SYMBOLS
Symbols
Err 3
Err 9
Function/Meaning
The pressure value
did not increase during
inflation. The cuff is not
applied correctly.
Blood pressure monitor
internal error.
Data is transferring
Airplane mode
Appears when a body
or arm movement is
detected.
Appears during
measurement when the
cuff is attached loosely.
Firmware is updating
Recommended Action
Apply the cuff correctly,
and take another
measurement.
Press the reset button.
After reset, clock data
will disappear. If the
error still appears,
contact A&D Medical.
Measurements can be
taken but device will not
transmit data wirelessly.
The reading may yield
an incorrect value. Take
another measurement.
Remain still during
measurement.
The reading may yield
an incorrect value.
Apply the cuff correctly,
and take another
measurement.