ALARM TEMPERATURE SETTING

This unit is provided with both high and low temperature alarms. The temperature at which the alarm is
activated may be changed.
The available set range for high temperature alarm is between +5
low temperature alarm against the set temperature.
Note: The temperature alarm is set at ±10
Display
F01
F02
As an example, Table 3 shows the procedure to set the high temperature alarm so that the alarm can
activate when the chamber temperature is 5
Table 4 shows the procedure to set the low temperature alarm so that the alarm can activate when the
chamber temperature is 5
Table 3 Procedure for setting high temperature alarm
Description of operation
1
Press numerical value shift key for
2
five seconds.
Press numerical value shift key and
3
scroll the figure to 1.
4
Press set key.
Set the temperature to 005 with the
5
digit shift key and numerical value
shift key.
6
Press set key.
The set mode returns to the temperature display mode automatically when ninety seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
Mode
High temperature alarm set
Low temperature alarm set
o
C lower than the set temperature.
o
C and +15
o
C of the set temperature at the factory.
o
C higher than the set temperature.
Key
operated
----
The current chamber temperature is
displayed.
The first digit is flashed.
When pressed, the figure of settable
digit increases.
SET
The right digit is flashed.
Pressing the key shifts the digit which can be set.
When pressed, the figure of settable
digit increases.
Alarm temperature is memorized and
SET
the current chamber temperature is
displayed.
15
o
o
C, and -5
C and -15
Function
See Table 3 on page 15
See Table 4 on page 16