ALARM TEMPERATURE SETTING

This unit is provided with the high and low temperature alarm and the temperature at which the alarm is
activated is changeable.
The following procedure shows the setting of alarm temperature according to the condition below:
High temperature alarm: activates at the temperature 5
Low temperature alarm: activates at the temperature 5
Note:
The alarm temperature is set at the factory 10
The available range of alarm temperature is between 5
Table 3. Procedure for setting high temperature alarm
Description of operation
1
Press numerical value shift key for
2
about 5 seconds.
Press numerical value shift key
3
and scroll the figure to 1.
4
Press set key.
Scroll the figure to 005 by using
5
digit shift key and numerical value
shift key
6
Press set key.
Note:
The alarm temperature is set at the factory 10
The available range of alarm temperature is between 5
Table 4. Procedure for setting low temperature alarm
Description of operation
1
Press numerical value shift key for
2
about 5 seconds
Press numerical value shift key
3
and scroll the figure to 2
4
Press set key
Scroll the figure to -05 by using
5
digit shift key and numerical value
shift key
6
Press set key.
• The alarm temperature set mode returns to the temperature display mode automatically when 90
seconds has passed without any key operation.
o
C higher than the set temperature
o
C lower than the set temperature
o
C higher than the set temperature.
o
o
C and 20
C higher than the set temperature.
Key operated
----
The current chamber temperature is
displayed.
The first digit is flashed.
The first digit is flashed.
SET
The first digit is flashed.
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
moves.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
o
C lower than the set temperature.
o
o
C and 20
C lower than the set temperature.
Key operated
----
The current chamber temperature is
displayed.
The first digit is flashed.
The first digit is flashed.
SET
The first digit is flashed.
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
moves.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
15
Indication after operation
Indication after operation