180.0 0.54 0.51 0.53 180.0...