Specifications
General Specifications
Display
Maximum Load
Meter Power
Auto Power OFF
Current output capability
Voltage input impedance
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Dimensions
Weight
Range Specifications
Mode
DC
Measure
DC Source
Dot matrix LCD
1000 ohms @ 24mA
6 AA batteries or AC adaptor
Meter automatically powers off after 10 minutes of inactivity
24mADC at 1000 ohms
10kohms minimum
41ºF to 104ºF (5ºC to 40ºC)
o
- 4
F to 140
Max 80% up to 87ºF (31ºC) decreasing linearly to 50% at 104ºF
(40ºC)
<80%
7000ft (2000meters) maximum
6.3 x 3.2 x 1.7") (159 x 80 x 44 mm)
8.3 oz. (234g)
Function
Range (Resolution)
Current
0 to 50mA (0.01mA)
Percent (%)
-25% to +230% (0.1%)
Voltage
0 to 1999mV (1mV)
(autorange)
2 to 20V (0.01V)
Current
0 to 24mA (0.01mA)
Percent (%)
-25% to +125% (0.1%)
Voltage
0 to 2000mV (1mV)
0 to 20V (0.01V)
o
F (5ºC to 40ºC)
8
Accuracy (% of reading)
± (0.01% + 1 digit)
PRC15-en-US_V1.5 03/21
Need help?
Do you have a question about the PRC15 and is the answer not in the manual?
Questions and answers