Programmable AC Power Supply ASD-1150 USER MANUAL ISO-9001 CERTIFIED MANUFACTURER...
Page 2
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company. The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
SAFETY INSTRUCTIONS Table of Contents SAFETY INSTRUCTIONS ..........5 GETTING STARTED ............9 Main Features ............10 Appearance .............. 12 INSTALLATION .............. 16 Preparation for Use ..........17 Requirements of Input Power ........17 Input Connection ............. 18 Output Connection ..........20 Power on ..............
Page 4
ASD-1150 User Manual REMOTE OPERAION ............74 Set GPIB address and RS-232 Parameters ....75 Programming ............77 Command List ............80 Command for Remote Control ......... 83 APPENDIX ..............114 Protection .............. 115 Specifications ............117 Certificate Of Compliance ........121 TTL Pin Assignment ..........
SAFETY INSTRUCTIONS AFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition. Safety Symbols These safety symbols may appear in this manual or on the instrument.
Page 6
ASD-1150 User Manual Safety Guidelines Do not place any heavy object on the ASD-1150. General Guideline Avoid severe impact or rough handling that leads to damaging the ASD-1150. CAUTION Do not discharge static electricity to the ASD- 1150.
Page 7
SAFETY INSTRUCTIONS Location: Indoor, no direct sunlight, dust free, Operation almost non-conductive pollution (Note below) Environment Relative Humidity: 20%~ 85% (no condensation) Altitude: < 2000m Temperature: 0°C to 50°C (Pollution Degree) EN61010-1:2010 and EN61010-2-030 specifies the pollution degrees and their requirements as follows. The ASD- 1150 falls under degree 2.
Page 8
ASD-1150 User Manual Power cord for the United Kingdom When using the power supply in the United Kingdom, make sure the power cord meets the following safety instructions. NOTE: This lead/appliance must only be wired by competent persons WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code: Green/ Yellow:...
GW Instek ASD-1150 programmable AC/DC Power Source. In this manual the term, AC/DC Source, stands for GW Instek ASD-1150 programmable AC/DC Source. The AC source is a high efficiency programmable AC power source, which provides low distortion sinusoid AC voltage output and measurement.
ASD-1150 User Manual Main Features Local operation from the front panel keypad. Configuration Protection against Over-power, Over-load, Over-temperature and Fan-fail. Fan speed controlled by output power. Built-in isolation output relays. Remote operation via GPIB or RS-232C ...
GETTING STARTED Accessories Please check the following accessories are included and not damaged during shipment. If there are damaged or shortage of the accessories, please contact GW Instek electronics or sales agent for help. Standard Accessories Part number Description Qty.
ASD-1150 User Manual Appearance Front Panel Menu/ Home SAVE RECALL Page up Page down METER INDIV ESC/ Enter Cancel Unlock OUTPUT TRIGGER AC 250V MAX. 10A POWER Item Symbol Description LCD panel, display setting and measurement results. LED indicator: OUT(green)indicates AC source normal output ...
Page 13
GETTING STARTED At main menu page, press this key to enter Menu/ “advanced parameter setting” page; at other pages, Home press this key to return to main menu. Short press: Cursor move upward or downward. SAVE Long press: long press at “main menu” “advanced parameters setting menu”...
Page 14
ASD-1150 User Manual Main switch: Turn on/off the AC power input POWER...
GETTING STARTED Rear Panel BREAKER MAX 10A INPUT AC100- 240V SENSING SENSING TTL I/O RS-232 SYNC GPIB OUTPUT AC 0V-300V Max Power 1500VA AC 0V-424V Max Power 750VA Item Name Description Power Line input Input connector. Please refer to page 18 Connector for connection SYNC port...
Page 16
If any damage is found, please file a claim against the carrier immediately. Do not return the instrument to the factory without obtaining the prior RMA acceptance from GW Instek. Preparation for Use ..........17 Requirements of Input Power ........17 Rating ....................
