Opera FLO AUTO Instructions Manual page 15

Mattress system
Table of Contents

Advertisement

Power unit
is working,
but mattress
replacement is not
inflating
and/or
Bottoming out is
occurring
and/or
Patient leaves a
deep indentation
at the contact area
which does not
return back to its
original shape.
Alternating or
Constant mode is
not available.
Patient's wounds
are not responding
to pressure relief
(reddening of skin).
Check CPR Pull Cord.
Check if mattress' air hose
couplers are properly
connected to power unit's
quick release couplers.
Verify that patient weight
setting is correct.
Inspect air filter for dust and
dirt.
Lift mattress coverlet up
to check if air cells are
connected correctly.
Lift mattress coverlet up to
check if air tubes are kinked or
obstructed.
Check if air cells are cut or
cracked
Possible control failure.
Unzip mattress cover and
inspect CPR latch is fully
inserted into socket.
Secure air hose couplers firmly
into place.
Increase or decrease weight
setting until appropriate
pressure is reached.
Clean or replace air filter.
Make sure all air cells are
properly linked to air supply.
Check and adjust air tubes
positions.
Please contact Opera on 0333
222 8584 for assistance.
Please contact Opera on 0333
222 8584 for assistance.
Contact your physician and/or
nursing service immediately.
operabeds.com | 0333 222 8584 | 14