ALARM TEMPERATURE SETTING

This unit is provided with both high and low temperature alarms. The temperature at which the alarm is
activated can be changed.
The available set range for high temperature alarm is between +5
low temperature alarm against the set temperature.
Note: The temperature alarm is set at ±10
Display
As an example, Table 3 shows the procedure to set the high temperature alarm so that the alarm can
activate when the chamber temperature is 5
Table 4 shows the procedure to set the low temperature alarm so that the alarm can activate when the
chamber temperature is 5
Table 3. Procedure for setting high temperature alarm
Description of operation
1
Press numerical value shift key for
2
about 5 seconds.
Press numerical value shift key
3
and scroll the figure to 1.
4
Press set key.
Scroll the figure to 005 by using
5
digit shift key and numerical value
shift key
6
Press set key.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
Table 4. Procedure for setting low temperature alarm
Description of operation
1
Press numerical value shift key for
2
about 5 seconds.
Press numerical value shift key
3
and scroll the figure to 2.
4
Press set key.
Scroll the figure to -05 by using
5
digit shift key and numerical value
shift key
6
Press set key.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
o
C of the set temperature at the factory.
Mode
High temperature alarm set
Low temperature alarm set
o
C higher than the set temperature.
o
C lower than the set temperature.
Key operated
Key operated
o
C and +15
See Table 3 on page 15
See Table 4 on page 15
Indication after operation
The current chamber temperature is
----
displayed.
The first digit is flashed.
The first digit is flashed.
SET
The first digit is flashed.
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
moves.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
Indication after operation
The current chamber temperature is
----
displayed.
The first digit is flashed.
The first digit is flashed.
SET
The first digit is flashed.
When pressed, the figure of settable
digit changes.
When pressed, the changeable digit
moves.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
15
o
o
o
C, and -5
C and -15
C for
Function