INSTALLATION Preparation for Use Before start using, the instrument must be connected with an appropriate AC line input. As the instrument is cooled by fans, it must be installed in an area with sufficient space for air circulation. It should be used in an area where the environment temperature does not exceed 40°C.
ASD-1150 User Manual Input Connection The input terminal block is located on the instrument rear panel. The temperature rating of the power cord must be rated at least for 85°C. The current rating of input power line must be greater than or equal to the maximum current rating of the AC source.
Page 19
INSTALLATION 100-240Vac input terminal To protect operators the wire connected to the GND Caution terminal must be connected to the ground. Under no circumstances shall this AC source be operated without an adequate ground connection. Installation of the AC input power cord must be done by professional personnel in accordance with local electrical codes.
ASD-1150 User Manual Output Connection The output terminal block is located at the rear of the AC source. Load connecting to the “N” and “L” is done at the output terminals. To meet the safety requirements, the safety cover must be fastened. The wires to the load must be sufficiently large gauges, so they will not overheat while carrying the output current.
INSTALLATION Power on Before turning on the instrument, all protective earth Caution terminals, extension cords and devices connected to the instrument must be connected to a protective ground. Any interruption of the protective grounding will cause a potential shock hazard that might injure people.
Page 22
ASD-1150 User Manual Error messages will show on the LCD panel if error Caution occurs during turn on process. Please refer to page 115 for error code. Digital circuit inside AC source may not operate properly if turn on AC source immediately after turn off.
LOCAL OERATION OCAL OERATION The AC source can be configured to operate in local or remote mode. The operation in remote mode through GPIB controller or RS-232C will be described in next chapter. This chapter describes the operation in local mode through the keypad on the front panel for data entry and test.
Page 24
ASD-1150 User Manual Couple (output mode selection) ............48 Inhibit Output by Remote Control Signal (Remote Inhibit) ..49 Power On State ................. 51...
LOCAL OERATION Operation through Keypad and Knob LCD panel will show the self-test message as below after AC switch is on. Then will enter main menu setting page as below. Operator can set output voltage and frequency for three phase. ASD-1150 provides a user-friendly programming interface by using the keypad on the front panel.
Page 26
ASD-1150 User Manual Press Enter key to enter individual parameter setting menu. After entering individual setting menu, press Up arrow or Down arrow key to move cursor to item which wants to modify. Use numerical key or knob to do the ...
LOCAL OERATION Menu Tree Main Menu Advanced parameter setting menu 1. SETUP 2. CONFIG 3. LIST 4. PULSE 5. STEP 6. SYNTHESIS Degree On Wave A Count Count Count Synthesis Waveform Degree Off Wave A CSIN CF List Degree Pulse V Step Vac, DV GAIN I Spike Start...
ASD-1150 User Manual Main Menu Setting After the AC source is on and initialization is finished, screen will show the main menu. Default output voltage and frequency is 110V and 60Hz. Voltage and frequency setting are shown on the left side of the screen, the right side of the screen shows the measurement value, example shown as below.
Page 29
LOCAL OERATION Ipeak/Irms. VA: Apparent power, calculation equation is Vrms x Irms. VAR: Reactive power. Calculation equation is √VA...
ASD-1150 User Manual Advanced Parameter Setting Menu Press Menu/ Home key at main menu to enter Operation step advanced parameter setting menu. User can select an item from item 1~6, then press Enter key to enter individual setting menu;...
LOCAL OERATION SETUP Function Operation step 1. Press numerical key on the front panel to select SETUP option when the advanced parameter setting menu appears. 2. Press Enter key to enter SETUP setting menu shown as figure below. 3. Press ► Page down key to move to next page. 4.
ASD-1150 User Manual Definitions for all parameters on SETUP setting menu Setting Range Parameter Definition Unit 150.0/ Maximum voltage setting V Limit 300.0 for all pages Maximum output Irms 16.0/ 8.0 A I Limit Delay time for over current OCP Delay protection Voltage Level Output voltage range 150 V/ 300 V...
