11. Routine Maintenance and Storage
Remove air filter from the rear panel of the power unit by opening up the filter
cap. Inspect the filter for dirt or dust, and clean it with mild soap and water.
Reinsert the dried air filter after cleaning and ensure that the cap is secure. If a
replacement is needed, contact Opera on 0333 222 8584.
Only disinfected and dry systems should be stored. Disconnect the air hoses
from the power unit. Roll up the mattress starting from the head end and working
down toward the foot end. Use the straps to secure it and store in clean plastic or
other storage bag. Store in a cool, dark place.
12. Troubleshooting
Fault
Power unit is not
working.
Low pressure alert
does not go out in
30 sec. after system
starts.
Alert LED is lit during
operation.
13 | Opera Beds | Opera Flo Auto Mattress System Instructions and Technical Specifications | Version 2
Inspection Procedure
Check if power cord is firmly
plugged into both the control
unit and the electrical outlet.
Check if the power switch is in
the on position.
Check if the power surge has
shut down the power unit.
Make sure there is no power
failure.
Power unit does not respond
to possible solutions.
Possible control failure.
Check if there is leakage in air
tubes* or air cells.
Remedy
Secure power cord into control
unit and/or electrical outlet.
Turn power switch to on position.
A power surge may overload the
circuitry temporarily. Turn the
unit off, and check the fuse for
damage. Turn the unit on again
with normal procedure.
Turn on and operate the unit
after power is restored to the
facility or home.
Please contact Opera on 0333
222 8584 for assistance.
Please contact Opera on 0333
222 8584 for assistance.