Selection
Frequency
Cal Factor
Samples per Measurement
RF Power
mW/dBm
Zero/Cal/Zero+Cal
Highest Value
Lowest Value
Captured Value
Information Window

Software Access from PS-Cal

1. In PS-Cal, select the Tools menu (see Figure 12 below).
2. Click 28xx RF Power
Figure 4: Accessing 28xx RF Power Display program from PS-Cal
2818A-900 Revision AB
Description
Select desired frequency to measure
Calibration factor for selected frequency. By default, this value is
downloaded from the EEPROM of the 2818A but it can be changed
manually if desired.
Number of samples between each measurement displayed. This
value can be between 1 and 20
Power measured at TEST SENSOR port
Measurement unit
Use to Zero or calibrate the unit to its internal 50 MHz reference
Highest value recorded. Use "Reset" button to clear
Lowest value recorded. Use "Reset" button to clear
Press "Capture" to hold value
Diagnostic information
Table 1: RF Power Display feature description list
RF Power Transfer Standard
12