LOCAL OERATION Voltage Level Setting ASD-1150 provides two output voltage levels used for low voltage and high voltage applications. At 300V level, AC source can provide 8.0A in total;At 150V level 16.0A in total can be provided, maximum output power is 1500VA. When switched from 300V to 150V level, AC source will check all voltage settings, and change the setting to 150V if any voltage setting checked is higher than 150V.
ASD-1150 User Manual Voltage Sense Voltage Sense determines output voltage measurement point is at AC output terminal (Vout) or input terminal at product under testing (Vsense). If it is set at Vsense, Remote Sense wire must be connected first, otherwise compensate error will occur. AC source will compensate the voltage drop of the output cable if Vsense is selected.
LOCAL OERATION 2. Rotate knob to change A to B, then press Enter key to confirm setting. Press OUT key when main menu appears, only A or Note B can be selected. If using LIST function, then A or B waveform can be output alternatively.
ASD-1150 User Manual Disable (Lock) the Keys at Front Panel ASD-1150 allows the user to lock the parameters at front panel to avoid being changed accidently. After the parameters are locked, only Menu/ Home key and OUT key are enabled. Menu/ Home key allows user to return to SETUP menu to enable or disable lock.
Page 37
LOCAL OERATION 2. Change value to 200.0V by numerical key or knob, then press Enter key to confirm setting. AC voltage limit setting is related to voltage level Caution range. Maximum limit AC voltage at 300V range is 300V, while is 150V at 150V output range. ...
ASD-1150 User Manual 5. Change parameter to -150.0V by using numerical key and - / Unlock key, then press Enter key to confirm setting. DC Voltage limit also is related to Voltage Level range. Caution Maximum DC limit voltage at 300V range is ±424.2V, and is ±212.1V at 150V output range.
Page 39
LOCAL OERATION Current Output Time Set I Limit is to 3A, OCP delay time to 1.5s. Example Operation step 1. Press Up arrow or Down arrow key to move cursor to I Limit. 2. Change value to 3A by numerical key or knob. Then press Enter key to confirm setting.
ASD-1150 User Manual Degree On, Off ASD-1150 can control the start phase angle (Degree On) or stop phase angle (Degree Off) of the output voltage waveform. Range for Degree On is 0.0~359.9, and is 0.0~360.0 for Degree Off. When Degree Off parameter is 0, the output stopped immediately no matter which output status is at.
LOCAL OERATION Is Start, Is Interval Both the parameter ranges for Is Start and Is Intervals are 0.0~1000.0 msec, these parameters are used for the surge current measurement. Figure shown below indicates the measurement of the surge current starting at Is Start after output enabled, and measurement time is Is Interval.
Page 42
ASD-1150 User Manual 4. Change parameter to 200 by numerical key or knob. Then press Enter key to confirm setting.
LOCAL OERATION CONFIG Function Operation step 1. Press numerical key on the front panel to select CONFIG option when the advanced parameter setting menu appears. 2. Press Enter key to enter CONFIG setting menu shown as figure below. 3. Press ► Page down key to enter page 2. 4.
ASD-1150 User Manual Definitions for all parameters on CONFIG setting menu Setting Range Parameter Definition Minimum Maximum Stored group A waveform Wave A Sin, CSin, SQUA, DST0 ~ DST31 Stored group B waveform Wave B Address of GPIB GPIB ADDR RS232 Odd/Even parity check NONE, EVEN, ODD RS232 Parity RS232 Baud Rate Baud Rate of RS232...
Page 45
LOCAL OERATION 2. Rotate Knob to change SIN to DST20. Then press Enter key to confirm setting. The processes to set group B waveform as cut Example sinusoidal waveform (Cut Sine) with CF 1.300. Operation step 1. Press Up arrow or Down arrow key to move cursor to Wave B.
ASD-1150 User Manual GPIB, RS232 Communication Setting ASD-1150 provides remote control operation, there are GPIB or RS232 interface for selection. Only one interface can be activated at the same time. Controller Menu Remote (GPIB) Remote (RS232) (front panel) Operation Mode Menu (front panel) Remote (GPIB) Remote (RS232)
Page 47
LOCAL OERATION Set GPIB address to 20 Example Operation step 1. Press Up arrow or Down arrow key to move cursor to GPIB ADDR. 2. Change parameter to 20 by using numerical key or knob. Then press Enter key to confirm setting.
Page 48
ASD-1150 User Manual 4. Rotate Knob to change parameter from NONE to ODD. Then press Enter key to confirm setting. Couple (output mode selection) AC/DC source can output AC, DC, AC+DC three kinds of output voltage. User can set the output mode at CONFIG menu. Set Couple as AC+DC Mode Example Operation step...
Page 49
LOCAL OERATION Inhibit Output by Remote Control Signal (Remote Inhibit) Remote Inhibit is an “Active HIGH” TTL signal, high level is 5V, low level is 0V. Its signals send to 9 pin TTL signal connector (D- SUB) at back panel of the AC source. Please refer to appendix “TTL Pin Assignment”...
Page 50
ASD-1150 User Manual TRIG: TTL signal changes from HIGH to LOW, and keeps low longer than 1ms, ASD-1150 turn off output, and stop detecting TTL signal, output stays at off state even TTL signal changes to high. User must press OUT key to resume output and enable TTL signal detecting.
Page 51
LOCAL OERATION Power On State ASD-1150 may set different operation modes or enable turn on output immediately next time after AC is on. Output voltage and frequency depend on stored parameter before AC input is off. The process to set output is ON, output mode is Example AC+DC, AC voltage is 100Vac, DC voltage is 50Vdc, frequency is 50Hz for next time after AC is...
Page 52
ASD-1150 User Manual 5. Change parameter value to 100.0V by using numerical key or knob. Then press Enter key to confirm setting. 6. Press Up arrow or Down arrow key to move cursor to Vdc. 7. Change parameter value to 50 by using numerical key or knob.
Page 53
LOCAL OERATION 11. Rotate Knob to change parameter value from AC to AC+DC. Then press Enter key to confirm setting. Refer to section “Save the main page setting“ on page 69 for Group save function, save all settings to Group 1, it will take effect next time after AC is on.
ASD-1150 User Manual PPLICAION ASD-1150 provides not only stable AC voltage but also three kinds of programmable dynamic output. Following paragraphs explain the definition of parameters and how to do the setting. LIST Mode ............... 55 Definitions for all parameters on LIST setting menu ....56 PULSE Mode ............
APPLICAION LIST Mode Let user to set 10 sets of user-defined voltage, frequency. This mode can make 10 different outputs consecutively. Operation step 1. Press numerical key on the front panel to select LIST option when the advanced parameter setting menu appears. 2.
ASD-1150 User Manual Definitions for all parameters on LIST setting menu Setting Range Parameter Definition Min. Max. Execution times for one complete List waveform. When 0 is setting, output 10000 Count recycles endless until user presses Enter key to stop. Start voltage(V): (high voltage range) 300.0 V Start...
Page 57
APPLICAION Screen shows Trigger On, Stop is current output status. Press Enter key to trigger output. Then screen shows as below, Trigger Off and Running are displayed, waiting user to stop LIST output. When all sequences and Count are executed completed, AC output turned off, and screen shows Stop.
Page 58
ASD-1150 User Manual Output waveform SEQ - 0 SEQ - 1 45° 72 ms 100 ms...
APPLICAION PULSE Mode Simulate repeatedly changed output voltage. Operation step 1. Press numerical key on the front panel to select PULSE option when the advanced parameter setting menu appears. 2. Press Enter key to enter PULSE setting menu shown as figure below. PULSE mode lets user insert fixed period, programmed special waveform to output waveform set at main menu.
Page 60
ASD-1150 User Manual Degree On Start phase angle of Pulse output 359.9 Select Go to Trigger Mode, press Enter key to Operation step do the setting. Screen shows information as below at trigger page. “Trigger On” represents that the trigger is ...
ASD-1150 User Manual STEP Mode Simulate gradually changed output voltage. Operation step 1. Press numerical key on the front panel to select STEP option when the advanced parameter setting menu appears. 2. Press Enter key to enter STEP setting menu shown as figure below.
Page 63
APPLICAION Frequency change for each step. May be -150.0 150.0 positive or negative Output duration for each step 60000 Dwell Degree On Initial phase angle for each STEP 359.9 For each STEP, voltage waveform changes satisfy following equation. + dV ...
Page 64
ASD-1150 User Manual When AC source is at Running (triggered on) status, press OUT key can stop the output. Press OUT key again, the outputs are according those settings at main menu. Press Enter key can switch output settings ...
APPLICAION SYNTHESIS Self-Defined Waveform Mode ASD-1150 also provides 39 orders of amplitude and phase angle settings for user to make self-defined waveforms. It can be used at all output modes once the self-defined waveforms are created. Operation step 1. Press numerical key on the front panel to select STEP option when the advanced parameter setting menu appears.
ASD-1150 User Manual Definitions for all parameters on STEP setting menu Setting Range Parameter Definition Min. Max. Name for two sets of SYNTHESIS DST30/ DST31 Waveform waveform Order 2~20 33.33 Voltage amplitude ratio of Order 21~30 Harmonic to fundamental wave Gain 30.00 for each order.
ASD-1150 User Manual AVE AND RECALL ASD-1150 provides two modes for users to save and recall the output setting or the system information. Save and recall the main page setting ...... 69 Save the main page setting .............. 69 Recall the main page setting ............70 Save and recall the system setting ......
SAVE AND RECALL Save and recall the main page setting AC power supply provides two modes for users to save and recall the output setting or the system information. Save the main page setting ASD-1150 provides 9 channels for users to save V and F. Users can save or recall the specific parameters at any time.
ASD-1150 User Manual Recall the main page setting The process to recall the setting stored in Channel Example 2 Memory (V=150V, F=80Hz) to the main page. Operation step 1. Press Recall key about 2 seconds in the main page, then it will enter the storage page as shown below.
SAVE AND RECALL Save and recall the system setting ASD-1150 provides 3 groups of memory for users to save the system data. System data contains the parameters in SETUP and CONFIG pages. Enter the PAGE SELECT screen, as shown below. Operation step Save the System Setting Operation step...
ASD-1150 User Manual Recall the System Setting Operation step 1. Press Recall key about 2 seconds in PAGE SELECT page, then it will enter the system data recalling mode, as shown below. 2. Choose a group of Group 1 to Group 3. Press Enter key and then the system data will saved in the memory of the chosen group will be recalled.
ASD-1150 User Manual EMOTE OPERAION ASD-1150 can be remotely controlled by GPIB or RS-232. GPIB interface is an 8-bit parallel data bus with other control lines to manage communication. RS-232 transmits data in series so communication speed is slower. If transmission speed is faster than data processing speed of micro-processor, micro- processor will neglect new received command when there is command under processing.
REMOTE OPERAION Set GPIB address and RS-232 Parameters GPIB and RS-232 parameters can be set at CONFIG page. Default address of GPIB is 30, baud rate of RS232 is 9600, and Even/odd parity check of RS232 is NONE. RS-232 pin definition RS-232 interface uses RXD and TXD signals only.
Page 76
ASD-1150 User Manual remotely. Press - / Unlock key to switch from remote control back to local control, other keys are inactive during remote control. Screen shows >REMOTE< when AC source is Operation step in remote control as below...
REMOTE OPERAION Programming All commands and responses are transferred in the form of ASCII code. Response must be received completely before sending a new command. Otherwise, error will occur due to message disorder. Conventions <> Items in angle brackets are abbreviations for Angle bracket parameter.
Page 78
ASD-1150 User Manual higher level commands are more close to left side, and the more detailed commands are more close to right side. ASD-1150 includes two distinctive types of command. The first one is command described in IEEE 488.2, it is common commands for GPIB, represented by “*”...
Page 79
REMOTE OPERAION Explanation The above instruction has two commands which are VOLT:AC 100 and FETCh:CURRent:AC?. Using (;:) can connect commands more than two. Space( ): Space is used to separate command and number or string after it. Instrument differentiates command and data by space. Example VOLT:AC 100 Star(*):...
REMOTE OPERAION Command for Remote Control IEEE 488.2 Standard Commands *ESE ..................83 *ESR ..................84 *IDN ..................84 *RCL ..................84 *SAV ..................85 *SRE ..................85 *STB ..................86 *CLS ..................86 *ESE Query Set and query flag mask for standard event status Description register.
ASD-1150 User Manual Power on Syntax *ESE<NR1> Query Syntax *ESE? Parameter/ <NR1> Return parameter *ESR Query Query value stored in standard events status Description register, 1 represents event occurs. Query Syntax *ESR? *IDN Query Return string of AC source identification. Description Query Syntax *IDN?
REMOTE OPERAION *SAV Save system setting Description Syntax *SAV<NR1> Return parameter 1 *SRE Query Set and return status of flag mask status for Description service events. This command is for a register used for IEE488.2 and instrument defined service events. It is used for flag mask or control enable.
REMOTE OPERAION FETCH and MEAURE Commands Differences between FETCH and MEASURE are: FETCH reads existing measurements value. Execution time is shorter because data is available immediately but the data may not be the newest. Instrument will send the next sampling data to the computer after receiving MEASURE command.
REMOTE OPERAION OUTPUT Commands :OUTPut ................. 91 :OUTPut:MODE ..............91 :OUTPut:COUPling ............. 92 :OUTPut Query Set or query status of turning on or turning off Description output. Syntax OUTPut Query Syntax OUTPut? Parameter/Return parameter Example OUTPut? OUTPut ON Query output status. Turn on output.
ASD-1150 User Manual Example OUTPut:MODE? OUTPut:MODE LIST Query output mode. Set output as LIST mode. :OUTPut:COUPling Query Set of query output coupling mode. Description Syntax OUTPut:COUPling Query Syntax OUTPut:COUPling? Parameter/Return parameter ACDC Example OUTPut:COUPling? OUTPut:COUPling AC Query output coupling mode. Set output coupling mode as AC.
ASD-1150 User Manual [:SOURce]:FUNCtion:SHAPe Query Set or query waveform buffer. AC source provides Description two waveform buffers, user has to select A or B waveform buffer. Syntax [:SOURce:]FUNCtion:SHAPe Query Syntax [:SOURce:]FUNCtion:SHAPe? Parameter/Return parameter Example [:SOURce:]FUNCtion:SHAPe? [:SOURce]:FUNCtion:SHAPe:A Query Set or query waveform of waveform buffer A. Description Syntax [:SOURce:]FUNCtion:SHAPe:A...
REMOTE OPERAION Effective rang: 1.200 ~ 1.414 Parameter/ <NR2> Return parameter Example [:SOURce:]FUNCtion:SHAPe:A:CF? 1.234 [:SOURce]:FUNCtion:SHAPe:B Query Set or query waveform of waveform buffer B. Description Syntax [:SOURce:]FUNCtion:SHAPe:B Query Syntax [:SOURce:]FUNCtion:SHAPe:B? Parameter/ SINE Return parameter CSIN DST<00~31> Example [:SOURce:]FUNCtion:SHAPe:B? CSIN [:SOURce]:FUNCtion:SHAPe:B:CF Query Set or query CF of the cut sin waveform in Description...
ASD-1150 User Manual Return parameter HIGH Example [:SOURce:]VOLTage:RANGe? HIGH [:SOURce]:VOLTage:SENSe Query Set or query remote sense function. REMOTE is in Description enable, VOUT is in disable Syntax [:SOURce:]VOLTage:SENSe Query Syntax [:SOURce:]VOLTage:SENSe? Parameter/ VOUT Return parameter REMOTE Example [:SOURce:]VOLTage:SENSe? REMOTE [:SOURce]:CONFigure:INHibit Query Set or query remote TTL inhibition operation.
REMOTE OPERAION [:SOURce]:PHASe:ON Query Set or query waveform start phase angle set at Description main menu. Syntax [:SOURce:]PHASe:ON Query Syntax [:SOURce:]PHASe:ON? Effective rang: 0.0 ~ 359.9 Parameter/ <NR2> Return parameter Example [:SOURce:]PHASe:ON? 200.5 [:SOURce]:PHASe:OFF Query Set or query waveform stop phase angle set at Description main menu.
ASD-1150 User Manual Example [:SOURce:]LIST:COUNt? [:SOURce]:LIST:DWELl Query Set or query execution time of 10 individual Description sequences in LIST mode. Syntax [:SOURce:]LIST:DWELl Query Syntax [:SOURce:]LIST:DWELl? Effective rang: 0 ~ 60000 (unit: msec) Parameter/ <NR2>… Return parameter <NR2> Example [:SOURce:]LIST:DWELI? 60000 1 23 95 [:SOURce]:LIST:SHAPe Query Set or query waveform buffer used for 10...
REMOTE OPERAION Effective rang: Parameter/ <NR2>… Return parameter 0.0 ~ 150.0 (low voltage range), <NR2> 0.0 ~ 300.0 (high voltage range) Example [:SOURce:]LIST:VOLTage:AC:STARt? 110 22.5 55.6 [:SOURce]:LIST:VOLTage:AC:END Query Set or query stop voltage of 10 individual Description sequences in LIST mode. Syntax [:SOURce:]LIST:VOLTage:AC:END Query Syntax...
ASD-1150 User Manual [:SOURce]:LIST:VOLTage:DC:END Query Set or query DC end voltage for 10 individual Description sequences in LIST mode. Syntax [:SOURce:]LIST:VOLTage:DC:END Query Syntax [:SOURce:]LIST:VOLTage:DC:END? Effective rang: Parameter/ <NR2>… Return parameter ±212.1 (low voltage range), <NR2> ±424.2 (high voltage range) Example [:SOURce:]LIST:VOLTage:DC:END? 1.2 -50 66.6 [:SOURce]:LIST:FREQency:STARt...
REMOTE OPERAION Effective rang: 0 ~ 10000 Parameter/ <NR1> Return parameter Example [:SOURce:]PULSe:COUNt? [:SOURce]:PULSe:DCYCle Query Set or query execution time of PULSE waveform, Description it must be no longer than total period in PULSE mode. Syntax [:SOURce:]PULSe:DCYCle Query Syntax [:SOURce:]PULSE:DCYCle? Effective rang: 1 ~ 59999 (unit: msec) Parameter/ <NR1>...
REMOTE OPERAION Syntax [:SOURce:]SYNThesis Query Syntax [:SOURce:]SYNThesis? Parameter/ DST30 Return parameter DST31 Example [:SOURce:]SYNThesis? DST30 [:SOURce]:SYNThesis:AMPLitude Query Set or query voltage amplitude for each order of Description harmonic waveforms. Maximum order is 39. Syntax [:SOURce:]SYNThesis:AMPLitude Query Syntax [:SOURce:]SYNThesis:AMPLitude? Effective rang <NR2>…...
Page 112
ASD-1150 User Manual Other Commands :TRIG ..................112 :VERion:DSP? ..............112 :VERion:LCM? ..............112 :VERion:UI? ................. 113 :TRIG Query Set or query trigger status. Disabled when Description OUTPut:MODE = FIXE. Syntax :TRIG Query Syntax :TRIG? Parameter Return parameter OFF RUNNING Example TRIG? TRIG ON...
Page 113
REMOTE OPERAION Return Parameter S00E02 Example :VERion:LCM? S00E02 :VERion:UI? Query Query firmware version of UI. Description Syntax :VERion:UI? Return Parameter S00E02 Example :VERion:UI? S00E02...
APPENDIX Protection ASD-1150 provide software and hardware protections. A1500 SERIES will disconnect output relay to turn off output and display protection status on screen when protection happens. To resume the output, following three errors can be released by long press any key.
Page 116
ASD-1150 User Manual Input current reaches trip point of built in no fuse Input OCP breaker* AC mains voltage higher or lower than rating ranges PFC AC Fault PFC OVP Fault Over voltage for PFC Board PFC OTP Fault Over temperature for PFC Board Output abnormal for PFC Board PFC PG Fault D2A OCP ranges: When Voltage Level is 300V, OCP trigger point...
APPENDIX Specifications The operation specifications of ASD-1150 programmable AC Power Source are listed below. All specifications have been tested according to the standard of Delta test procedures. All specifications are based on resistor load with remote sense connection, under 25±2°C unless specified otherwise. Input ratings (AC rms) 100 Vac to 240 Vac (single phase) Nominal input voltage...
Page 118
ASD-1150 User Manual Setting resolution 0.1 Hz (30.0 to 1000.0 Hz) 0.15% Accuracy 0.0° to 359.9° variable (setting Output on phase resolution 0.1°) Output rating for DC mode -212 V to +212 V/ -424 V to +424 V Setting Range Setting Resolution 0.1 V Voltage ±...
Page 119
APPENDIX 0.000 to 1.000 Range Load power factor 0.001 Resolution 1.2 to 1.414 Range Load crest factor 0.001 Resolution Up to 39th order of the Harmonic voltage Range fundamental wave Others UVP, OCP, OPP, SCP, OTP, FAN FAIL Protections STN-LCD, 6 inch Display Store and recall settings, Basic settings: 4 (0~3 Memory Function...
Page 120
ASD-1150 User Manual EN 61010-1 Safety Operating Indoor use, Overvoltage Category II environment Operating 0 °C to 40 °C temperature range Environment Storage -40 °C to 85 °C temperature range Storage humidity 90 % RH or less (no condensation) range Up to 2000 m Altitude 425(W) ×...
APPENDIX Certificate Of Compliance GOOD WILL INSTRUMENT CO., LTD. declare that the CE marking mentioned product satisfies all the technical relations application to the product within the scope of council: Directive: EMC; LVD; WEEE; RoHS The product is in conformity with the following standards or other normative documents: ◎...
APPENDIX Built In Waveforms DST 0 Order Gain Phase 2.07 9.80 15.80 0.0 2.16 DST 1 Order Gain Phase 1.50 1.50 2.00 DST 2 Order Gain Phase 2.00 1.40 2.00 1.40 1.00...
Page 124
ASD-1150 User Manual DST 3 Order Gain Phase 2.50 1.90 2.50 1.90 1.10 1.50 1.10 DST 4 Order Gain Phase 1.10 2.80 1.40 2.30 1.50 DST 5 Order Gain Phase 1.65 4.20 3.45 1.05 3.00...
Page 125
APPENDIX DST 6 Order Gain Phase 2.20 5.60 2.80 4.60 3.00 1.40 1.00 DST 7 Order Gain Phase 4.90 1.60 2.70 1.40 2.00 1.10 DST 8 Order Gain Phase 7.35 2.40 4.05 2.10 1.05 3.00 1.65 1.05 1.05 1.20 1.05...
Page 126
ASD-1150 User Manual DST 9 Order Gain Phase 9.80 3.20 5.40 1.20 2.80 1.40 4.00 2.20 1.40 1.40 1.60 1.40 DST 10 Order Gain Phase 17.75 0.0 DST 11 Order Gain Phase 21.25 0.0...
Page 127
APPENDIX DST 12 Order Gain Phase 24.50 0.0 DST 13 Order Gain Phase 2.30 9.80 15.80 0.0 2.50 DST 14 Order Gain Phase 1.15 4.90 7.90 1.25...
Page 128
ASD-1150 User Manual DST 15 Order Gain Phase 1.15 4.90 DST 16 Order Gain Phase 11.00 180.0 4.05 2.00 180.0 1.30 DST 17 Order Gain Phase 7.17 3.42 180.0 0.80...